Preventricular far-field sensing in the atrial channel of dual chamber pacemakers--an occasional cause of inappropriate mode switch.

Christof Kolb1, Selim Aratma, Bernhard Zrenner, Claus Schmitt.   

Abstract

INTRODUCTION AND AIMS OF THE STUDY: Atrial oversensing may trigger false positive mode switch to an asynchronous mode in dual chamber pacemakers. While myopotential oversensing and far-field R wave sensing within the postventricular atrial refractory period are well characterized, data about oversensing (near-field P wave or far-field R wave) within the atrioventricular delay is limited. Aim of the study was to determine the incidence of preventricular oversensing in the atrial channel of current dual chamber pacemakers.
METHODS: Consecutive patients with dual chamber pacemakers who were in sinus rhythm and who showed no myopotential oversensing were included in the study. Atrial sensitivity was programmed to the maximal available value and atrioventricular delay was prolonged when necessary for intrinsic atrioventricular conduction.
RESULTS: Ten out of 100 (10%) patients showed oversensing within the atrioventricular delay at highest levels of atrial sensitivity, median sensing threshold for the signals was 0.35 mV, and the median coupling interval between preventricular atrial oversensing and the ventricular sensed event was 20 ms.
CONCLUSION: The incidence of inappropriate mode switch caused by atrial preventricular oversensing related to near-field P wave or far-field R wave oversensing in dual chamber pacemakers is up to 10%. Interpretation of pacemaker Holters should consider oversensing within the atrioventricular delay as a differential diagnosis to true episodes of atrial tachyarrhythmias.

Entities:  

Mesh:

Year:  2004        PMID: 15133360     DOI: 10.1023/B:JICE.0000026917.82516.eb

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  10 in total

Review 1.  A reappraisal of pacemaker timing cycles pertaining to automatic mode switching.

Authors:  R X Stroobandt; S S Barold; F D Vandenbulcke; R J Willems; A F Sinnaeve
Journal:  J Interv Card Electrophysiol       Date:  2001-12       Impact factor: 1.900

2.  Stability of far field R wave signals in different conditions.

Authors:  F J C Cools; R R C van Twembeke; J Backers; G A Verpooten
Journal:  Europace       Date:  2003-04       Impact factor: 5.214

3.  Bipolar ventricular far-field signals in the atrium.

Authors:  G Fröhlig; Z Helwani; O Kusch; M Berg; H Schieffer
Journal:  Pacing Clin Electrophysiol       Date:  1999-11       Impact factor: 1.976

4.  Impact and prevention of far-field sensing in fallback mode switches.

Authors:  Pierre Bordacher; Serge Cazeau; Laurence Graindorg; Philippe Ritter
Journal:  Pacing Clin Electrophysiol       Date:  2003-01       Impact factor: 1.976

5.  P wave and far-field R wave detection in pacemaker patient atrial electrograms.

Authors:  H Theres; W Sun; W Combs; E Panken; H Mead; G Baumann; K Stangl
Journal:  Pacing Clin Electrophysiol       Date:  2000-04       Impact factor: 1.976

6.  Far-field R wave oversensing in a dual chamber arrhythmia management device: predisposing factors and practical implications.

Authors:  S Weretka; R Becker; T Hilbel; C Karle; B R Osswald; W Kuebler; W Schoels
Journal:  Pacing Clin Electrophysiol       Date:  2001-08       Impact factor: 1.976

7.  High rate atrial tachyarrhythmia detections in implantable pulse generators: low incidence of false-positive detections. The PA Clinical Trial Investigators.

Authors:  S M Fitts; M R Hill; R Mehra; A M Gillis
Journal:  Pacing Clin Electrophysiol       Date:  2000-07       Impact factor: 1.976

8.  Sensing of atrial fibrillation by a dual chamber pacemaker: how should atrial sensing be programmed to ensure adequate mode shifting?

Authors:  H Walfridsson; M Aunes; M Capocci; N Edvardsson
Journal:  Pacing Clin Electrophysiol       Date:  2000-07       Impact factor: 1.976

9.  Far-field QRS complex sensing: prevalence and timing with bipolar atrial leads.

Authors:  J Brandt; W Worzewski
Journal:  Pacing Clin Electrophysiol       Date:  2000-03       Impact factor: 1.976

10.  Detection of atrial fibrillation and flutter by a dual-chamber implantable cardioverter-defibrillator. For the Worldwide Jewel AF Investigators.

Authors:  C D Swerdlow; W Schsls; B Dijkman; W Jung; N V Sheth; W H Olson; B D Gunderson
Journal:  Circulation       Date:  2000-02-29       Impact factor: 29.690

  10 in total
  2 in total

1.  Inappropriate mode switching clarified by using a chest radiograph.

Authors:  Brian Marino; Abhishek Jaiswal; Seth Goldbarg
Journal:  J Arrhythm       Date:  2015-02-14

2.  Use of an atrial lead with very short tip-to-ring spacing avoids oversensing of far-field R-wave.

Authors:  Christof Kolb; Georg Nölker; Carsten Lennerz; Hansmartin Jetter; Verena Semmler; Klaus Pürner; Klaus-Jürgen Gutleben; Tilko Reents; Klaus Lang; Ulrich Lotze
Journal:  PLoS One       Date:  2012-06-22       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.