| Literature DB >> 12841788 |
David S Followill1, Marilyn S Stovall, Stephen F Kry, Geoffrey S Ibbott.
Abstract
The shielding calculations for high energy (>10 MV) linear accelerators must include the photoneutron production within the head of the accelerator. Procedures have been described to calculate the treatment room door shielding based on the neutron source strength (Q value) for a specific accelerator and energy combination. Unfortunately, there is currently little data in the literature stating the neutron source strengths for the most widely used linear accelerators. In this study, the neutron fluence for 36 linear accelerators, including models from Varian, Siemens, Elekta/Philips, and General Electric, was measured using gold-foil activation. Several of the models and energy combinations had multiple measurements. The neutron fluence measured in the patient plane was independent of the surface area of the room, suggesting that neutron fluence is more dependent on the direct neutron fluence from the head of the accelerator than from room scatter. Neutron source strength, Q, was determined from the measured neutron fluences. As expected, Q increased with increasing photon energy. The Q values ranged from 0.02 for a 10 MV beam to 1.44(x10(12)) neutrons per photon Gy for a 25 MV beam. The most comprehensive set of neutron source strength values, Q, for the current accelerators in clinical use are presented for use in calculating room shielding. (c) 2003 American College of Medical Physics.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12841788 PMCID: PMC5724447 DOI: 10.1120/jacmp.v4i3.2514
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Location of the gold foils and moderators (points A‐E) for the neutron fluence measurements around the linear accelerators.
Neutron source strength (Q) values for various linac model and energy combinations.
| Manufacturer | Model | Nominal MV |
| Std. Deviation ( | No. of Linac s | Published |
|---|---|---|---|---|---|---|
| Varian | 1800 | 10 | 0.06 | |||
| Varian | 1800 | 15 | 0.76 | |||
| Varian | 1800 | 18 | 1.22 | |||
| Varian | 2100C | 18 | 0.96 | 0.11 | 17 | |
| Varian | 2100C | 18 | 0.87 | 1 | ||
| Varian | 2300CD | 18 | 0.95 | 0.03 | 2 | |
| Varian | 2500 | 24 | 0.77 | 1 | ||
| Siemens | MD2 | 10 | 0.08 | 1 | ||
| Siemens | MD | 15 | 0.20 | 0.02 | 2 | |
| Siemens | KD | 18 | 0.88 | 0.10 | 2 | |
| Siemens | KD | 20 | 0.92 | |||
| Siemens | Primus | 10 | 0.02 | 1 | ||
| Siemens | Primus | 15 | 0.12 | 1 | ||
| Siemens | Primus | 15 | 0.21 | 1 | ||
| Siemens | Primus | 15 | 1 | 0.20 | ||
| Elekta | SL‐20 | 17 | 0.69 | |||
| Elekta | SL‐20 | 18 | 0.46 | 1 | ||
| Elekta | SL‐25 | 18 | 0.46 | 1 | ||
| Elekta | SL‐25 | 22 | 2.37 | |||
| Elekta | SL‐25 | 25 | 1.44 | 0.31 | 3 | |
| GE | Saturne 41 | 12 | 0.24 | |||
| GE | Saturne 41 | 15 | 0.47 | |||
| GE | Saturne 43 | 18 | 1.32 | 1 | 1.50 | |
| GE | Saturne 43 | 25 | 2.40 |
With MiMIC device in place.
With the multileaf collimator set to a field.
Figure 2The measured neutron fluence ( perphoton Gy) produced by the 18 MV x‐ray beam from 15 Clinac 2100C linear accelerators at six locations in the patient plane as a function of room surface area.