Non-invasive cardiac output determination: comparison of a new partial-rebreathing technique with thermodilution.

J C Binder1, W G Parkin.   

Abstract

This study compares a derivative Fick technique using carbon dioxide (CO2) with the thermodilution pulmonary artery catheter (PAC), for determination of cardiac output (CO). Subjects were sedated, mechanically ventilated adults following elective cardiac surgery Microprocessor controlled deadspace activation and side-stream capnography in a ventilator circuit enabled calculation of CO (CO(CO2)) every four minutes. Thermodilution CO (CO(TD)) was performed as clinically indicated and at 20-minute intervals. Simultaneous CO(TD)/CO(CO2) pairs were recorded from time of admission to ICU for a minimum period of two hours for each patient. There were 358 CO(TD)/CO(CO2) pairs recorded from 41 patients. Cardiac output measurements ranged from 2. 7 to 10.6 l/min. The bias (Bland-Altman) was 0.050 l/min (95% CI -0.024 to 0.125 l/min). The 95% limits of agreement were -1.354 to 1.455 l/min. This simple, non-invasive partial-rebreathing technique is a valid alternative to thermodilution for cardiac output determination in sedated, mechanically ventilated patients. There are significant implications for improved safety, reduced complexity and reduced cost in anaesthesia and intensive care.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11261904     DOI: 10.1177/0310057X0102900103

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  9 in total

1.  Development and in vitro validation of a device for measuring non-shunt cardiac output by nitrous oxide throughflow.

Authors:  G M Vartuli; R B Burfoot; G J B Robinson; P J Peyton; P A Junor
Journal:  Med Biol Eng Comput       Date:  2002-07       Impact factor: 2.602

2.  A novel approach to measure cardiac output noninvasively: a comparison with the thermodilution method on critical care patients.

Authors:  John Kabal; Bruce K Lagerman
Journal:  J Clin Monit Comput       Date:  2004-06       Impact factor: 2.502

3.  Non-invasive automated measurement of cardiac output during stable cardiac surgery using a fully integrated differential CO(2) Fick method.

Authors:  Philip J Peyton; Daniel Thompson; Paul Junor
Journal:  J Clin Monit Comput       Date:  2008-07-12       Impact factor: 2.502

4.  Cross-comparisons of trending accuracies of continuous cardiac-output measurements: pulse contour analysis, bioreactance, and pulmonary-artery catheter.

Authors:  Bouchra Lamia; Hyung Kook Kim; Donald A Severyn; Michael R Pinsky
Journal:  J Clin Monit Comput       Date:  2017-02-10       Impact factor: 2.502

5.  A comparative evaluation of thermodilution and partial CO2 rebreathing techniques for cardiac output assessment in critically ill patients during assisted ventilation.

Authors:  Monica Rocco; Gustavo Spadetta; Andrea Morelli; Donatella Dell'Utri; Patrizia Porzi; Giorgio Conti; Paolo Pietropaoli
Journal:  Intensive Care Med       Date:  2003-12-03       Impact factor: 17.440

6.  The effect of carbon dioxide on near-death experiences in out-of-hospital cardiac arrest survivors: a prospective observational study.

Authors:  Zalika Klemenc-Ketis; Janko Kersnik; Stefek Grmec
Journal:  Crit Care       Date:  2010-04-08       Impact factor: 9.097

7.  Noninvasive cardiac output assessment during heart surgery.

Authors:  Piero Ceriana; Marco Mau Relli; Antonio Braschi; Paola Baiardi; Donatella De Amici
Journal:  J Clin Monit Comput       Date:  2002-02       Impact factor: 2.502

8.  Invasive and noninvasive cardiovascular monitoring options for cardiac surgery.

Authors:  Dominic P Recco; Nathalie Roy; Alexander J Gregory; Kevin W Lobdell
Journal:  JTCVS Open       Date:  2022-04-11

9.  Difference in end-tidal CO2 between asphyxia cardiac arrest and ventricular fibrillation/pulseless ventricular tachycardia cardiac arrest in the prehospital setting.

Authors:  Stefek Grmec; Katja Lah; Ksenija Tusek-Bunc
Journal:  Crit Care       Date:  2003-09-24       Impact factor: 9.097

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.