Abstract

OBJECTIVE: To assess the accuracy and precision of a software-aided system to measure migration of femoral components after total hip replacement (THR) on digitised radiographs. DESIGN AND PATIENTS: Subsidence and varus-valgus tilt of THR stems within the femur were measured on digitised anteroposterior pelvic radiographs. The measuring software (UMA, GEMED, Germany) relies on bony landmarks and comparability parameters of two consecutive radiographs. Its accuracy and precision were calculated by comparing it with the gold standard in migration measurements, radiostereometric analysis (RSA). Radiographs and corresponding RSA measurements were performed in 60 patients (38-69 years) following cementless THR surgery. RESULTS AND
CONCLUSIONS: The UMA software measured the subsidence of the stems with an accuracy of +/-2.5 mm and varus-valgus tilt with an accuracy of +/-1.8 degrees (95% confidence interval). A good interobserver and intraobserver reliability was calculated with Cronbach's alpha ranging from 0.86 to 0.97. Measuring the subsidence of THR stems within the femur is an important parameter in the diagnosis of implant loosening. Software systems such as UMA improve the accuracy of migration measurements and are easy to use on routinely performed radiographs of operated hip joints.

Entities:  

Mesh:

Year:  2003        PMID: 12898059     DOI: 10.1007/s00256-003-0670-9

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  8 in total

1.  Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement. Einzel-Bild-Röntgen-Analyse-femoral component analysis.

Authors:  R Biedermann; M Krismer; B Stöckl; P Mayrhofer; E Ornstein; H Franzén
Journal:  J Bone Joint Surg Br       Date:  1999-03

2.  Evaluation of accuracy and precision of bone markers for the measurement of migration of hip prostheses. A comparison of conventional measurements.

Authors:  R Biedermann; B Stöckl; M Krismer; P Mayrhofer; E Ornstein; H Franzén
Journal:  J Bone Joint Surg Br       Date:  2001-07

3.  Assessing the results of hip replacement. A comparison of five different rating systems.

Authors:  J J Callaghan; S H Dysart; C F Savory; W J Hopkinson
Journal:  J Bone Joint Surg Br       Date:  1990-11

4.  Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses.

Authors:  J Kärrholm; B Borssén; G Löwenhielm; F Snorrason
Journal:  J Bone Joint Surg Br       Date:  1994-11

5.  Radiostereometry of hip prostheses. Review of methodology and clinical results.

Authors:  J Kärrholm; P Herberts; P Hultmark; H Malchau; B Nivbrant; J Thanner
Journal:  Clin Orthop Relat Res       Date:  1997-11       Impact factor: 4.176

6.  Prediction of clinical outcome of THR from migration measurements on standard radiographs. A study of cemented Charnley and Stanmore femoral stems.

Authors:  P S Walker; S F Mai; A G Cobb; G Bentley; J Hua
Journal:  J Bone Joint Surg Br       Date:  1995-09

Review 7.  The problem is osteolysis.

Authors:  W H Harris
Journal:  Clin Orthop Relat Res       Date:  1995-02       Impact factor: 4.176

8.  Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients.

Authors:  H Malchau; J Kärrholm; Y X Wang; P Herberts
Journal:  Acta Orthop Scand       Date:  1995-10
  8 in total
  3 in total

1.  Component version in modular total hip revision.

Authors:  Michael A Kopec; Aaron Pemberton; Joseph C Milbrandt; Gordon Allan
Journal:  Iowa Orthop J       Date:  2009

2.  Discrepancy of Trial Rasp and Femoral Stem Relative Position Within the Femoral Canal of a Coated Tapered System: An Intraoperative, Intrapatient Controlled Study.

Authors:  Thomas Hofstaedter; Michael Najfeld; Gion Fessel; Luca C Orlandini; Robert Hube
Journal:  Arthroplast Today       Date:  2020-09-25

Review 3.  Adherence of hip and knee arthroplasty studies to RSA standardization guidelines. A systematic review.

Authors:  Rami Madanat; Tatu J Mäkinen; Hannu T Aro; Charles Bragdon; Henrik Malchau
Journal:  Acta Orthop       Date:  2014-06-23       Impact factor: 3.717

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.