STATEMENT OF PROBLEM: Primary stability at the time of implant placement is related to the level of primary bone contact. The level of bone contact with implant is affected by thread design, surgical procedure and bone quality, etc.Entities:
Keywords: Bicortical engagement; ISQ; Removal torque value; Taper implant; Thread design
Year: 2009 PMID: 21165250 PMCID: PMC2994669 DOI: 10.4047/jap.2009.1.1.19
Source DB: PubMed Journal: J Adv Prosthodont ISSN: 2005-7806 Impact factor: 1.904
The characteristics of 6 different implants
Fig. 1Characteristics of thread design used in this study.
a. GS III® (Osstem, Seoul, South Korea), b. Osseotite NT®(3i, Florida, USA), c. Replace® Select (Nobel Biocare, Göteborg, Sweden), d. Sinus Quick (Neoplant, Seoul, South Korea), e. US III® (Osstem, Seoul, South Korea), f. Hexplant (Warantec, Seoul, South Korea).
Fig. 2a. One side of cortical bone was removed. b. Bone height was 10 mm for Group 1. c. Bone height was 13 mm for Group 2.
Fig. 3a. 6 implants were installed with engaging (Group 1). b. 6 implants were installed without engaging (Group 2).
Fig. 4Surgical procedures were performed following to technique recommended by each manufacturers.
Fig. 5Osstell Mentor®.
Fig. 6The RFA device (Smartpegs; Integration Diagnosis AB., Geml estadrägen was placed by hand tightening 5 - 10 Ncm on to the implant fixtures). The types of smartpegs were #1 for universal external hex, #6 for GSIII® and #13 for Replace® Select.
Fig. 7ISQ value were measured parallel and perpendicular to the bone specimens.
Fig. 8The photograph of torque gauge; removal torque was measured with MGT50 torque gauge.
Fig. 9For under-drilling the last step drill was skipped.
ISQ and removal torque value depending on thread designs in group 1. Data were represented means and standard deviation
1) Statistical significances were tested by one-way analysis of variance among groups.
Fig. 10ISQ and RTV in Group 1.
ISQ and removal torque depending on thread design in group 2
1) Statistical significances were tested by one-way analysis of variance among groups.
Fig. 11ISQ and RTV in Group 2.
Change of ISQ and removal torque value depending on the reduction of step drilling
1) Statistical significances were tested by Mann-Whitney's test.