Literature DB >> 15364707

Dynamic contour tonometry: a comparative study on human cadaver eyes.

Christoph Kniestedt1, Michelle Nee, Robert L Stamper.   

Abstract

OBJECTIVE: To compare intraocular pressure measurements obtained by recently introduced dynamic contour tonometry (DCT), Goldmann applanation tonometry (GAT), pneumatonometry (PTG), and intracameral manometry in human cadaver eyes.
METHODS: Sixteen freshly enucleated human cadaver eyes were deepithelialized and dehydrated with dextran. A tube was placed in the anterior chamber and connected to a transducer and to a bottle system filled with balanced salt solution. The pressure in the eye was then altered between 5 mm Hg and 58 mm Hg. Intraocular pressure measurements were obtained with DCT, GAT, and PTG at each manometric pressure reading.
RESULTS: On average, DCT values measured 0.58 +/- 0.70 mm Hg higher than real intracameral pressure. The GAT and PGT showed consistently lower values, -4.01 +/- 1.76 mm Hg and -5.09 +/- 2.61 mm Hg, respectively. At all bottle heights, DCT values were significantly closer to the reference pressure than GAT and PTG (P<.001).
CONCLUSIONS: Measurement with DCT provides IOP values significantly closer to true manometric levels than either GAT or PTG. Further studies are warranted to determine its reliability in patients and the effect of corneal thickness.

Entities:  

Mesh:

Year:  2004        PMID: 15364707     DOI: 10.1001/archopht.122.9.1287

Source DB:  PubMed          Journal:  Arch Ophthalmol        ISSN: 0003-9950


  32 in total

1.  Ocular pulse amplitude after trabeculectomy.

Authors:  Sandra R von Schulthess; Claude Kaufmann; Lucas M Bachmann; Ahmet Yanar; Michael A Thiel
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-07-23       Impact factor: 3.117

Review 2.  [Dynamic contour tonometry].

Authors:  C Kniestedt; H E Kanngiesser
Journal:  Ophthalmologe       Date:  2006-08       Impact factor: 1.059

3.  Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplasties.

Authors:  A R Ismail; M Lamont; S Perera; D Khan-Lim; R Mehta; J D A Macleod; D F Anderson
Journal:  Br J Ophthalmol       Date:  2007-07       Impact factor: 4.638

4.  Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Fatih Özcura; Nilgün Yildirim; Afsun Şahin; Ertuğrul Çolak
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

5.  Effect of hevin deletion in mice and characterization in trabecular meshwork.

Authors:  Min Hyung Kang; Dong-Jin Oh; Douglas J Rhee
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-04-06       Impact factor: 4.799

6.  Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients.

Authors:  Milena Pahlitzsch; Jeanette Brünner; Johannes Gonnermann; Anna-Karina B Maier; Necip Torun; Eckart Bertelmann; Matthias Kj Klamann
Journal:  Int J Ophthalmol       Date:  2016-11-18       Impact factor: 1.779

7.  [Mobile intraocular pressure measurement. From palpation to initial clinical experience with the handheld dynamic contour tonometer].

Authors:  U Schmid; C Kniestedt
Journal:  Ophthalmologe       Date:  2010-07       Impact factor: 1.059

8.  Evaluation of a contact lens-embedded sensor for intraocular pressure measurement.

Authors:  Michael D Twa; Cynthia J Roberts; Huikai J Karol; Ashraf M Mahmoud; Paul A Weber; Robert H Small
Journal:  J Glaucoma       Date:  2010-08       Impact factor: 2.503

9.  [Evaluation of dynamic contour tonometry in keratoconus].

Authors:  N Schädle; J D Unterlauft; T Klink; G Geerling
Journal:  Ophthalmologe       Date:  2009-06       Impact factor: 1.059

10.  Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients.

Authors:  J Jimenez-Roman; F Gil-Carrasco; A Martinez; O Albis-Donado; J D De la Torre-Tovar
Journal:  Int Ophthalmol       Date:  2012-11-16       Impact factor: 2.031

View more