Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer.

P Gunvant1, M Baskaran, L Vijaya, I S Joseph, R J Watkins, M Nallapothula, D C Broadway, D J O'Leary.   

Abstract

AIMS: To investigate the effect of central corneal thickness and corneal curvature on intraocular pressure measurements using the pulsatile ocular blood flow tonograph and the Goldmann applanation tonometer, and to assess the agreement between the pulsatile ocular blood flow tonograph and the Goldmann applanation tonometer in intraocular pressure measurement.
METHODS: 479 subjects underwent intraocular pressure measurements with the Goldmann applanation tonometer and the pulsatile ocular blood flow tonograph. Of these, 334 patients underwent additional measurement of central corneal thickness with an ultrasonic pachymeter and corneal curvature measurement with a keratometer.
RESULTS: The intraocular pressure measurements obtained with both the Goldmann applanation tonometer and the pulsatile ocular blood flow tonograph varied with central corneal thickness and mean keratometric reading. Intraocular pressure measured using the Goldmann applanation tonometer increased by 0.027 mm Hg per micro m increase in central corneal thickness. Intraocular pressure measured using the pulsatile ocular blood flow tonograph increased by 0.048 mm Hg per micro m increase in central corneal thickness. For an increase of 1 mm of mean corneal curvature there was rise in intraocular pressure of 1.14 mm Hg measured by the Goldmann applanation tonometer and of 2.6 mm Hg measured by the pulsatile ocular blood flow tonograph. When compared to the Goldmann applanation tonometer, the pulsatile ocular blood flow tonograph underestimated at low intraocular pressure and overestimated at higher intraocular pressure.
CONCLUSION: Central corneal thickness and corneal curvature affected measurements obtained with the pulsatile ocular blood flow tonograph more than they affected measurements obtained with the Goldmann applanation tonometer.

Entities:  

Mesh:

Year:  2004        PMID: 15031169      PMCID: PMC1772116          DOI: 10.1136/bjo.2003.019331

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  16 in total

1.  Intraocular pressure and pulse rate measurements by the OBF tonograph--comparison to reference instruments.

Authors:  Y C Yang; B Illango; A Cook; M Batterbury
Journal:  Ophthalmic Physiol Opt       Date:  2000-09       Impact factor: 3.117

2.  Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen.

Authors:  Archana Bhan; Andrew C Browning; Sunil Shah; Robin Hamilton; Dinesh Dave; Harminder S Dua
Journal:  Invest Ophthalmol Vis Sci       Date:  2002-05       Impact factor: 4.799

3.  Applanation tonometry and central corneal thickness.

Authors:  N Ehlers; T Bramsen; S Sperling
Journal:  Acta Ophthalmol (Copenh)       Date:  1975-03

4.  Tonometer utilization, accuracy, and calibration under field conditions.

Authors:  I F Wessels; Y Oh
Journal:  Arch Ophthalmol       Date:  1990-12

5.  Central corneal thickness, tonometry, and ocular dimensions in glaucoma and ocular hypertension.

Authors:  R P Singh; I Goldberg; S L Graham; A Sharma; M Mohsin
Journal:  J Glaucoma       Date:  2001-06       Impact factor: 2.503

6.  Corneal curvature in applanation tonometry.

Authors:  H H Mark
Journal:  Am J Ophthalmol       Date:  1973-08       Impact factor: 5.258

7.  Elevated tonometer readings caused by a thick cornea.

Authors:  F K Hansen; N Ehlers
Journal:  Acta Ophthalmol (Copenh)       Date:  1971

8.  Falsely elevated intraocular pressure due to increased central corneal thickness.

Authors:  A M Bron; C Creuzot-Garcher; S Goudeau-Boutillon; P d'Athis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1999-03       Impact factor: 3.117

9.  Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic.

Authors:  S Shah; A Chatterjee; M Mathai; S P Kelly; J Kwartz; D Henson; D McLeod
Journal:  Ophthalmology       Date:  1999-11       Impact factor: 12.079

10.  Ocular hypertension and central corneal thickness.

Authors:  W A Argus
Journal:  Ophthalmology       Date:  1995-12       Impact factor: 12.079

View more
  27 in total

1.  Cornea plana associated with open-angle glaucoma: a case report.

Authors:  Bilge Ozturk Sahin; Goktug Seymenoglu; Esin F Baser
Journal:  Int Ophthalmol       Date:  2011-12-11       Impact factor: 2.031

2.  Effects of phacoemulsification surgery on ocular hemodynamics.

Authors:  Adem Turk; Suleyman Mollamehmetoglu; Halil Ibrahim Imamoglu; Mehmet Kola; Hidayet Erdol; Nurettin Akyol
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

Review 3.  [Glaucoma diagnostics and corneal thickness].

Authors:  J von Eicken; M Kohlhaas; H Höh
Journal:  Ophthalmologe       Date:  2005-09       Impact factor: 1.059

4.  The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.

Authors:  P-A Tonnu; T Ho; T Newson; A El Sheikh; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

5.  A comparison of four methods of tonometry: method agreement and interobserver variability.

Authors:  P-A Tonnu; T Ho; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

6.  Pulsatile ocular blood flow: relationship with flow velocities in vessels supplying the retina and choroid.

Authors:  Itay Ben Zion; Alon Harris; Brent Siesky; Shiri Shulman; Lynne McCranor; Hanna J Garzozi
Journal:  Br J Ophthalmol       Date:  2007-07       Impact factor: 4.638

7.  Effects of central corneal thickness, central corneal power, and axial length on intraocular pressure measurement assessed with goldmann applanation tonometry.

Authors:  Fatih Özcura; Sayime Aydın; Nevin Uzgören
Journal:  Jpn J Ophthalmol       Date:  2008-11-11       Impact factor: 2.447

8.  Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Fatih Özcura; Nilgün Yildirim; Afsun Şahin; Ertuğrul Çolak
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

9.  The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry.

Authors:  A Kotecha; E T White; J M Shewry; D F Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-12       Impact factor: 4.638

10.  Clinical evaluation of multiparameter correction equations for Goldmann applanation tonometry.

Authors:  P G Davey; A Elsheikh; D F Garway-Heath
Journal:  Eye (Lond)       Date:  2013-03-15       Impact factor: 3.775

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.