Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method.

Susan M Renner1, Tae-Hong Lim, Whoan-Jeang Kim, Leonid Katolik, Howard S An, Gunnar B J Andersson.   

Abstract

STUDY
DESIGN: Axial pullout tests using fresh cadaveric thoracolumbar vertebral bodies.
OBJECTIVES: To evaluate the effect of a new injectable calcium phosphate cement on the axial pullout strength of both revised and augmented pedicle screws in comparison with polymethyl methacrylate and in terms of injection method. SUMMARY OF BACKGROUND DATA: Failure of pedicle screws by loosening and back out remains a significant clinical problem and is of particular concern for patients with low bone quality. Polymethyl methacrylate was shown to significantly improve the screw pullout strength. However, polymethyl methacrylate is known to have a high polymerization temperature, which may damage surrounding tissues, and a short handling time, and it lacks long-term biocompatibility. Bone mineral cements such as calcium phosphate have a longer working time, very low thermal effect, and are biodegradable as well as having good mechanical strength. Recently, new calcium phosphate cement with improved infiltration properties for better injectability has been introduced, but its performance in augmenting the pedicle screw fixation has not been tested yet.
METHODS: The bone mineral densities of 52 vertebral bodies (T11-L5) were measured using dual-energy x-ray absorptiometry. In each vertebral body, a 6.5-mm-diameter and 45 +/- 5-mm-long pedicle screw was inserted into either the right or left pedicle, representing an initial intact implantation. These intact screws were pulled axially until failure at 10 mm/min. Following failure of the intact pedicle, 3.0 cc of cement was injected into the failed screw hole, representing a revision case, and the prepared screw hole in the contralateral intact pedicle representing an augmentation case. The cement was injected either to the distal tip of the screw hole (calcium phosphate-1 group, n = 19) or along the entire length of the screw hole (calcium phosphate-2 group, n = 20), and the screws were inserted. The cement was then allowed to cure for 24 hours at room temperature before both screws were pulled to failure. In 13 specimens, polymethyl methacrylate was injected along the entire length of the screw hole (polymethyl methacrylate group). Kruskal-Wallis and Mann-Whitney tests were used to compare the screw pullout strengths for study groups, whereas linear relationships between variables were assessed with scatter plots and Spearman correlation coefficients with a significance level of 0.05.
RESULTS: Mean bone mineral densities of all groups were similar. A significant positive correlation was seen between bone mineral density and intact pullout strength. In revision, the pullout strength of calcium phosphate-1 was similar to that of intact, whereas the pullout strength of calcium phosphate-2 and polymethyl methacrylate was significantly greater than that of intact. In augmentation, all 3 injection methods significantly improved the pullout strength over intact. Injection of the calcium phosphate cement along the entire screw length was found to produce significantly higher pullout strengths than injection only at the distal tip of the screw in revision case. Injection of polymethyl methacrylate produced significantly higher pullout strengths than the injection of calcium phosphate by either method in both revision and augmentation.
CONCLUSION: Results of this study demonstrate that the new calcium phosphate cement can improve the axial pullout strength of revised and augmented pedicle screws when injected along the entire length of the screw. This suggests that the injection method may be crucial for revision of failed pedicle screws. Considering inherent properties more favorable for in vivo application, such as nonexothermal polymerization and longer working time, and significant improvement in pullout strength, the new calcium phosphate cement may be a good alternative to polymethyl methacrylate for the augmentation of pedicle screw fixation.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15167670     DOI: 10.1097/00007632-200406010-00020

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  45 in total

1.  Biomechanical evaluation of the Total Facet Arthroplasty System® (TFAS®): loading as compared to a rigid posterior instrumentation system.

Authors:  Simon G Sjovold; Qingan Zhu; Anton Bowden; Chad R Larson; Peter M de Bakker; Marta L Villarraga; Jorge A Ochoa; David M Rosler; Peter A Cripton
Journal:  Eur Spine J       Date:  2012-03-10       Impact factor: 3.134

2.  [Pedicle screw augmentation from a biomechanical perspective].

Authors:  V Bullmann; U R Liljenqvist; R Rödl; T L Schulte
Journal:  Orthopade       Date:  2010-07       Impact factor: 1.087

3.  Biomechanical evaluation of an expansive pedicle screw in calf vertebrae.

Authors:  Wei Lei; Zixiang Wu
Journal:  Eur Spine J       Date:  2005-04-30       Impact factor: 3.134

4.  Stability of anterior vertebral body screws after kyphoplasty augmentation. An experimental study to compare anterior vertebral body screw fixation in soft and cured kyphoplasty cement.

Authors:  O Linhardt; C Lüring; J Matussek; C Hamberger; T Herold; W Plitz; J Grifka
Journal:  Int Orthop       Date:  2006-04-04       Impact factor: 3.075

5.  A case of pedicle screw loosening treated by modified transpedicular screw augmentation with polymethylmethacrylate.

Authors:  Suk-Hyung Kang; Kyoung-Tae Kim; Seung Won Park; Young-Baeg Kim
Journal:  J Korean Neurosurg Soc       Date:  2011-01-31

6.  Local bone quality measurements correlates with maximum screw torque at the femoral diaphysis.

Authors:  Christopher M McAndrew; Avinesh Agarwalla; Adam C Abraham; Eric Feuchtbaum; William M Ricci; Simon Y Tang
Journal:  Clin Biomech (Bristol, Avon)       Date:  2018-02-03       Impact factor: 2.063

7.  Pull-out strength of patient-specific template-guided vs. free-hand fluoroscopically controlled thoracolumbar pedicle screws: a biomechanical analysis of a randomized cadaveric study.

Authors:  A Aichmair; M Moser; M R Bauer; E Bachmann; J G Snedeker; M Betz; M Farshad
Journal:  Eur Spine J       Date:  2017-03-04       Impact factor: 3.134

8.  Analysis of bone cement distribution around fenestrated pedicle screws in low bone quality lumbosacral vertebrae.

Authors:  Sergio Gómez González; Gerard Cabestany Bastida; Maria Daniela Vlad; José López López; Pablo Buenestado Caballero; Luis Alvarez-Galovich; Maria Rodríguez-Arguisjuela; Enrique Fernández Aguado
Journal:  Int Orthop       Date:  2018-08-23       Impact factor: 3.075

9.  Bioabsorbable tricalcium phosphate bone cement strengthens fixation of suture anchors.

Authors:  Rayshad Oshtory; Derek P Lindsey; Nicholas J Giori; Faisal M Mirza
Journal:  Clin Orthop Relat Res       Date:  2010-06-03       Impact factor: 4.176

10.  Salvaging the pullout strength of stripped screws in osteoporotic bone.

Authors:  Pierre H M Pechon; Simon C Mears; Evan R Langdale; Stephen M Belkoff
Journal:  Geriatr Orthop Surg Rehabil       Date:  2013-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.