Alan W Eberhardt1, R Travis McKee, John M Cuckler, Donald W Peterson, Preston R Beck, Jack E Lemons.
Abstract
Metal transfer to femoral heads may result from impingement against the metallic acetabular shell following subluxation/dislocation, or when metallic debris enters the articulation zone. Such transfers roughen the head surface, increasing polyethylene wear in total hip replacements. Presently, we examined the surface roughness of retrieved femoral heads with metallic transfer. Profilometry revealed roughness averages in regions of metal transfer averaging 0.380 mum for CoCr and 0.294 mum for ZrO(2) which were one order of magnitude higher than those from non-implanted controls. Scanning electron microscopy (SEM) revealed adherent transfers on these retrievals, with titanium presence confirmed by electron dispersive spectroscopy. Due to the concern for increased wear, metal transfer was induced on non-implanted heads, which were then articulated against flat polyethylene discs in multidirectional sliding wear tests. Increased polyethylene wear was associated with these specimens as compared to unaltered controls. SEM imaging provided visual evidence that the transfers remained adherent following the wear tests. Pre- and post-test roughness averages exceeded 1 mum for both the CoCr and ZrO(2) heads. Overall, these results suggest that metal transfer increases the surface roughness of CoCr and ZrO(2) femoral heads and that the transfers may remain adherent following articulation against polyethylene, leading to increased polyethylene wear.Entities:
Year: 2009 PMID: 20126576 PMCID: PMC2811347 DOI: 10.1155/2009/185456
Source DB: PubMed Journal: Int J Biomater ISSN: 1687-8787
Figure 1Example retrievals with clear visual evidence of metal transfer on the femoral heads.
Retrieval information (blank cells indicate unavailable information).
| Implant data | Donor data | |||||
|---|---|---|---|---|---|---|
| Manufacturer | Size (mm)/Material | Mos. in vivo | Age (yrs) | Sex | Weight (lbs.) | Height (inches) |
| J&J | 28/CoCr | 36 | 55 | M | 145 | 70 |
| Richards | 28/CoCr | 60 | 75 | F | 119 | 62 |
| Richards | 28/CoCr | 30 | 49 | F | 215 | 65 |
| Depuy | 32/CoCr | 108 | 82 | F | 145 | 65 |
| Biomet | 28/CoCr | 0.3 | 60 | F | ||
| 28/CoCr | M | 143 | 69 | |||
|
| ||||||
| Richards | 28/ZrO2 | 54 | ||||
| Richards | 28/ZrO2 | 48 | 39 | M | 215 | 72 |
| Biomet | 28/ZrO2 | 12 | 27 | |||
| Richards | 28/ZrO2 | 60 | 68 | M | ||
| Biomet | 28/ZrO2 | 2 | 46 | M | 139 | 69 |
| Richards | 28/ZrO2 | F | ||||
| Richards | 32/ZrO2 | 120 | 40 | F | ||
| Biomet | 28/ZrO2 | |||||
| Biomet | 28/ZrO2 | 49 | M | 185 | 73 | |
| 32/ZrO2 | 50 | M | ||||
Figure 2The OrthoPod wear simulator with mounted CoCr test head and transfer disc.
Figure 3(a) Control CoCr surface reveals residual polishing marks; (b) region of metal transfer (1000X).
Figure 4(a) Control ZrO2 surfaces appeared smooth; (b) region of metal transfer (1000X).
Surface parameters for retrievals (mean ± standard deviation).
| Head type |
|
|
|
|---|---|---|---|
| CoCr control | 0.012 ± 0.002 | 0.016 ± 0.004 | −0.85 ± 1.09 |
| ZrO2 control | 0.013 ± 0.002 | 0.018 ± 0.003 | −0.204 ± 0.114 |
| CoCr transfer | 0.380 ± 0.308 | 0.540 ± 0.512 | 0.597 ± 1.319 |
| ZrO2 transfer | 0.294 ± 0.294 | 0.363 ± 0.387 | 1.397 ± 1.365 |
Pre- and posttest surface parameters for induced-transfer specimens and controls used in wear tests.
| Material | Pretest | Posttest | ||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
| CoCr | 1.213 ± .156 | 2.208 ± .276 | 3.500 ± .638 | 1.010 ± .180 | 1.630 ± .250 | 2.800 ± .231 |
| ZrO2 | 1.024 ± .295 | 1.694 ± .504 | 3.400 ± .505 | 1.327 ± .960 | 2.034 ± 1.381 | 2.767 ± .437 |
Figure 5Wear (mass loss) for the control femoral heads and those with induced metal transfer.
Figure 61000X images of postwear transfer regions: (a) CoCr (b) ZrO2.