Literature DB >> 11559607

The clinical evaluation of the Respi-check mask: a new oxygen mask incorporating a breathing indicator.

A Breakell1, C Townsend-Rose.   

Abstract

UNLABELLED: Study objective-To investigate the correlation between the Respi-check sensor and simultaneous chest auscultation in determining the respiratory rates in adults.
METHODS: Random visits to a local accident and emergency (A&E) department were made and all patients wearing oxygen masks were recruited into the study. The new sensor was attached to the outside of the mask. One researcher auscultated the chest to count breaths, the other counted the sensor activity. Each was blinded to the activities of the other. Breaths were counted by each researcher simultaneously and independently over one minute. A total of 40 patients were recruited into the study. A difference of more than two breaths/min compared with chest auscultation was deemed as a sensor failure.
RESULTS: The respiratory rates of 40 patients were measured. There were 28 men, 12 women. Twenty six patients were wearing an Intersurgical high concentration (flow 12l/min) mask, 14 were wearing an aerosol mask with variable venturi (flow 3-12l/min) by Medicaid. Over one minute rates determined by the two methods were the same in 28 cases (70%). It was accurate to within one breath in 37 cases (93%) and to within two breaths in 39 (98%) cases and in one case (2.5%) the sensor failed. The mean difference (mean of the differences between rates obtained from auscultation and the new sensor) was -0.1282 breaths/min, with limits of agreement (d (2SD) between -1.414 to 1.157 breaths/min.
CONCLUSION: The Respi-check sensor provides an accurate method of estimating the respiratory rate in adult patients attending the A&E department.

Entities:  

Mesh:

Year:  2001        PMID: 11559607      PMCID: PMC1725676          DOI: 10.1136/emj.18.5.366

Source DB:  PubMed          Journal:  Emerg Med J        ISSN: 1472-0205            Impact factor:   2.740


  16 in total

1.  Inability to assess breath sounds during air medical transport by helicopter.

Authors:  R C Hunt; D M Bryan; V S Brinkley; T W Whitley; N H Benson
Journal:  JAMA       Date:  1991-04-17       Impact factor: 56.272

2.  The respiratory rate.

Authors:  M J Clancy; M J Williams
Journal:  Arch Emerg Med       Date:  1991-09

3.  End-tidal carbon dioxide monitoring in emergency medicine, Part 2: Clinical applications.

Authors:  K R Ward; D M Yealy
Journal:  Acad Emerg Med       Date:  1998-06       Impact factor: 3.451

4.  AURA: a new respiratory monitor.

Authors:  V Kulkarni; A Cyna; J M Hutchison; M E Tunstall; J R Mallard
Journal:  Biomed Sci Instrum       Date:  1990

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Measurement of respiratory rate and timing using a nasal thermocouple.

Authors:  M K Marks; M South; B G Carter
Journal:  J Clin Monit       Date:  1995-05

7.  Experimental evaluation of two new sensors for respiratory rate monitoring.

Authors:  M Vegfors; H Ugnell; B Hök; P A Oberg; C Lennmarken
Journal:  Physiol Meas       Date:  1993-05       Impact factor: 2.833

8.  Physiologic chest sounds and helicopter engine noise.

Authors:  T J Poulton; D W Worthington; T R Pasic
Journal:  Aviat Space Environ Med       Date:  1994-04

9.  Evaluating trauma care: the TRISS method. Trauma Score and the Injury Severity Score.

Authors:  C R Boyd; M A Tolson; W S Copes
Journal:  J Trauma       Date:  1987-04

10.  A new respiratory rate monitor: development and initial clinical experience.

Authors:  B Hök; L Wiklund; S Henneberg
Journal:  Int J Clin Monit Comput       Date:  1993-05
View more
  1 in total

1.  A capaciflector provides continuous and accurate respiratory rate monitoring for patients at rest and during exercise.

Authors:  Nick Hayward; Mahdi Shaban; James Badger; Isobel Jones; Yang Wei; Daniel Spencer; Stefania Isichei; Martin Knight; James Otto; Gurinder Rayat; Denny Levett; Michael Grocott; Harry Akerman; Neil White
Journal:  J Clin Monit Comput       Date:  2022-01-18       Impact factor: 1.977

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.