Pannus formation in aortic valve prostheses in the late postoperative period.

Yukio Kuniyoshi1, Kageharu Koja, Kazufumi Miyagi, Mituyoshi Shimoji, Tooru Uezu, Katuya Arakaki, Satoshi Yamashiro, Katuhito Mabuni, Shigenobu Senaha, Yoshiyuki Nakasone.   

Abstract

We present three patients who underwent repeat aortic valve replacement for prosthetic valve dysfunction caused by tissue ingrowth in the late postoperative period. These patients (three women aged 48-51 years, mean 49.3 +/- 1.53 years) underwent operations for restriction of prosthetic valve leaflet movement by pannus in the left ventricular outflow tract. The interval from the previous operation ranged from 8.0 to 9.6 years (mean 9.6 +/- 2.0 years). The symptoms of the patients were New York Heart Association functional class I, II, and IV in one patient each. Diagnosis was made by cinefluoroscopy in two patients and aortography in one patient. The operative procedures consisted of aortic valve replacement ( n = 1) and aortic valve replacement with mitral valve replacement ( n = 2). Pannus was found at the left ventricular aspect of the prosthetic valve in all patients. In two patients, the pannus directly restricted movement of the leaflet and also severely narrowed the inflow orifice of the prosthetic valve. In the other patient, the pannus had grown at a distance of 7 mm from the valve and narrowed the left ventricular outflow tract circularly. The postoperative course was uneventful and all three patients were discharged in a good condition. One patient died of pneumonia 8 months after surgery and the other two patients have remained well and have been followed up for one and a half years. In conclusion, there may be a discrepancy between the clinical symptoms and the grade of subvalvular stenosis caused by pannus. Therefore, it is essential for satisfactory operative results that early diagnosis be made by various means.

Entities:  

Mesh:

Year:  2003        PMID: 14598101     DOI: 10.1007/s10047-003-0226-8

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  10 in total

1.  Utility of cardiac computed tomography for evaluation of pannus in mechanical aortic valve.

Authors:  Young Joo Suh; Young Jin Kim; Sak Lee; Yoo Jin Hong; Hye-Jeong Lee; Jin Hur; Byoung Wook Choi; Byung-Chul Chang
Journal:  Int J Cardiovasc Imaging       Date:  2015-05-21       Impact factor: 2.357

2.  Prevalence of pannus formation after aortic valve replacement: clinical aspects and surgical management.

Authors:  Yoshimasa Sakamoto; Kazuhiro Hashimoto; Hiroshi Okuyama; Shinichi Ishii; Taguchi Shingo; Hiroshi Kagawa
Journal:  J Artif Organs       Date:  2006       Impact factor: 1.731

3.  Recurrent pannus formation causing prosthetic aortic valve dysfunction: is excision without valve re-replacement applicable?

Authors:  Ahmad K Darwazah
Journal:  J Cardiothorac Surg       Date:  2012-06-29       Impact factor: 1.637

4.  Redo surgery risk in patients with cardiac prosthetic valve dysfunction.

Authors:  Marek Maciejewski; Katarzyna Piestrzeniewicz; Agata Bielecka-Dąbrowa; Monika Piechowiak; Ryszard Jaszewski
Journal:  Arch Med Sci       Date:  2011-05-17       Impact factor: 3.318

5.  Measurement of Opening and Closing Angles of Aortic Valve Prostheses In Vivo Using Dual-Source Computed Tomography: Comparison with Those of Manufacturers' in 10 Different Types.

Authors:  Young Joo Suh; Young Jin Kim; Yoo Jin Hong; Hye-Jeong Lee; Jin Hur; Dong Jin Im; Yun Jung Kim; Byoung Wook Choi
Journal:  Korean J Radiol       Date:  2015-08-21       Impact factor: 3.500

6.  Investigation of failure modes of explanted porcine valves in the mitral position.

Authors:  Kun Liu; Wentao Feng; Xianda Yang; Jinglun Shen; Haibo Zhang; Yubo Fan
Journal:  J Thorac Dis       Date:  2021-05       Impact factor: 2.895

7.  Can postoperative mean transprosthetic pressure gradient predict survival after aortic valve replacement?

Authors:  Bart M Koene; Mohamed A Soliman Hamad; Wobbe Bouma; Massimo A Mariani; Kathinka C Peels; Jan-Melle van Dantzig; Albert H van Straten
Journal:  Clin Res Cardiol       Date:  2013-10-18       Impact factor: 5.460

8.  Artificial aortic valve dysfunction due to pannus and thrombus - different methods of cardiac surgical management.

Authors:  Stanisław Ostrowski; Anna Marcinkiewicz; Anna Kośmider; Andrzej Walczak; Radosław Zwoliński; Ryszard Jaszewski
Journal:  Kardiochir Torakochirurgia Pol       Date:  2015-09-28

9.  Pannus-related prosthetic valve dysfunction. Case report.

Authors:  Maria-Sînziana Moldovan; Daniela Bedeleanu; Emese Kovacs; Lorena Ciumărnean; Adrian Molnar
Journal:  Clujul Med       Date:  2016-01-15

10.  Effects of pannus formation on the flow around a bileaflet mechanical heart valve.

Authors:  Woojin Kim; Haecheon Choi; Jihoon Kweon; Dong Hyun Yang; Young-Hak Kim
Journal:  PLoS One       Date:  2020-06-12       Impact factor: 3.240

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.