PURPOSE: To evaluate intraindividual visual acuity, wavefront errors and modulation transfer functions in patients implanted with two diffractive multifocal intraocular lenses.Entities:
Keywords: Aberrometer; Modulation transfer function; Multifocal intraocular lenses; Visual acuity; Wavefront
Mesh:
Year: 2009 PMID: 19841701 PMCID: PMC2763069 DOI: 10.1590/S1807-59322009001000004
Source DB: PubMed Journal: Clinics (Sao Paulo) ISSN: 1807-5932 Impact factor: 2.365
Visual performance of diffractive multifocal IOLs
| Intraocular lens | Distance uncorrected visual acuity | Distance corrected visual acuity | Near uncorrected visual acuity | Near corrected visual acuity | Spherical equivalent | |
|---|---|---|---|---|---|---|
| ReSTOR | 20 | 0.74 ± 0.20 | 1.00 ± 0.00 | 0.96 ± 0.10 | 1.00 ± 0.00 | 0.21 ± 0.17 |
| Tecnis | 20 | 0.76 ± 0.22 | 1.00 ± 0.00 | 0.93 ± 0.14 | 1.00 ± 0.00 | 0.22 ± 0.17 |
| 0.286 | 0.963 | 0.840 |
The visual acuity values were converted to the logarithm of the minimal angle resolution scale (logMAR) for statistical analysis and are thus presented in decimal scale ± standard deviation.
Figure 1Visual acuity at various defocusing levels in the ReSTOR group and Tecnis group. The values are the mean of the logMAR visual acuity. The binocular defocusing curve showed a good range for the pseudo-accommodative amplitude
Wavefront data with 5 mm pupil diameter
| Mean (μm) ± SD | |||
|---|---|---|---|
| ReSTOR | Tecnis | ||
| HOA RMS | 0.46 ± 0.63 μm | 0.34 ± 0.19 μm | 0.575 |
| Coma aberration RMS | 0.17 ± 0.16 μm | 0.16 ± 0.07 μm | 0.455 |
| Spherical aberration RMS | 0.08 ± 0.07 μm | 0.03 ± 0.03 μm | 0.004 |
| Total aberration RMS | 0.75 ± 0.76 μm | 0.56 ± 0.23 μm | 0.737 |
HOA = higher-order aberration; RMS = root mean square; SD = standard deviation;
statistically significant.
Figure 2Bar graph showing the wavefront data with a 5 mm pupil diameter for two diffractive multifocal intraocular lenses. The root-mean-square of the total aberration, the higher-order aberration (HOA) and coma aberration were not statistically different between the groups. The spherical aberration was significantly lower in the Tecnis group than in the ReSTOR group (p=0.004)
Figure 3The modulation transfer function (MTF) curve analyzed for a 5 mm pupil diameter of the IOL groups at different spatial frequencies. The MTF shows the contrast transferred at different spatial frequencies. The aspheric full diffractive multifocal IOL Tecnis ZM900 performed similar to the spherical apodized diffractive multifocal ReSTOR (p>0.05 at all spatial frequencies studied)