Impact and prevention of far-field sensing in fallback mode switches.

Pierre Bordacher1, Serge Cazeau, Laurence Graindorg, Philippe Ritter.   

Abstract

UNLABELLED: Far-field oversensing (FFOS) promoted by high atrial sensitivity and short atrial refractory periods induces false positive mode switches. We evaluated the incidence of ventricular FFOS in a population of DDD paced patients.
METHODS: One hundred thirty-seven patients (71 +/- 10 years, 76 men) implanted with a Talent DR pacemaker were studied. Before discharge, an analysis of internal data stored in the memories of the PM was performed by the specific software incorporated in the programmer in parallel with a 24-hour Holter recording. Data were validated by a panel of experts. One and 4 months follow-up was based only on the data stored in the PM memories.
RESULTS: Pacing indications were atrioventricular block (n = 75), sinus node dysfunction (n = 57), and other (n = 5). Sustained far-field oversensing was observed in 12/137 patients (9%). Out of a total of 3,511 triggered mode switch episodes, FFOS accounted for 20% and 7% of a 311 days cumulative time in mode switch. Inappropriate mode switch episodes induced by far-field were more numerous but shorter than episodes prompted by atrial arrhythmias. Atrial sensitivity was increased in eight patients, successfully in four. Reprogramming of the atrial refractory period (156 +/- 11 ms) was successful in five of six patients.
CONCLUSIONS: A 9% rate of ventricular FFOS was observed in an unselected population, easily and automatically diagnosed using the internal memory function and the automatic analysis provided by the programmer. Prolongation of the atrial refractory period was more effective than resetting of the atrial sensitivity in eliminating FFOS.

Entities:  

Mesh:

Year:  2003        PMID: 12687813     DOI: 10.1046/j.1460-9592.2003.00017.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Preventricular far-field sensing in the atrial channel of dual chamber pacemakers--an occasional cause of inappropriate mode switch.

Authors:  Christof Kolb; Selim Aratma; Bernhard Zrenner; Claus Schmitt
Journal:  J Interv Card Electrophysiol       Date:  2004-06       Impact factor: 1.900

2.  Ambulatory Holter monitoring in asymptomatic patients with DDD pacemakers - do we need ACC/AHA Guidelines revision?

Authors:  Michal Chudzik; Artur Klimczak; Jerzy Krzysztof Wranicz
Journal:  Arch Med Sci       Date:  2013-11-05       Impact factor: 3.318

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.