Literature DB >> 12720274

Pair of resonant fiducial markers for localization of endovascular catheters at all catheter orientations.

Titus Kuehne1, Rebecca Fahrig, Kim Butts.   

Abstract

PURPOSE: To test wireless resonance circuits (RC) to be used as fiducial marker of endovascular catheters during MR-guided interventions. Current markers loose their signal enhancement for certain catheter orientations. The purpose of this study was to test a marker setup which overcomes this orientation problem.
MATERIALS AND METHODS: The markers were constructed from a pair of two RCs. The RCs were individually tuned and the coil axes were oriented perpendicular to each other in order to decouple the two RCs. The markers were mounted on the tip of endovascular catheters and tested in vitro and in one porcine in vivo experiment.
RESULTS: An intense MR signal at similar signal levels was noted at all catheter orientations. In the in vivo experiment the markers allowed for fast and reliable MR guidance of the catheters.
CONCLUSION: A pair of two individually tuned and decoupled RCs is well suited for MR guidance of endovascular catheters. Copyright 2003 Wiley-Liss, Inc.

Mesh:

Year:  2003        PMID: 12720274     DOI: 10.1002/jmri.10307

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  10 in total

Review 1.  Cardiovascular interventional magnetic resonance imaging.

Authors:  Robert J Lederman
Journal:  Circulation       Date:  2005-11-08       Impact factor: 29.690

2.  Magnetic resonance imaging guided catheterisation for assessment of pulmonary vascular resistance: in vivo validation and clinical application in patients with pulmonary hypertension.

Authors:  T Kuehne; S Yilmaz; I Schulze-Neick; E Wellnhofer; P Ewert; E Nagel; P Lange
Journal:  Heart       Date:  2005-08       Impact factor: 5.994

3.  Using a low-amplitude RF pulse at echo time (LARFET) for device localization in MRI.

Authors:  Murat Tümer; Baykal Sarioglu; Senol Mutlu; Yekta Ulgen; Arda Yalcinkaya; Cengizhan Ozturk
Journal:  Med Biol Eng Comput       Date:  2014-08-31       Impact factor: 2.602

4.  Endovascular MR-guided Renal Embolization by Using a Magnetically Assisted Remote-controlled Catheter System.

Authors:  Prasheel V Lillaney; Jeffrey K Yang; Aaron D Losey; Alastair J Martin; Daniel L Cooke; Bradford R H Thorne; David C Barry; Andrew Chu; Carol Stillson; Loi Do; Ronald L Arenson; Maythem Saeed; Mark W Wilson; Steven W Hetts
Journal:  Radiology       Date:  2016-03-28       Impact factor: 11.105

5.  Interventional magnetic resonance imaging guided carotid embolectomy using a novel resonant marker catheter: demonstration of preclinical feasibility.

Authors:  Jeffrey K Yang; Andre M Cote; Caroline D Jordan; Sravani Kondapavulur; Aaron D Losey; David McCoy; Andrew Chu; Jay F Yu; Teri Moore; Carol Stillson; Fabio Settecase; Matthew D Alexander; Andrew Nicholson; Daniel L Cooke; Maythem Saeed; Dave Barry; Alastair J Martin; Mark W Wilson; Steven W Hetts
Journal:  Biomed Microdevices       Date:  2017-09-25       Impact factor: 2.838

Review 6.  Magnetic resonance imaging-guided vascular interventions.

Authors:  Cengizhan Ozturk; Michael Guttman; Elliot R McVeigh; Robert J Lederman
Journal:  Top Magn Reson Imaging       Date:  2005-10

7.  Wireless Resonant Circuits Printed Using Aerosol Jet Deposition for MRI Catheter Tracking.

Authors:  Caroline D Jordan; Bradford R H Thorne; Arjun Wadhwa; Aaron D Losey; Eugene Ozhinsky; Sravani Kondapavulur; Vincent Fratello; Teri Moore; Carol Stillson; Colin Yee; Ronald D Watkins; Greig C Scott; Alastair J Martin; Xiaoliang Zhang; Mark W Wilson; Steven W Hetts
Journal:  IEEE Trans Biomed Eng       Date:  2019-06-21       Impact factor: 4.538

Review 8.  Interventional cardiovascular magnetic resonance imaging: a new opportunity for image-guided interventions.

Authors:  Christina E Saikus; Robert J Lederman
Journal:  JACC Cardiovasc Imaging       Date:  2009-11

9.  Whole shaft visibility and mechanical performance for active MR catheters using copper-nitinol braided polymer tubes.

Authors:  Ozgur Kocaturk; Christina E Saikus; Michael A Guttman; Anthony Z Faranesh; Kanishka Ratnayaka; Cengizhan Ozturk; Elliot R McVeigh; Robert J Lederman
Journal:  J Cardiovasc Magn Reson       Date:  2009-08-12       Impact factor: 5.364

Review 10.  Interventional cardiovascular magnetic resonance: still tantalizing.

Authors:  Kanishka Ratnayaka; Anthony Z Faranesh; Michael A Guttman; Ozgur Kocaturk; Christina E Saikus; Robert J Lederman
Journal:  J Cardiovasc Magn Reson       Date:  2008-12-29       Impact factor: 5.364

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.