Literature DB >> 11792782

Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study.

Moussa Hamadouche1, Pierre Boutin, Jacques Daussange, Mark E Bolander, Laurent Sedel.   

Abstract

BACKGROUND: The purpose of this retrospective study was to report the results, after a minimum of 18.5 years of follow-up, in a consecutive series of total hip arthroplasties performed with an alumina-on-alumina combination.
METHODS: One hundred and eighteen consecutive total hip arthroplasties were performed in 106 patients between 1979 and 1980. The prostheses combined a 32-mm alumina head and an all-alumina socket. Both components were cemented in eighty-five hips, both components were implanted without cement in twenty-nine, and only the stem was cemented in four. The mean age of the patients at the time of the index arthroplasty was 62.2 years (range, thirty-two to eighty-nine years).
RESULTS: At the 18.5 to 20.5-year follow-up evaluation, forty-five patients (fifty-one hips) were alive and had not had a revision, twenty-five patients (twenty-five hips) had undergone revision of one or both components, twenty-seven patients (thirty hips) had died, and nine patients (twelve hips) had been lost to follow-up. The mean Merle d'Aubigné hip score (and standard deviation) was 16.2 +/- 1.8 points at the latest follow-up evaluation. The rate of survival at twenty years, with revision for any reason as the end-point, was 85.6% for the cementless cups compared with 61.2% for the cemented cups and 84.9% for the cementless stems compared with 87.3% for the cemented stems. Wear of the prosthetic components was undetectable on plain radiographs. Periprosthetic cystic or scalloped lesions requiring the use of allograft bone during revision were present in three of the twenty-five revised hips. In addition, seven hips had moderate acetabular osteolysis treated with a 4-mm-larger cup. No fracture of the alumina socket or head was recorded. The mean acetabular wear rate in this series was <0.025 mm/yr.
CONCLUSION: With the alumina-on-alumina total hip arthroplasty, minimal wear rates and limited osteolysis can be expected up to twenty years after the operation, provided that sound acetabular component fixation is obtained.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11792782

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  63 in total

1.  A squeaky reputation: the problem may be design-dependent.

Authors:  Javad Parvizi; Bahar Adeli; Justin C Wong; Camilo Restrepo; Richard H Rothman
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

2.  Is routine mid-term total hip arthroplasty surveillance beneficial?

Authors:  James A Keeney; Bradley S Ellison; William J Maloney; John C Clohisy
Journal:  Clin Orthop Relat Res       Date:  2012-06-06       Impact factor: 4.176

3.  Cup loosening after cemented Metasul® total hip replacement: a retrieval analysis.

Authors:  Christophe Nich; Moussa Hamadouche
Journal:  Int Orthop       Date:  2010-06-11       Impact factor: 3.075

4.  Do "premium" joint implants add value?: analysis of high cost joint implants in a community registry.

Authors:  Terence J Gioe; Amit Sharma; Penny Tatman; Susan Mehle
Journal:  Clin Orthop Relat Res       Date:  2011-01       Impact factor: 4.176

5.  Clinical and biological assessment of cemented titanium femoral stems: an 11-year experience.

Authors:  Patrick Boyer; Jean-Yves Lazennec; Joel Poupon; Marc-Antoine Rousseau; Philippe Ravaud; Yves Catonné
Journal:  Int Orthop       Date:  2008-10-28       Impact factor: 3.075

6.  The first 50 years of total hip arthroplasty: lessons learned.

Authors:  William H Harris
Journal:  Clin Orthop Relat Res       Date:  2008-11-04       Impact factor: 4.176

7.  Incidence of 'squeaking' after ceramic-on-ceramic total hip arthroplasty.

Authors:  Kenny Mai; Christopher Verioti; Kace A Ezzet; Steven N Copp; Richard H Walker; Clifford W Colwell
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

8.  Ceramic-on-ceramic bearing decreases the cumulative long-term risk of dislocation.

Authors:  Philippe Hernigou; Yasuhiro Homma; Olivier Pidet; Isaac Guissou; Jacques Hernigou
Journal:  Clin Orthop Relat Res       Date:  2013-12       Impact factor: 4.176

9.  Cementless total hip arthroplasty with ceramic-on-ceramic bearing in patients younger than 45 years with femoral-head osteonecrosis.

Authors:  Young-Hoo Kim; Yoowang Choi; Jun-Shik Kim
Journal:  Int Orthop       Date:  2009-09-26       Impact factor: 3.075

10.  Alumina inlay failure in cemented polyethylene-backed total hip arthroplasty.

Authors:  Kentaro Iwakiri; Hiroyoshi Iwaki; Yukihide Minoda; Hirotsugu Ohashi; Kunio Takaoka
Journal:  Clin Orthop Relat Res       Date:  2008-02-21       Impact factor: 4.176