Erik G Larsson1, Deniz Erdogmus, Rui Yan, Jose C Principe, Jeffrey R Fitzsimmons.
Abstract
We consider the commonly used "Sum-of-Squares" (SoS) reconstruction method for phased-array magnetic resonance imaging with unknown coil sensitivities. We show that the signal-to-noise ratio (SNR) in the image produced by SoS is asymptotically (as the input SNR--> infinity ) equal to that of maximum-ratio combining, which is the best unbiased reconstruction method when the coil sensitivities are known. Finally, we discuss the implications of this result.Entities:
Mesh:
Year: 2003 PMID: 12852915 DOI: 10.1016/s1090-7807(03)00132-0