MR imaging-related heating of deep brain stimulation electrodes: in vitro study.

Daniel A Finelli1, Ali R Rezai, Paul M Ruggieri, Jean A Tkach, John A Nyenhuis, Greg Hrdlicka, Ashwini Sharan, Jorge Gonzalez-Martinez, Paul H Stypulkowski, Frank G Shellock.   

Abstract

BACKGROUND AND
PURPOSE: Recent work has shown a potential for excessive heating of deep brain stimulation electrodes during MR imaging. This in vitro study investigates the relationship between electrode heating and the specific absorption rate (SAR) of several MR images.
METHODS: In vitro testing was performed by using a 1.5-T MR imaging system and a head transmit-receive coil, with bilateral deep brain stimulation systems positioned in a gel saline-filled phantom, and temperature monitoring with a fluoroptic thermometry system. Standardized fast spin-echo sequences were performed over a range of high, medium, and low SAR values. Several additional, clinically important MR imaging techniques, including 3D magnetization prepared rapid acquisition gradient-echo imaging, echo-planar imaging, quantitative magnetization transfer imaging, and magnetization transfer-suppressed MR angiography, were also tested by using typical parameters.
RESULTS: A significant, highly linear relationship between SAR and electrode heating was found, with the temperature elevation being approximately 0.9 times the local SAR value. Minor temperature elevations, <1 degrees C, were found with the fast spin-echo, magnetization prepared rapid acquisition gradient-echo, and echo-planar clinical imaging sequences. The high dB/dt echo-planar imaging sequence had no significant heating independent of SAR considerations. Sequences with magnetization transfer pulses produced temperature elevations in the 1.0 to 2.0 degrees C range, which was less than theoretically predicted for the relatively high SAR values.
CONCLUSION: A potential exists for excessive MR imaging-related heating in patients with deep brain stimulation electrodes; however, the temperature increases are linearly related to SAR values. Clinical imaging sequences that are associated with tolerable temperature elevations in the <or=2.0 degrees C range at the electrode tips can be performed safely within an SAR range <2.4 W/kg local (0.9 W/kg whole body averaged).

Entities:  

Mesh:

Year:  2002        PMID: 12427641      PMCID: PMC8185821     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  27 in total

1.  A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation.

Authors:  M K Jain; P D Wolf
Journal:  Ann Biomed Eng       Date:  2000-09       Impact factor: 3.934

2.  MR imaging and electronically activated devices.

Authors:  F G Shellock
Journal:  Radiology       Date:  2001-04       Impact factor: 11.105

3.  Optimizing accuracy in magnetic resonance imaging-guided stereotaxis: a technique with validation based on the anterior commissure-posterior commissure line.

Authors:  C G diPierro; P C Francel; T R Jackson; T Kamiryo; E R Laws
Journal:  J Neurosurg       Date:  1999-01       Impact factor: 5.115

Review 4.  Magnetization transfer in neuroimaging.

Authors:  D A Finelli
Journal:  Magn Reson Imaging Clin N Am       Date:  1998-02       Impact factor: 2.266

5.  Magnetic resonance imaging and spinal cord stimulation systems.

Authors:  L A Liem; V C van Dongen
Journal:  Pain       Date:  1997-03       Impact factor: 6.961

6.  Correlation between MRI-based stereotactic thalamic deep brain stimulation electrode placement, macroelectrode stimulation and clinical response to tremor control.

Authors:  F Mobin; A A De Salles; E J Behnke; R Frysinger
Journal:  Stereotact Funct Neurosurg       Date:  1999       Impact factor: 1.875

7.  Thalamotomy caused by cardioversion in a patient treated with deep brain stimulation.

Authors:  T Yamamoto; Y Katayama; C Fukaya; J Kurihara; H Oshima; M Kasai
Journal:  Stereotact Funct Neurosurg       Date:  2000       Impact factor: 1.875

Review 8.  Tumor ablation with radio-frequency energy.

Authors:  G S Gazelle; S N Goldberg; L Solbiati; T Livraghi
Journal:  Radiology       Date:  2000-12       Impact factor: 11.105

9.  Radio-frequency tissue ablation: effect of pharmacologic modulation of blood flow on coagulation diameter.

Authors:  S N Goldberg; P F Hahn; E F Halpern; R M Fogle; G S Gazelle
Journal:  Radiology       Date:  1998-12       Impact factor: 11.105

10.  The effect of magnetic resonance imagers on implanted neurostimulators.

Authors:  C A Gleason; N F Kaula; H Hricak; R A Schmidt; E A Tanagho
Journal:  Pacing Clin Electrophysiol       Date:  1992-01       Impact factor: 1.976

View more
  29 in total

1.  Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: in vitro and in vivo assessment of safety and function at 1.5 T.

