Literature DB >> 1636392

Repeatability, reproducibility and intersession variability of the Allergan Humphrey ultrasonic biometer.

A R Rudnicka1, C F Steele, D P Crabb, D F Edgar.   

Abstract

We evaluated repeatability and reproducibility of the Allergan Humphrey model 820 ultrasonic biometer. Sixty eyes (30 subjects) free from ophthalmological abnormality were examined on two separate occasions by two experimenters, one experienced and one inexperienced. The 95% confidence limits for repeatability lie between +0.12/-0.13 mm and reproducibility between +0.15/-0.14 mm for AC depth, between +0.11/-0.13 mm and +0.12/-0.16 mm for lens thickness, between +0.17/-0.19 mm and +0.19/-0.18 mm for right axial length and between +0.26/-0.40 mm and +0.22/-0.25 mm for left axial length. The 40% gain used to measure left axial lengths produced poor retinal echospikes, which may account for reduced accuracy compared to the right eyes using 60% gain. Ten subjects (20 eyes) returned for 3 further visits and the variability over 5 sessions was examined using a two-factor ANOVA with repeated measures on one factor. There were no significant experimenter or session effects and no interaction effect. This biometer gives clinically repeatable and reproducible results and is convenient to use.

Entities:  

Mesh:

Year:  1992        PMID: 1636392     DOI: 10.1111/j.1755-3768.1992.tb08573.x

Source DB:  PubMed          Journal:  Acta Ophthalmol (Copenh)        ISSN: 0001-639X


  5 in total

1.  High spatial resolution automated perimetry in glaucoma.

Authors:  M C Westcott; A I McNaught; D P Crabb; F W Fitzke; R A Hitchings
Journal:  Br J Ophthalmol       Date:  1997-06       Impact factor: 4.638

2.  Improvements on Littmann's method of determining the size of retinal features by fundus photography.

Authors:  A G Bennett; A R Rudnicka; D F Edgar
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1994-06       Impact factor: 3.117

3.  Evaluation of the validity and reliability of A-scan ultrasound biometry with a single use disposable cover.

Authors:  K Cass; C M Thompson; C Tromans; I C J Wood
Journal:  Br J Ophthalmol       Date:  2002-03       Impact factor: 4.638

4.  Validation of optical coherence tomography-based crystalline lens thickness measurements in children.

Authors:  Bret M Lehman; David A Berntsen; Melissa D Bailey; Karla Zadnik
Journal:  Optom Vis Sci       Date:  2009-03       Impact factor: 1.973

5.  Measurement of optic disc size: equivalence of methods to correct for ocular magnification.

Authors:  D F Garway-Heath; A R Rudnicka; T Lowe; P J Foster; F W Fitzke; R A Hitchings
Journal:  Br J Ophthalmol       Date:  1998-06       Impact factor: 5.908

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.