Rami Madanat1, Tatu J Mäkinen, Hannu T Aro, Charles Bragdon, Henrik Malchau.
Abstract
BACKGROUND ANDEntities:
Mesh:
Year: 2014 PMID: 24954489 PMCID: PMC4164860 DOI: 10.3109/17453674.2014.934187
Source DB: PubMed Journal: Acta Orthop ISSN: 1745-3674 Impact factor: 3.717
Figure 1.PRISMA flow chart of study selection.
List of the 13 guidelines and subcategories used to evaluate study adherence (adapted from Valstar et al. 2005)
| Guideline number Subcategory | Guideline | |
|---|---|---|
| 1 | A | Translation units in millimeters |
| B | Rotation units in degrees | |
| 2 | A | Accuracy presented |
| B | Precision presented | |
| C | Measurement interval mentioned | |
| D | Window of tolerance mentioned | |
| 3 | Type of calibration cage mentioned | |
| 4 | Fixed or portable X-ray sources used | |
| 5 | A | Subject positioning described |
| B | Orientation of global coordinate system presented | |
| 6 | A | Method of image acquisition stated |
| B | Scanner or system details mentioned | |
| 7 | A | Software name mentioned |
| B | Software version mentioned | |
| 8 | A | Marker size mentioned |
| B | Marker size validated (or manufacturer mentioned) | |
| 9 | RSA method mentioned (marker, model, etc.) | |
| 10 | A | Cutoff level for condition number noted |
| B | Cutoff level for mean error of rigid-body fitting noted | |
| 11 | A | Rigid-body fixed coordinate frames defined |
| B | Angular rotation sequence mentioned if relevant | |
| 12 | A | Precision determined using double examinations |
| B | Double examinations of all patients performed | |
| 13 | A | All 6 degrees of freedom reported |
| B | Point or segment motion used | |
| C | Point(s) used to measure translation indicated | |
Figure 2.The annual numbers of hip and knee arthroplasty RSA studies published from 2000 to 2011 that met the inclusion criteria (bars) and the proportion of studies with high adherence (line). The RSA guidelines where published at the end of 2005.
Figure 3.Adherence of studies published after the guidelines to each of the 13 individual guideline items. The levels of adherence are presented in green (full adherence), yellow (partial adherence), and red (no adherence). Guidelines are presented left to right from highest to lowest adherence. The guideline number is given in parentheses.
Adherence of studies published during 2006–2011 to the RSA guidelines according to region of publication
| Total no. of studies | Number (%) of studies with high adherence | |
|---|---|---|
| Nordic countries | 55 | 29 (53) |
| Rest of Europe | 21 | 8 (38) |
| USA/Canada | 9 | 3 (33) |
| Australia | 7 | 3 (43) |
Partial or full adherence to at least 10 of 13 RSA guidelines.
Denmark, Finland, Norway, and Sweden.
Germany, Italy, the Netherlands, UK.
Adherence of studies published during 2006–2011 to the RSA guidelines according to journal
| Total no. of studies | No. of studies with high adherence | |
|---|---|---|
| Acta Orthopaedica | 23 | 9 |
| Clinical Orthopaedics and | ||
| Related Research | 18 | 12 |
| Journal of Arthroplasty | 13 | 4 |
| Journal of Bone and Joint | ||
| Surgery (British Volume) | 12 | 3 |
| International Orthopaedics | 6 | 2 |
| The Knee | 4 | 2 |
| Other | 16 | 11 |
Partial or full adherence to at least 10 of 13 RSA guidelines.
11 journals with 2 publications or less that met the inclusion criteria during the study period.
Suggested presentation of standardized output for a clinical RSA study
|
|
| Time period of patient recruitment |
| Positioning of subject and calibration cage |
| Orientation of the global coordinate system (consider a diagram) |
| Rigid body fixed coordinate frames defined |
| RSA method (marker-based, geometrical or model-based) and version of RSA software |
| Units used for translation millimeters and the units for rotations degrees |
| Measurement interval |
| The points used to measure translation indicated (consider an x-ray or diagram) |
| Translations and angular rotations with all six degrees of freedom if applicable |
| Clinical precision |
| Number of patients for whom RSA was not successful |
|
|
| Accuracy of the arrangement (if phantom studies have been performed) |
| Size and manufacturer of RSA markers |
| Type and manufacturer of calibration cage |
| Fixed or portable x-ray sources and manufacturer |
| Radiographic exposure settings (kV and mAs) |
| Window of tolerance for both postoperative and follow-up RSA examinations (days/weeks) |
| Method of image acquisition and scanner or system details as well as image resolution |
| Number of patients for whom double examinations were performed |
| Cut-off level for CN and ME |
| Formula used to calculate confidence interval for precision |
| Angular rotation sequence defined (if relevant) |