O Linhardt1, C Lüring, J Matussek, C Hamberger, T Herold, W Plitz, J Grifka.
Abstract
The goal of this cadaver study was to compare the stability of anterior vertebral body screws after implantation in soft or cured kyphoplasty cement. Anterior vertebral body screws were inserted in a total of 30 thoracolumbar vertebrae of ten different human specimens: ten screws were implanted in non-augmented vertebrae (group 1), ten screws were placed in soft cement (group 2), and ten screws were placed in cured cement (group 3). The screws were then tested for biomechanical axial pullout resistance. Mean axial pullout strength was 192 N (range: 10-430 N) in group 1, 364 N (range: 65-875 N) in group 2, and 271 N (range: 35-625 N) in group 3. The paired Student's t-test demonstrated a significant difference between pullout strength of groups 1 and 2 (p= 0.0475). No significant difference was seen between pullout strength of groups 1 and 3 (p= 0.2646) and between groups 2 and 3 (p= 0.3863). We achieved a 1.9 times higher pullout strength with kyphoplasty augmentation of osteoporotic vertebrae compared with the pullout strength of non-augmented vertebrae. Implantation of anterior vertebral body screws in cured cement is a satisfactory method. With this method we found a 1.4 times higher pullout strength than non-augmented vertebrae.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16586135 PMCID: PMC3172758 DOI: 10.1007/s00264-006-0100-4
Source DB: PubMed Journal: Int Orthop ISSN: 0341-2695 Impact factor: 3.075