Room scattered neutrons.

R C McCall1, P H McGinley, K E Huffman.   

Abstract

A method of evaluating the photoneutron fluence in the maze of accelerator facilities due to room scattered neutrons is presented. Measurements were made to demonstrate that the room scattered fluence is reduced by a factor of 2pi in going from the treatment room to the inner maze entrance.

Mesh:

Year:  1999        PMID: 10076975     DOI: 10.1118/1.598505

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071