The new glucose standard POCT12-A3 misses the mark.

Jan S Krouwer1.   

Abstract

POCT12-A3 is a Clinical Laboratory Standards Institute standard for hospitals about hospital glucose meter procedures and performance standards. I have reviewed this standard based on the attributes of an ideal performance standard. POCT12-A3 has tighter limits than its predecessor for 95% of results, the limits widen for 98% of results, and there are no limits for 2% of results. It is hard to fathom that 2% of the results are unspecified and could cause life-threatening results, as glucose meters do not perform this poorly. There should be a specification for unreported results since, by definition, point-of-care-testing assays are time sensitive. POCT12-A3 provides useful advice about the glucose testing procedure but provides evaluation guidance only about analytical performance. Moreover, the recommended protocol to assess meter performance is biased and likely to underestimate the observed performance. The guideline would be improved if its specification were based on an error grid and contained evaluation protocols for user errors.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24124969      PMCID: PMC3876386          DOI: 10.1177/193229681300700532

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  6 in total

1.  Wrong thinking about glucose standards.

Authors:  Jan S Krouwer
Journal:  Clin Chem       Date:  2010-01-28       Impact factor: 8.327

2.  Analysis of the performance of the OneTouch SelectSimple blood glucose monitoring system: why ease of use studies need to be part of accuracy studies.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

3.  Evaluation of the analytical performance of the coulometry-based Optium Omega blood glucose meter: what do such evaluations show?

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

4.  Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician.

Authors:  Svein Skeie; Geir Thue; Kari Nerhus; Sverre Sandberg
Journal:  Clin Chem       Date:  2002-07       Impact factor: 8.327

5.  Accuracy of roche accu-chek inform whole blood capillary, arterial, and venous glucose values in patients receiving intensive intravenous insulin therapy after cardiac surgery.

Authors:  Brad S Karon; Gunjan Y Gandhi; Gregory A Nuttall; Sandra C Bryant; Hartzell V Schaff; M Molly McMahon; Paula J Santrach
Journal:  Am J Clin Pathol       Date:  2007-06       Impact factor: 2.493

Review 6.  A review of standards and statistics used to describe blood glucose monitor performance.

Authors:  Jan S Krouwer; George S Cembrowski
Journal:  J Diabetes Sci Technol       Date:  2010-01-01
  6 in total
  2 in total

1.  Why the New FDA Glucose Meter POCT Guidance Is Disappointing.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2016-12-05

2.  Falsely Elevated Glucose Concentrations in Peritoneal Dialysis Patients Using Icodextrin.

Authors:  Kübra Dogan; Damla Kayalp; Gözde Ceylan; Alper Azak; Mehmet Senes; Murat Duranay; Dogan Yucel
Journal:  J Clin Lab Anal       Date:  2015-10-29       Impact factor: 2.352

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.