Literature DB >> 16521011

Eleven-year results of the ABG I hip replacement.

W Y Kim1, B N Muddu.   

Abstract

The ABG I uncemented hip prosthesis has demonstrated an unacceptably high failure rate and poor functional outcome in our patients. A prospective review was performed of 38 implanted hips with a mean follow-up of 11 years. Sixteen hips have been revised or are candidates for revision hip surgery, due principally to polyethylene wear and pelvic osteolysis. Contributing factors were likely to be the poor wear characteristics of the polyethylene liners--which were gamma irradiated in air--and the thinness of these liners. Of further concern were the poor pain and function scores in the remaining hips [mean Harris Hip Scores (HHS) of 70 (SD 22) at 11-year follow-up]. The influence of the surgical learning curve on clinical results may have been an important factor. Undersized femoral components were associated with poor pain and functional scores in five patients with HHS < 60. All ABG I hip implants must remain under careful, long-term review.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16521011      PMCID: PMC2532099          DOI: 10.1007/s00264-006-0075-1

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  9 in total

1.  Medium-term results of the ABG total hip arthroplasty in young patients.

Authors:  Konstantinos A Giannikas; Robert Din; Sahzad Sadiq; Timothy H Dunningham
Journal:  J Arthroplasty       Date:  2002-02       Impact factor: 4.757

2.  Premature wear and osteolysis in an HA-coated, uncemented total hip arthroplasty.

Authors:  P Duffy; J L Sher; P F Partington
Journal:  J Bone Joint Surg Br       Date:  2004-01

3.  Measurement of polyethylene wear in cementless total hip arthroplasty.

Authors:  C H Shih; P C Lee; J H Chen; C L Tai; L F Chen; J S Wu; W H Chang
Journal:  J Bone Joint Surg Br       Date:  1997-05

4.  Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation.

Authors:  W H Harris
Journal:  J Bone Joint Surg Am       Date:  1969-06       Impact factor: 5.284

5.  The Otto Aufranc Award. Impact of gamma sterilization on clinical performance of polyethylene in the hip.

Authors:  L C Sutula; J P Collier; K A Saum; B H Currier; J H Currier; W M Sanford; M B Mayor; R E Wooding; D K Sperling; I R Williams
Journal:  Clin Orthop Relat Res       Date:  1995-10       Impact factor: 4.176

6.  Association between hospital and surgeon procedure volume and outcomes of total hip replacement in the United States medicare population.

Authors:  J N Katz; E Losina; J Barrett; C B Phillips; N N Mahomed; R A Lew; E Guadagnoli; W H Harris; R Poss; J A Baron
Journal:  J Bone Joint Surg Am       Date:  2001-11       Impact factor: 5.284

7.  Measurement of polyethylene wear in metal-backed acetabular cups. II. Clinical application.

Authors:  P A Devane; R B Bourne; C H Rorabeck; S MacDonald; E J Robinson
Journal:  Clin Orthop Relat Res       Date:  1995-10       Impact factor: 4.176

8.  High osteolysis and revision rate with the hydroxyapatite-coated ABG hip prostheses: 65 hips in 56 young patients followed for 5-9 years.

Authors:  Jan Blacha
Journal:  Acta Orthop Scand       Date:  2004-06

9.  Polyethylene and metal debris generated by non-articulating surfaces of modular acetabular components.

Authors:  O L Huk; M Bansal; F Betts; C M Rimnac; J R Lieberman; M H Huo; E A Salvati
Journal:  J Bone Joint Surg Br       Date:  1994-07
  9 in total
  7 in total

1.  The influence of surgical approach on outcome in Birmingham hip resurfacing.

Authors:  Callum W McBryde; Matthew P Revell; Andrew M Thomas; Ronan B Treacy; Paul B Pynsent
Journal:  Clin Orthop Relat Res       Date:  2008-01-26       Impact factor: 4.176

2.  Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty.

Authors:  Jiri Gallo; Vitezslav Havranek; Jana Zapletalova
Journal:  Int Orthop       Date:  2009-02-13       Impact factor: 3.075

3.  THA with the ABG I prosthesis at 15 years. Excellent survival with minimal osteolysis.

Authors:  P N Baker; I A McMurtry; G Chuter; A Port; J Anderson
Journal:  Clin Orthop Relat Res       Date:  2010-07       Impact factor: 4.176

4.  Letter to the editor: THA with the ABG I prosthesis at 15 years: excellent survival with minimal osteolysis.

Authors:  Jiri Gallo
Journal:  Clin Orthop Relat Res       Date:  2010-04       Impact factor: 4.176

5.  The Bicontact cementless primary total hip arthroplasty: long-term results.

Authors:  Girish Swamy; Alistair Pace; Conal Quah; Peter Howard
Journal:  Int Orthop       Date:  2010-09-10       Impact factor: 3.075

Review 6.  Recent advances in glycerol polymers: chemistry and biomedical applications.

Authors:  Heng Zhang; Mark W Grinstaff
Journal:  Macromol Rapid Commun       Date:  2014-10-13       Impact factor: 5.734

7.  Optimisation of Glycerol and Itaconic Anhydride Polycondensation.

Authors:  Krzysztof Kolankowski; Magdalena Miętus; Paweł Ruśkowski; Agnieszka Gadomska-Gajadhur
Journal:  Molecules       Date:  2022-07-20       Impact factor: 4.927

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.