Authors:  Ariel Roguin; Menekhem M Zviman; Glenn R Meininger; E Rene Rodrigues; Timm M Dickfeld; David A Bluemke; Albert Lardo; Ronald D Berger; Hugh Calkins; Henry R Halperin
Journal:  Circulation       Date:  2004-07-26       Impact factor: 29.690

Review 2.  Network effects of deep brain stimulation.

Authors:  Ahmad Alhourani; Michael M McDowell; Michael J Randazzo; Thomas A Wozny; Efstathios D Kondylis; Witold J Lipski; Sarah Beck; Jordan F Karp; Avniel S Ghuman; R Mark Richardson
Journal:  J Neurophysiol       Date:  2015-08-12       Impact factor: 2.714

3.  Apparent diffusion coefficient reveals gray and white matter disease, and T2 mapping detects white matter disease in the brain in feline alpha-mannosidosis.

Authors:  C H Vite; S Magnitsky; D Aleman; P O'Donnell; K Cullen; W Ding; S Pickup; J H Wolfe; H Poptani
Journal:  AJNR Am J Neuroradiol       Date:  2007-11-01       Impact factor: 3.825

4.  Compatibility of temporary pacemaker myocardial pacing leads with magnetic resonance imaging: an ex vivo tissue study.

Authors:  Alexander Pfeil; Stefanie Drobnik; Reinhard Rzanny; Anas Aboud; Joachim Böttcher; Peter Schmidt; Christian Ortmann; Gita Mall; Khosro Hekmat; Bernhard Brehm; Juergen Reichenbach; Thomas E Mayer; Gunter Wolf; Andreas Hansch
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-24       Impact factor: 2.357

5.  Quantitatively validating the efficacy of artifact suppression techniques to study the cortical consequences of deep brain stimulation with magnetoencephalography.

Authors:  Matthew J Boring; Zachary F Jessen; Thomas A Wozny; Michael J Ward; Ashley C Whiteman; R Mark Richardson; Avniel Singh Ghuman
Journal:  Neuroimage       Date:  2019-05-31       Impact factor: 6.556

6.  Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants.

Authors:  Laleh Golestanirad; Maria Ida Iacono; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Michael D Fox; Todd Herrington; Elfar Adalsteinsson; Cristen LaPierre; Azma Mareyam; Lawrence L Wald
Journal:  Neuroimage       Date:  2016-12-21       Impact factor: 6.556

7.  Computational electromagnetic analysis in a human head model with EEG electrodes and leads exposed to RF-field sources at 915 MHz and 1748 MHz.

Authors:  Leonardo M Angelone; Giorgi Bit-Babik; Chung-Kwang Chou
Journal:  Radiat Res       Date:  2010-07       Impact factor: 2.841

8.  Effect of the extracranial deep brain stimulation lead on radiofrequency heating at 9.4 Tesla (400.2 MHz).

Authors:  Devashish Shrivastava; Aviva Abosch; Timothy Hanson; Jinfeng Tian; Akshay Gupte; Paul A Iaizzo; J Thomas Vaughan
Journal:  J Magn Reson Imaging       Date:  2010-09       Impact factor: 4.813

9.  Safety of retained microcatheters: an evaluation of radiofrequency heating in endovascular microcatheters with nitinol, tungsten, and polyetheretherketone braiding at 1.5 T and 3 T.

Authors:  Aaron D Losey; Prasheel Lillaney; Alastair J Martin; Van V Halbach; Daniel L Cooke; Christopher F Dowd; Randall T Higashida; David A Saloner; Mark W Wilson; Maythem Saeed; Steven W Hetts
Journal:  J Neurointerv Surg       Date:  2013-05-18       Impact factor: 5.836

10.  Safety of localizing epilepsy monitoring intracranial electroencephalograph electrodes using MRI: radiofrequency-induced heating.

Authors:  David W Carmichael; John S Thornton; Roman Rodionov; Rachel Thornton; Andrew McEvoy; Philip J Allen; Louis Lemieux
Journal:  J Magn Reson Imaging       Date:  2008-11       Impact factor: 4.813

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.