| Literature DB >> 12815689 |
Guillaume Duhamel1, Cédric de Bazelaire, David C Alsop.
Abstract
Velocity-selective (VS) sequences potentially permit arterial spin labeling (ASL) perfusion imaging with labeling applied very close to the tissue. In this study the effects of cerebrospinal fluid (CSF) motion, radiofrequency (RF) field imperfections, and sequence timing parameters on the appearance and quantitative perfusion values obtained with VS-ASL were evaluated. Large artifacts related to CSF motion were observed with moderate velocity weighting, which were removed by inversion recovery preparation at the cost of increased imaging time. Imperfect refocusing and excitation pulses resulting from nonuniform RF fields produced systematic errors in the ASL subtraction images. A phase cycling scheme was introduced to eliminate these errors. Quantitative perfusion images were obtained with CSF suppression and phase cycling. Gray matter blood flow of 27.7 ml 100 g(-1) min(-1), approximately half the value reported in studies using spatially-selective ASL, was measured. Potential sources for this underestimation are discussed. Copyright 2003 Wiley-Liss, Inc.Mesh:
Substances:
Year: 2003 PMID: 12815689 DOI: 10.1002/mrm.10510
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668