Accuracy evaluation of surface-based registration methods in a computer navigation system for hip surgery performed through a posterolateral approach.

N Sugano1, T Sasama, Y Sato, Y Nakajima, T Nishii, K Yonenobu, S Tamura, T Ochi.   

Abstract

OBJECTIVE: Many computer navigation systems have recently been developed for brain surgery, and the use of such systems in orthopedic surgery is increasing. Intraoperative registration of preoperative images is one of the most important steps in controlling the overall accuracy of computer navigation systems. Various parameters, such as CT-scan slice thickness, reconstruction pitch, intraoperative data sampling area, and data sampling volume, may affect the accuracy of registration. The purpose of this study was to evaluate the effect of the aforementioned parameters on the accuracy of registration for hip surgery performed through a posterolateral approach, and to find a clinically suitable trade-off between accuracy and surgical invasiveness.
MATERIALS AND METHODS: One cadaveric pelvis and one cadaveric femur were used for this study. Four alumina ceramic balls with a diameter of 28 mm and within 1 micrometer of sphericity were attached to the pelvis, and three similar balls attached to the femur, to determine relative position. CT-scan images of the pelvis and femur were obtained with a helical scanner. Three sets of slice thickness and slice pitch were chosen for data acquisition, and two additional sets of reconstructed data were made. Bone contours were extracted by cutting out the surrounding substrate at a given CT number threshold, and surface models of the bone were made from the resultant data. The positions of the pelvis and femur were tracked by LED markers attached to the bone using an optical three-dimensional position sensor (OPTOTRAK). Registration of the computer models to the real objects was performed by measuring the position of a certain number of surface points on each object with an OPTOTRAK pen-probe. RESULTS AND
CONCLUSION: Slice thickness and reconstruction pitch affected the accuracy of registration. As the sampling area was expanded from the periarticular area to the distant peripheral area, accuracy increased slightly. Accuracy did not increase when the whole area was used, but in fact decreased, especially in the femur. The positive effect of increasing the number of sampling points was saturated at 30 points when the surface of the periarticular area was sampled. The following trade-off between accuracy and invasiveness, in terms of various parameters of preoperative and intraoperative data, is proposed as clinically optimal: perform the CT scan with 3-mm slice thickness and 1-mm reconstruction pitch, and sample the periarticular area with 30 sampling points. With these parameters, the accuracy of registration was 1.2 mm and 0.9 degrees of bias with 0.7 mm and 0.3 degrees of RMS in the pelvis, and 1.4 mm and 0.6 degrees of bias with 1.3 mm and 0.3 degrees of RMS in the femur. Copyright 2001 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2001        PMID: 11835614     DOI: 10.1002/igs.10011

Source DB:  PubMed          Journal:  Comput Aided Surg        ISSN: 1092-9088


  12 in total

1.  Evaluation of achievable registration accuracy of the femur during minimally invasive total hip replacement.

Authors:  F C Popescu; M Viceconti; F Traina; A Toni
Journal:  Med Biol Eng Comput       Date:  2005-07       Impact factor: 2.602

2.  [A-mode ultrasonic pointer for navigated pelvic surgery].

Authors:  M Oszwald; M Citak; J Kowal; C Amstutz; D Kendoff; T Kirchhoff; L P Nolte; C Krettek; T Hüfner
Journal:  Unfallchirurg       Date:  2008-03       Impact factor: 1.000

3.  Six sigma analysis of minimally invasive acetabular arthroplasty: a preliminary investigation.

Authors:  David A Heck; James B Stiehl
Journal:  Clin Orthop Relat Res       Date:  2009-05-22       Impact factor: 4.176

4.  Does CT-based navigation improve the long-term survival in ceramic-on-ceramic THA?

Authors:  Nobuhiko Sugano; Masaki Takao; Takashi Sakai; Takashi Nishii; Hidenobu Miki
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

5.  The value of computed tomography based navigation in revision total hip arthroplasty.

Authors:  Kazunari Kuroda; Tamon Kabata; Toru Maeda; Yoshitomo Kajino; Shin Watanabe; Shintaro Iwai; Fujita Kenji; Kazuhiro Hasegawa; Daisuke Inoue; Hiroyuki Tsuchiya
Journal:  Int Orthop       Date:  2013-11-19       Impact factor: 3.075

6.  Robot-assisted primary cementless total hip arthroplasty using surface registration techniques: a short-term clinical report.

Authors:  Nobuo Nakamura; Nobuhiko Sugano; Takashi Nishii; Hidenobu Miki; Akihiro Kakimoto; Mitsuyoshi Yamamura
Journal:  Int J Comput Assist Radiol Surg       Date:  2009-02-13       Impact factor: 2.924

7.  Effect of improved navigation performance on the accuracy of implant placement in total hip arthroplasty with a CT-based navigation system.

Authors:  Ichiro Nakahara; Takayuki Kyo; Yasuo Kuroda; Hidenobu Miki
Journal:  J Artif Organs       Date:  2018-04-02       Impact factor: 1.731

8.  Computer-assisted, Le Fort-based, face-jaw-teeth transplantation: a pilot study on system feasiblity and translational assessment.

Authors:  Ryan J Murphy; Chad R Gordon; Ehsan Basafa; Peter Liacouras; Gerald T Grant; Mehran Armand
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-09-18       Impact factor: 2.924

9.  Comparison of rotational acetabular osteotomy performed with navigation by surgeons with different levels of experience of osteotomies.

Authors:  Masaki Takao; Takashi Nishii; Takashi Sakai; Nobuhiko Sugano
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-10-25       Impact factor: 2.924

10.  Accessing 3D Location of Standing Pelvis: Relative Position of Sacral Plateau and Acetabular Cavities versus Pelvis.

Authors:  E Berthonnaud; R Hilmi; J Dimnet
Journal:  Radiol Res Pract       Date:  2012-04-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.