Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography.

C E Floyd1, R J Warp, J T Dobbins, H G Chotas, A H Baydush, R Vargas-Voracek, C E Ravin.   

Abstract

PURPOSE: To evaluate the imaging characteristics of an amorphous silicon flat-panel detector (FPD) for digital chest radiography.
MATERIALS AND METHODS: The 41 x 41-cm digital FPD is constructed on a single monolithic glass substrate with a structured cesium iodide scintillator layer and an amorphous silicon thin-film transistor array for image readout. Basic imaging characteristics of the FPD and associated image processing system were assessed on acquired images, including linearity, repeatability, uniformity of response, modulation transfer function (MTF), noise power spectrum, detective quantum efficiency (DQE), contrast sensitivity, and scatter content. Results with the FPD system were compared to those with a storage phosphor computed radiography (CR) system.
RESULTS: Images obtained with the FPD demonstrated excellent uniformity, repeatability, and linearity, as well as MTF and DQE that were superior to those with the storage phosphor CR system. The contrast and scatter content of images acquired with the FPD were equivalent to those acquired with the storage phosphor system.
CONCLUSION: The FPD provides radiographic images with excellent inherent physical image quality.

Entities:  

Mesh:

Year:  2001        PMID: 11230640     DOI: 10.1148/radiology.218.3.r01fe45683

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  18 in total

1.  Performance of a flat-panel detector in the detection of artificial erosive changes: comparison with conventional screen-film and storage-phosphor radiography.

Authors:  Karl Ludwig; Andreas Henschel; Thomas M Bernhardt; Horst Lenzen; Dag Wormanns; Stefan Diederich; Walter Heindel
Journal:  Eur Radiol       Date:  2002-11-29       Impact factor: 5.315

2.  Solid-state fluoroscopic imager for high-resolution angiography: physical characteristics of an 8 cm x 8 cm experimental prototype.

Authors:  Srinivasan Vedantham; Andrew Karellas; Sankararaman Suryanarayanan; Steven K Onishi
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

3.  Effects of dual-energy subtraction chest radiography on detection of small pulmonary nodules with varying attenuation: receiver operating characteristic analysis using a phantom study.

Authors:  Seitaro Oda; Kazuo Awai; Yoshinori Funama; Daisuke Utsunomiya; Yumi Yanaga; Koichi Kawanaka; Yasuyuki Yamashita
Journal:  Jpn J Radiol       Date:  2010-05-01       Impact factor: 2.374

4.  Dynamic dual-energy chest radiography: a potential tool for lung tissue motion monitoring and kinetic study.

Authors:  Tong Xu; Justin L Ducote; Jerry T Wong; Sabee Molloi
Journal:  Phys Med Biol       Date:  2011-02-01       Impact factor: 3.609

5.  Assessment of detective quantum efficiency: intercomparison of a recently introduced international standard with prior methods.

Authors:  Nicole T Ranger; Ehsan Samei; James T Dobbins; Carl E Ravin
Journal:  Radiology       Date:  2007-06       Impact factor: 11.105

6.  Rejection and redistribution of scattered radiation in scan equalization digital radiography (SEDR): simulation with spot images.

Authors:  Xinming Liu; Chris C Shaw
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

7.  Reduction of radiation dose for cerebral angiography using flat panel detector of direct conversion type: a vascular phantom study.

Authors:  Y Hatakeyama; S Kakeda; N Ohnari; J Moriya; N Oda; K Nishino; W Miyamoto; Y Korogi
Journal:  AJNR Am J Neuroradiol       Date:  2007-04       Impact factor: 3.825

8.  Scatter rejection and low-contrast performance of a slot-scan digital chest radiography system with electronic aft-collimation: a chest phantom study.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Mustafa C Altunbas; Lingyun Chen; Tao Han; Tianpeng Wang
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

Review 9.  Application of QC_DR software for acceptance testing and routine quality control of direct digital radiography systems: initial experiences using the Italian Association of Physicist in Medicine quality control protocol.

Authors:  Andrea Nitrosi; Marco Bertolini; Giovanni Borasi; Andrea Botti; Adriana Barani; Stefano Rivetti; Luisa Pierotti
Journal:  J Digit Imaging       Date:  2008-09-03       Impact factor: 4.056

10.  Effective DQE (eDQE) and speed of digital radiographic systems: an experimental methodology.

Authors:  Ehsan Samei; Nicole T Ranger; Alistair MacKenzie; Ian D Honey; James T Dobbins; Carl E Ravin
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.