Abstract

We measured tibiofemoral contact stresses and the load-bearing contact area of fixed and mobile-bearing inlay knee prostheses under dynamic loading conditions. An electronic resistive pressure-measuring sensor was used to detect contact stresses and contact area in five cadaver knees. Stresses were measured with the tibial component aligned normally, as well as in internally- and externally-rotated positions. The average peak contact stresses measured on the fixed inlay were greater (medial 21 MPa and lateral 21 MPa) than those on the mobile inlay (medial/lateral 7.7/5.3 MPa, p = 0.04). Although the average peak contact stresses of the fixed standard inlay greatly exceeded the contact stresses of the other two inlay designs in each malrotated position tested, no statistically significant differences were seen. The data suggest that the ability of the inlay to translate on the tibial baseplate permits the inlay to align itself on the femoral component so that the contact surface area is maximized and contact stresses are reduced.

Mesh:

Year:  2002        PMID: 12553510     DOI: 10.1080/000164702321039598

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  28 in total

1.  In vitro investigation of the influence of tibial slope on quadriceps extension force after total knee arthroplasty.

Authors:  Sven Ostermeier; Christof Hurschler; Henning Windhagen; Christina Stukenborg-Colsman
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-04-21       Impact factor: 4.342

2.  Femoro-tibial kinematics after TKA in fixed- and mobile-bearing knees in the sagittal plane.

Authors:  Kiriakos Daniilidis; Steffen Höll; Georg Gosheger; Ralf Dieckmann; Nicolo Martinelli; Sven Ostermeier; Carsten O Tibesku
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-04-04       Impact factor: 4.342

3.  Mobile-bearing prosthesis did not improve mid-term clinical results of total knee arthroplasty.

Authors:  Shuichi Matsuda; Hideki Mizu-uchi; Shingo Fukagawa; Hiromasa Miura; Ken Okazaki; Hideo Matsuda; Yukihide Iwamoto
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-04-21       Impact factor: 4.342

4.  Infrapatellar fat pad pressure and volume changes of the anterior compartment during knee motion: possible clinical consequences to the anterior knee pain syndrome.

Authors:  Michael Bohnsack; Christof Hurschler; Taskin Demirtas; Oliver Rühmann; Christina Stukenborg-Colsman; Carl-Joachim Wirth
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2004-10-29       Impact factor: 4.342

5.  The influence of pre-tensioning of meniscal transplants on the tibiofemoral contact area.

Authors:  G von Lewinski; C Hurschler; C Allmann; C J Wirth
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2005-11-22       Impact factor: 4.342

6.  Mobile versus fixed-bearing total knee arthroplasty: mid-term comparative clinical results of 216 prostheses.

Authors:  D Biau; M M Mullins; T Judet; P Piriou
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-03-25       Impact factor: 4.342

7.  Effects of initial graft tension on the tibiofemoral compressive forces and joint position after anterior cruciate ligament reconstruction.

Authors:  Mark F Brady; Michael P Bradley; Braden C Fleming; Paul D Fadale; Michael J Hulstyn; Rahul Banerjee
Journal:  Am J Sports Med       Date:  2007-01-11       Impact factor: 6.202

8.  Dynamic measurement of patellofemoral kinematics and contact pressure after lateral retinacular release: an in vitro study.

Authors:  Sven Ostermeier; Marc Holst; Christof Hurschler; Henning Windhagen; Christina Stukenborg-Colsman
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-01-16       Impact factor: 4.342

9.  In vitro measurement of patellar kinematics following reconstruction of the medial patellofemoral ligament.

Authors:  Sven Ostermeier; Marc Holst; Michael Bohnsack; Christof Hurschler; Christina Stukenborg-Colsman; Carl-Joachim Wirth
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-10-10       Impact factor: 4.342

10.  Confirmation of long-term in vivo bearing mobility in eight rotating-platform TKAs.

Authors:  Michael T LaCour; Adrija Sharma; Christopher B Carr; Richard D Komistek; Douglas A Dennis
Journal:  Clin Orthop Relat Res       Date:  2014-04-26       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.