Evaluation of the validity and reliability of A-scan ultrasound biometry with a single use disposable cover.

K Cass1, C M Thompson, C Tromans, I C J Wood.   

Abstract

BACKGROUND: The UK Medical Devices Agency has suggested that ophthalmic practitioners should, where practicable and not compromising clinical outcome, restrict corneal contact devices to single patient use to minimise a remote theoretical risk of transmission of new variant Creutzfeldt-Jakob disease (vCJD). This study reports on a modified technique of ultrasound A-scan biometry that complies with the MDA recommendations.
METHODS: The right eyes of 37 consecutive hospital patients had a series of biometry readings taken with a Humphrey 820 A-scan instrument with a plane wave transducer use d conventionally and with the addition of a disposable latex cover.
RESULTS: Intrasessional repeatability of axial length measurements was similar for conventional readings--mean difference 0.027 mm, 95% confidence intervals (CI) +/- 0.44 mm and those taken with a disposable cover (0.028 mm, CI +/- 0.38). Intersessional repeatability was equivalent with (0.002 mm, CI +.- 0.51) and without a cover (0.03 mm, CI +/- 0.51). Readings with a cover were not significantly different from those without (paired t test; p >0.05), but tended to be greater (mean difference 0.085 mm, CI +/- 0.60).
CONCLUSIONS: These findings suggest that corneal contact biometry with a disposable cover is a viable and theoretically safer alternative to the conventional technique.

Entities:  

Mesh:

Year:  2002        PMID: 11864896      PMCID: PMC1771057          DOI: 10.1136/bjo.86.3.344

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  16 in total

1.  Assessment of a new through-the-eyelid technique for 'A' scan ultrasound ocular axial length measurement.

Authors:  F Laws; D Laws; I Wood; D Clark
Journal:  Ophthalmic Physiol Opt       Date:  1998-09       Impact factor: 3.117

2.  CJD and the eye.

Authors:  A Tullo; R Buckley; M Painter
Journal:  Eye (Lond)       Date:  2000-06       Impact factor: 3.775

3.  Repeatability, reproducibility and intersession variability of the Allergan Humphrey ultrasonic biometer.

Authors:  A R Rudnicka; C F Steele; D P Crabb; D F Edgar
Journal:  Acta Ophthalmol (Copenh)       Date:  1992-06

4.  Effects of different ocular fixation conditions on A-scan ultrasound biometry measurements.

Authors:  C F Steele; D P Crabb; D F Edgar
Journal:  Ophthalmic Physiol Opt       Date:  1992-10       Impact factor: 3.117

5.  The repeatability of automated and clinician refraction.

Authors:  M A Bullimore; R E Fusaro; C W Adams
Journal:  Optom Vis Sci       Date:  1998-08       Impact factor: 1.973

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Comparison of corneal versus through-the-lid A-scan ultrasound biometry.

Authors:  J D Twelker; S Kirschbaum; K Zadnik; D O Mutti
Journal:  Optom Vis Sci       Date:  1997-10       Impact factor: 1.973

8.  Partial coherence interferometry: a novel approach to biometry in cataract surgery.

Authors:  W Drexler; O Findl; R Menapace; G Rainer; C Vass; C K Hitzenberger; A F Fercher
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

9.  A new variant of Creutzfeldt-Jakob disease in the UK.

Authors:  R G Will; J W Ironside; M Zeidler; S N Cousens; K Estibeiro; A Alperovitch; S Poser; M Pocchiari; A Hofman; P G Smith
Journal:  Lancet       Date:  1996-04-06       Impact factor: 79.321

10.  Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis.

Authors:  W Haigis; B Lege; N Miller; B Schneider
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2000-09       Impact factor: 3.117

View more
  2 in total

1.  Comparison axial length measurements from three biometric instruments in high myopia.

Authors:  Xiao-Gang Wang; Jing Dong; Yu-Lan Pu; Hui-Jun Liu; Qiang Wu
Journal:  Int J Ophthalmol       Date:  2016-06-18       Impact factor: 1.779

2.  Comparison of axial length, anterior chamber depth and intraocular lens power between IOLMaster and ultrasound in normal, long and short eyes.

Authors:  Jing Dong; Yaqin Zhang; Haining Zhang; Zhijie Jia; Suhua Zhang; Xiaogang Wang
Journal:  PLoS One       Date:  2018-03-15       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.