Literature DB >> 20521128

Bioabsorbable tricalcium phosphate bone cement strengthens fixation of suture anchors.

Rayshad Oshtory1, Derek P Lindsey, Nicholas J Giori, Faisal M Mirza.   

Abstract

BACKGROUND: Failure of suture anchor fixation in rotator cuff repair can occur at different interfaces. Prior studies show fixation at the bone-anchor interface can be augmented using polymethylmethacrylate (PMMA) cement, and screw fixation into bone can be strengthened using bioabsorbable tricalcium phosphate cement. QUESTIONS/PURPOSES: We wished to determine whether augmentation of suture anchor fixation using bioabsorbable tricalcium phosphate cement would increase pullout strength of suture anchors from bone and the number of cycles to failure, to determine the mode of failure after cement augmentation, and to compare strength and mode of failure with those after augmentation with PMMA.
METHODS: We used 10 matched pairs of cadaveric proximal humeri and implanted a metal screw-type suture anchor in one side and on the other side injected tricalcium phosphate cement into the anchor holes before anchor placement. We tested all specimens to failure using a ramped cyclic loading protocol.
RESULTS: Tricalcium phosphate cement augmentation increased the final load to failure by 29% and the number of cycles to failure by 20%. Visual inspection confirmed that failure occurred at the cement-bone interface.
CONCLUSIONS: Tricalcium phosphate cement appears to augment suture anchor fixation into bone, reducing the risk of anchor pullout and failure. CLINICAL RELEVANCE: When relying on suture anchor fixation in bone of questionable quality, we suggest considering augmentation of suture anchor fixation with bioabsorbable cement. This method also provides potential for bioabsorbability and may be more amenable to arthroscopic application.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20521128      PMCID: PMC2974857          DOI: 10.1007/s11999-010-1412-7

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  21 in total

1.  Mode of failure for rotator cuff repair with suture anchors identified at revision surgery.

Authors:  Craig A Cummins; George A C Murrell
Journal:  J Shoulder Elbow Surg       Date:  2003 Mar-Apr       Impact factor: 3.019

2.  The use of acrylic bone cement for suture anchoring.

Authors:  D C Meyer; H A C Jacob; W Pistoia; A von Roll; C Gerber
Journal:  Clin Orthop Relat Res       Date:  2003-05       Impact factor: 4.176

3.  Pullout strength of suture anchors used in rotator cuff repair.

Authors:  Markus J Tingart; Maria Apreleva; David Zurakowski; Jon J P Warner
Journal:  J Bone Joint Surg Am       Date:  2003-11       Impact factor: 5.284

4.  Association of osteopenia of the humeral head with full-thickness rotator cuff tears.

Authors:  Dominik C Meyer; Sandro F Fucentese; Bruno Koller; Christian Gerber
Journal:  J Shoulder Elbow Surg       Date:  2004 May-Jun       Impact factor: 3.019

5.  Revision rotator cuff repair: factors influencing results.

Authors:  M Djurasovic; G Marra; J S Arroyo; R G Pollock; E L Flatow; L U Bigliani
Journal:  J Bone Joint Surg Am       Date:  2001-12       Impact factor: 5.284

6.  Biomechanical evaluation of calcium phosphate cement-augmented fixation of unstable intertrochanteric fractures.

Authors:  S Elder; E Frankenburg; J Goulet; D Yetkinler; R Poser; S Goldstein
Journal:  J Orthop Trauma       Date:  2000-08       Impact factor: 2.512

7.  Biomechanical study of pedicle screw fixation in severely osteoporotic bone.

Authors:  Stephen D Cook; Samantha L Salkeld; Tom Stanley; Albert Faciane; Scot D Miller
Journal:  Spine J       Date:  2004 Jul-Aug       Impact factor: 4.166

8.  Complications of rotator cuff surgery in which bioabsorbable anchors are used.

Authors:  Thomas Magee; Marc Shapiro; Galen Hewell; David Williams
Journal:  AJR Am J Roentgenol       Date:  2003-11       Impact factor: 3.959

9.  Evaluation of calcium sulfate paste for augmentation of lumbar pedicle screw pullout strength.

Authors:  Michael T Rohmiller; Dugan Schwalm; R Chris Glattes; Tarek G Elalayli; Dan M Spengler
Journal:  Spine J       Date:  2002 Jul-Aug       Impact factor: 4.166

10.  Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method.

Authors:  Susan M Renner; Tae-Hong Lim; Whoan-Jeang Kim; Leonid Katolik; Howard S An; Gunnar B J Andersson
Journal:  Spine (Phila Pa 1976)       Date:  2004-06-01       Impact factor: 3.468

View more
  6 in total

1.  Self-setting calcium orthophosphate formulations.

Authors:  Sergey V Dorozhkin
Journal:  J Funct Biomater       Date:  2013-11-12

2.  [Footprint and augmented anchorage].

Authors:  L K L Postl; C Kirchhoff
Journal:  Oper Orthop Traumatol       Date:  2012-11       Impact factor: 1.154

3.  Suture anchor fixation strength with or without augmentation in osteopenic and severely osteoporotic bones in rotator cuff repair: a biomechanical study on polyurethane foam model.

Authors:  Mehmet Serhan Er; Levent Altinel; Mehmet Eroglu; Ozgur Verim; Teyfik Demir; Halil Atmaca
Journal:  J Orthop Surg Res       Date:  2014-08-22       Impact factor: 2.359

4.  Unsupervised Machine Learning-Based Analysis of Clinical Features, Bone Mineral Density Features and Medical Care Costs of Rotator Cuff Tears.

Authors:  Tong-Fu Wang; De-Sheng Chen; Jia-Wang Zhu; Bo Zhu; Zeng-Liang Wang; Jian-Gang Cao; Cai-Hong Feng; Jun-Wei Zhao
Journal:  Risk Manag Healthc Policy       Date:  2021-09-22

5.  Biomechanical properties of a suture anchor system from human allogenic mineralized cortical bone matrix for rotator cuff repair.

Authors:  Jakob E Schanda; Barbara Obermayer-Pietsch; Gerhard Sommer; Philipp R Heuberer; Brenda Laky; Christian Muschitz; Klaus Pastl; Eva Pastl; Christian Fialka; Rainer Mittermayr; Johannes Grillari; Ines Foessl
Journal:  BMC Musculoskelet Disord       Date:  2022-05-05       Impact factor: 2.562

6.  The Primary Stability of a Bioabsorbable Poly-L-Lactic Acid Suture Anchor for Rotator Cuff Repair Is Not Improved with Polymethylmethacrylate or Bioabsorbable Bone Cement Augmentation.

Authors:  Mehmet F Güleçyüz; Michael Kraus-Petersen; Christian Schröder; Andreas Ficklscherer; Markus U Wagenhäuser; Christian Braun; Peter E Müller; Matthias F Pietschmann
Journal:  HSS J       Date:  2017-10-17