OBJECTIVE: The aim of this study was to validate the quantification of absolute renal perfusion (RP) determined by dynamic magnetic resonance imaging (MRI) and contrast media using an experimental model in the rabbit and a transit-timed ultrasound flow probe around the left renal artery as comparison.
MATERIAL AND METHODS: An MR-compatible ultrasonic time-of-flight flow-probe was placed around the left renal artery in 9 New Zealand white rabbits. Absolute RP in basal state, after mechanical renal artery stenosis, intravenous dopamine, angiotensin II, or colloid infusion was measured using dynamic MRI and intravenous injection of gadoteridol. The results were correlated to the renal artery flow measured inside the magnet with the transit-timed flow-probe. For the signal intensity concentration conversion, we applied different calibrations according to various velocities measured in the aorta by a phase contrast sequence to correct for inflow effect. MRI-derived RP (in mL/min) was calculated by the maximum upslope method, where RP/volume was defined as the ratio of the cortex contrast enhancement slope over the maximum of the arterial input function determined in the aorta.
RESULTS: Reproducible arterial and renal transit curve with excellent contrast to noise ratio were obtained. The MRI derived perfusion was systematically underestimated by comparison to the ultrasonic transit-timed flow-probe but was linearly correlated with these measures (r = 0.80, P < 0.001).
CONCLUSIONS: Using a flow-sensitive calibration, an accurate arterial input function can be measured from the blood MR signal and used in a realistic model to assess the RP. There was a good correlation between the MR-derived RP and the renal artery blood flow measured by the flow-meter. This experimental study validates absolute RP quantification by MRI and contrast media injection and justifies further clinical studies.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12960528     DOI: 10.1097/01.RLI.0000077127.11949.8c

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  14 in total

1.  Evaluation of flow measurement from the first pass bolus T1 weighted images using inversion recovery sequence.

Authors:  M Nazarpoor
Journal:  Br J Radiol       Date:  2010-10-19       Impact factor: 3.039

Review 2.  MR urography in children.

Authors:  J Damien Grattan-Smith; Richard A Jones
Journal:  Pediatr Radiol       Date:  2006-06-22

Review 3.  Analysis of contrast-enhanced MR images to assess renal function.

Authors:  N Michoux; J-P Vallée; A Pechère-Bertschi; X Montet; L Buehler; B E Van Beers
Journal:  MAGMA       Date:  2006-08-12       Impact factor: 2.310

Review 4.  Renal relevant radiology: renal functional magnetic resonance imaging.

Authors:  Behzad Ebrahimi; Stephen C Textor; Lilach O Lerman
Journal:  Clin J Am Soc Nephrol       Date:  2013-12-26       Impact factor: 8.237

5.  Placental perfusion in uterine ischemia model as evaluated by dynamic contrast enhanced MRI.

Authors:  Alexander Drobyshevsky; P V Prasad
Journal:  J Magn Reson Imaging       Date:  2015-04-09       Impact factor: 4.813

6.  Effect of concentration of contrast agent on the inflow effect for measuring absolute perfusion by use of inversion recovery T₁-weighted TurboFLASH images.

Authors:  Mahmood Nazarpoor
Journal:  Radiol Phys Technol       Date:  2011-11-24

7.  Non-Invasive Blood Perfusion Measurements Using a Combined Temperature and Heat Flux Surface Probe.

Authors:  Patricia L Ricketts; Ashvinikumar V Mudaliar; Brent E Ellis; Clay A Pullins; Leah A Meyers; Otto I Lanz; Elaine P Scott; Thomas E Diller
Journal:  Int J Heat Mass Transf       Date:  2008-11-01       Impact factor: 5.584

8.  Noninvasive blood perfusion measurements of an isolated rat liver and an anesthetized rat kidney.

Authors:  Ashvinikumar V Mudaliar; Brent E Ellis; Patricia L Ricketts; Otto I Lanz; Charles Y Lee; Thomas E Diller; Elaine P Scott
Journal:  J Biomech Eng       Date:  2008-12       Impact factor: 2.097

9.  A phantom tissue system for the calibration of perfusion measurements.

Authors:  Ashvinikumar V Mudaliar; Brent E Ellis; Patricia L Ricketts; Otto I Lanz; Elaine P Scott; Thomas E Diller
Journal:  J Biomech Eng       Date:  2008-10       Impact factor: 2.097

10.  Quantitative MR measures of intrarenal perfusion in the assessment of transplanted kidneys: initial experience.

Authors:  Andrew L Wentland; Elizabeth A Sadowski; Arjang Djamali; Thomas M Grist; Bryan N Becker; Sean B Fain
Journal:  Acad Radiol       Date:  2009-06-17       Impact factor: 3.173

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.