The precision of radiostereometric measurements. Manual vs. digital measurements.

Niclas Börlin1, Truike Thien, Johan Kärrholm.   

Abstract

The precision of digital vs. manual radiostereometric measurements in total hip arthroplasty was evaluated using repeated stereoradiographic exposures with an interval of 10-15 min. Ten Lubinus SP2 stems cemented into bone specimens and 12 patients with the same stem design were used to evaluate the precision of stem translations and rotations. The precision of translations and rotations of the cup and femoral head penetration was studied in 12 patients with whole polyethylene cups. The use of a measurement method based on digitised radiographs improved the precision for some of the motion parameters, whereas many of them did not change. A corresponding pattern was observed for both the intra- and interobserver error. Of the wear parameters, the most pronounced improvements were the 3D wear and in the proximal-distal direction, although the anterior-posterior precision was also improved. The mean errors of rigid body and elliptic fitting decreased in all evaluations but one, consistent with a more reproducible identification of the markers centres and the edge of the femoral head. Increased precision of radiostereometric measurements may be used to increase the statistical power of future randomised studies and to study new fields in orthopaedics requiring higher precision than has been available with RSA based on manual measurements.

Entities:  

Mesh:

Year:  2002        PMID: 11747885     DOI: 10.1016/s0021-9290(01)00162-2

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  17 in total

1.  Early migration of the cemented tibial component of unicompartmental knee arthroplasty: a radiostereometry study.

Authors:  Andrea Ensini; Paolo Barbadoro; Alberto Leardini; Fabio Catani; Sandro Giannini
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-06-04       Impact factor: 4.342

2.  The effect of standard lumbar discectomy on segmental motion: 5-year follow-up using radiostereometry.

Authors:  K Halldin; B Zoëga; P Nyberg; J Kärrholm; B I Lind
Journal:  Int Orthop       Date:  2005-03-01       Impact factor: 3.075

3.  Is increased segmental motion early after lumbar discectomy related to poor clinical outcome 5 years later?

Authors:  K Halldin; B Zoëga; J Kärrholm; B I Lind; P Nyberg
Journal:  Int Orthop       Date:  2005-06-04       Impact factor: 3.075

4.  Long-term results and bone remodeling after THA with a short, metaphyseal-fitting anatomic cementless stem.

Authors:  Young-Hoo Kim; Jang-Won Park; Jun-Shik Kim; Jun-Seok Kang
Journal:  Clin Orthop Relat Res       Date:  2013-10-26       Impact factor: 4.176

5.  In vitro quantification of the performance of model-based mono-planar and bi-planar fluoroscopy for 3D joint kinematics estimation.

Authors:  Luca Tersi; Arnaud Barré; Silvia Fantozzi; Rita Stagni
Journal:  Med Biol Eng Comput       Date:  2012-11-11       Impact factor: 2.602

6.  Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  George K Karidakis; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

7.  The Elite-Plus stem migrates more than the flanged Charnley stem.

Authors:  Thord von Schewelov; Lennart Sanzén; Jack Besjakov; Ake Carlsson
Journal:  Acta Orthop       Date:  2010-06       Impact factor: 3.717

8.  Analysis of polyethylene wear in plain radiographs.

Authors:  Maiken Stilling; Kjeld Søballe; Niels Trolle Andersen; Kristian Larsen; Ole Rahbek
Journal:  Acta Orthop       Date:  2009-12       Impact factor: 3.717

9.  Notes on the use and interpretation of radiostereometric analysis.

Authors:  Brian Derbyshire; Robin J Prescott; Martyn L Porter
Journal:  Acta Orthop       Date:  2009-02       Impact factor: 3.717

10.  Dependence of model-based RSA accuracy on higher and lower implant surface model quality.

Authors:  Frank Seehaus; Judith Emmerich; Bart L Kaptein; Henning Windhagen; Christof Hurschler
Journal:  Biomed Eng Online       Date:  2013-04-16       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.