[Markerless patient registration. A new technique for image-guided surgery of the lateral base of the skull].

R Marmulla1, J Mühling, G Eggers, S Hassfeld.   

Abstract

BACKGROUND: Markerless patient registration is a new procedure that may reduce logistical efforts and possibly also the radiation load on the patients prior to a computer-assisted intervention. Congruent surfaces, such as bone surfaces or skin surfaces, represented in a data set and in the surgical site, can be overlapped with the help of surface matching. Hitherto, these techniques were performed by using the periorbital and frontal skin of the face. The present clinical study assessed the usability of the auricle as a skin structure for markerless patient registration in lateral skull base surgery. PATIENTS AND METHODS: The periauricular precision of the navigation system SSN was investigated clinically in ten patients. Radiopaque registration markers in the periauricular region served as a target in order to evaluate the markerless registration method.
RESULTS: Exact patient registration failed in five cases due to temporary deformation of the auricle caused by the head support during CT imaging. An excellent patient registration (mean target registration error = 0.9+/-0.2 mm) was found as long as the auricles had not been temporarily deformed during CT imaging.
CONCLUSION: Laser-scan registration of the auricle reduces logistics in image-guided lateral skull base surgery. The problem of temporary deformation of the auricle during CT imaging might be solved by an appropriate opening in the head support at the level of the auricles.

Entities:  

Mesh:

Year:  2005        PMID: 15221084     DOI: 10.1007/s00106-004-1101-5

Source DB:  PubMed          Journal:  HNO        ISSN: 0017-6192            Impact factor:   1.284


  23 in total

1.  Accuracy of computer navigation in ear, nose, throat surgery: the influence of matching strategy.

Authors:  J Claes; E Koekelkoren; F L Wuyts; G M Claes; L Van den Hauwe; P H Van de Heyning
Journal:  Arch Otolaryngol Head Neck Surg       Date:  2000-12

2.  Comparison of six three-dimensional navigation systems during sinus surgery.

Authors:  M Cartellieri; F Vorbeck; J Kremser
Journal:  Acta Otolaryngol       Date:  2001-06       Impact factor: 1.494

3.  [Applications for a robot in the lateral skull base. Evaluation of robot-assisted mastoidectomy in an anatomic specimen].

Authors:  P K Plinkert; B Plinkert; A Hiller; J Stallkamp
Journal:  HNO       Date:  2001-07       Impact factor: 1.284

Review 4.  [Robotics. A new dimension in otorhinolaryngology?].

Authors:  P A Federspil; J Stallkamp; P K Plinkert
Journal:  HNO       Date:  2001-07       Impact factor: 1.284

5.  [A new computer-aided surgical approach for reconstruction of the orbit].

Authors:  R Marmulla; H Niederdellmann; B Lorenz; R Dammer; C Niederdellmann
Journal:  Klin Monbl Augenheilkd       Date:  1998-11       Impact factor: 0.700

6.  Malleable registration mask: application of a novel registration method in image guided sinus surgery.

Authors:  F D Albritton; T T Kingdom; J M DelGaudio
Journal:  Am J Rhinol       Date:  2001 Jul-Aug

7.  A noninvasive, reattachable skull fiducial marker system. Technical note.

Authors:  M A Howard; M B Dobbs; T M Simonson; W E LaVelle; M A Granner
Journal:  J Neurosurg       Date:  1995-08       Impact factor: 5.115

8.  Automated rejection of contaminated surface measurements for improved surface registration in image guided neurosurgery.

Authors:  R Bucholz; W Macneil; P Fewings; A Ravindra; L McDurmont; C Baumann
Journal:  Stud Health Technol Inform       Date:  2000

9.  Experience with various 3-dimensional navigation systems in head and neck surgery.

Authors:  A R Gunkel; W Freysinger; W F Thumfart
Journal:  Arch Otolaryngol Head Neck Surg       Date:  2000-03

10.  Computer-assisted orbital surgery.

Authors:  L Klimek; M Wenzel; R Mösges
Journal:  Ophthalmic Surg       Date:  1993-06
View more
  3 in total

1.  [New developments in computer-assisted surgery (CAS). From intraoperative imaging to ultrasound-based navigation].

Authors:  P A Federspil
Journal:  HNO       Date:  2009-10       Impact factor: 1.284

2.  Validity of the 3D VECTRA photogrammetric surface imaging system for cranio-maxillofacial anthropometric measurements.

Authors:  Philipp Metzler; Yi Sun; Wolfgang Zemann; Alexander Bartella; Marc Lehner; Joachim Anton Obwegeser; Astrid L Kruse-Gujer; Heinz-Theo Lübbers
Journal:  Oral Maxillofac Surg       Date:  2013-04-05

3.  Does Head Orientation Influence 3D Facial Imaging? A Study on Accuracy and Precision of Stereophotogrammetric Acquisition.

Authors:  Giuditta Battistoni; Diana Cassi; Marisabel Magnifico; Giuseppe Pedrazzi; Marco Di Blasio; Benedetta Vaienti; Alberto Di Blasio
Journal:  Int J Environ Res Public Health       Date:  2021-04-17       Impact factor: 3.390

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.