Attenuation correction of myocardial SPECT images with X-ray CT: effects of registration errors between X-ray CT and SPECT.

Yasuyuki Takahashi1, Kenya Murase, Hiroshi Higashino, Teruhito Mochizuki, Nobutoku Motomura.   

Abstract

PURPOSE: Attenuation correction with an X-ray CT image is a new method to correct attenuation on SPECT imaging, but the effect of the registration errors between CT and SPECT images is unclear. In this study, we investigated the effects of the registration errors on myocardial SPECT, analyzing data from a phantom and a human volunteer.
METHODS: Registerion (fusion) of the X-ray CT and SPECT images was done with standard packaged software in three dimensional fashion, by using linked transaxial, coronal and sagittal images. In the phantom study, an X-ray CT image was shifted 1 to 3 pixels on the x, y and z axes, and rotated 6 degrees clockwise. Attenuation correction maps generated from each misaligned X-ray CT image were used to reconstruct misaligned SPECT images of the phantom filled with 201Tl. In a human volunteer, X-ray CT was acquired in different conditions (during inspiration vs. expiration). CT values were transferred to an attenuation constant by using straight lines; an attenuation constant of 0/cm in the air (CT value = -1,000 HU) and that of 0.150/cm in water (CT value = 0 HU). For comparison, attenuation correction with transmission CT (TCT) data and an external gamma-ray source (99mTc) was also applied to reconstruct SPECT images.
RESULTS: Simulated breast attenuation with a breast attachment, and inferior wall attenuation were properly corrected by means of the attenuation correction map generated from X-ray CT. As pixel shift increased, deviation of the SPECT images increased in misaligned images in the phantom study. In the human study, SPECT images were affected by the scan conditions of the X-ray CT.
CONCLUSION: Attenuation correction of myocardial SPECT with an X-ray CT image is a simple and potentially beneficial method for clinical use, but accurate registration of the X-ray CT to SPECT image is essential for satisfactory attenuation correction.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12416584     DOI: 10.1007/bf02990083

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  9 in total

1.  Mutidetector-row CT and quantitative gated SPECT for the assessment of left ventricular function in small hearts: the cardiac physical phantom study using a combined SPECT/CT system.

Authors:  Daisuke Utsunomiya; Seiji Tomiguchi; Kazuo Awai; Shinya Shiraishi; Takeshi Nakaura; Yasuyuki Yamashita
Journal:  Eur Radiol       Date:  2006-02-03       Impact factor: 5.315

2.  Prevalence of misregistration between SPECT and CT for attenuation-corrected myocardial perfusion SPECT.

Authors:  Sibyll Goetze; Richard L Wahl
Journal:  J Nucl Cardiol       Date:  2007-04       Impact factor: 5.952

Review 3.  Technological development and advances in single-photon emission computed tomography/computed tomography.

Authors:  Youngho Seo; Carina Mari; Bruce H Hasegawa
Journal:  Semin Nucl Med       Date:  2008-05       Impact factor: 4.446

4.  Is SPECT or CT Based Attenuation Correction More Quantitatively Accurate for Dedicated Breast SPECT Acquired with Non-Traditional Trajectories?

Authors:  Kristy L Perez; Steve D Mann; Jan H Pachon; Priti Madhav; Martin P Tornai
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2010 Oct-Nov

5.  A single CT for attenuation correction of both rest and stress SPECT myocardial perfusion imaging: a retrospective feasibility study.

Authors:  Mark A Ahlman; Paul J Nietert; Amy E Wahlquist; Jill M Serguson; Max W Berry; Pal Suranyi; Songtao Liu; Kenneth M Spicer
Journal:  Int J Clin Exp Med       Date:  2014-01-15

6.  Impact of attenuation correction and gated acquisition in SPECT myocardial perfusion imaging: results of the multicentre SPAG (SPECT Attenuation Correction vs Gated) study.

Authors:  Dario Genovesi; Assuero Giorgetti; Alessia Gimelli; Annette Kusch; Irene D'Aragona Tagliavia; Mirta Casagranda; Giorgio Cannizzaro; Raffaele Giubbini; Francesco Bertagna; Giorgio Fagioli; Massimiliano Rossi; Annadina Romeo; Pietro Bertolaccini; Rita Bonini; Paolo Marzullo
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-06-09       Impact factor: 9.236

7.  A Bayesian iterative transmission gradient reconstruction algorithm for cardiac SPECT attenuation correction.

Authors:  James A Case; Bai Ling Hsu; Timothy M Bateman; S James Cullom
Journal:  J Nucl Cardiol       Date:  2007-04-27       Impact factor: 5.952

8.  Optimisation of whole-body PET/CT scanning protocols.

Authors:  H Zaidi
Journal:  Biomed Imaging Interv J       Date:  2007-04-01

9.  Attenuation artifact, attenuation correction, and the future of myocardial perfusion SPECT.

Authors:  Bhupinder Singh; Timothy M Bateman; James A Case; Gary Heller
Journal:  J Nucl Cardiol       Date:  2007-04       Impact factor: 3.872

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.