[Evaluation of sphygmomanometers used by family physicians practicing outside the hospital environment in Bas-Saint-Laurent].

A Vanasse1, J Courteau.   

Abstract

OBJECTIVE: To assess the precision and integrity of all aneroid and mercury sphygmomanometers regularly used by family physicians practising outside hospitals.
DESIGN: Cross-sectional study.
SETTING: Private medical clinics and local community health centres in Bas-Saint-Laurent, Qué. PARTICIPANTS: A total of 151 of the 166 physicians in this administrative region. MAIN OUTCOME MEASURES: Precision of the mercury sphygmomanometers was measured using the difference between a reading in the absence of pressure and level 0. Precision of the aneroid sphygmomanometers was measured using variations at pressures of 140 mm Hg and 90 mm Hg compared with those on a calibrated mercury sphygmomanometer. Integrity of sphygmomanometers, arm cuffs, and inflating bulbs was also assessed.
RESULTS: In all, 258 sphygmomanometers met the inclusion criteria (111 mercury sphygmomanometers and 147 aneroid sphygmomanometers). Discrepancies of > or = 4 mm Hg were found in 15.5% of these instruments (12.6% and 17.7% of the mercury and aneroid sphygmomanometers, respectively). In 31.0% of the instruments (52.3% and 15.0% of the mercury and aneroid sphygmomanometers, respectively), one component was malfunctioning.
CONCLUSION: Sphygmomanometers that measure patients' blood pressure inaccurately could result in an incorrect diagnosis of hypertension or in a normal blood pressure reading in a hypertensive patient.

Entities:  

Mesh:

Year:  2001        PMID: 11228028      PMCID: PMC2016254     

Source DB:  PubMed          Journal:  Can Fam Physician        ISSN: 0008-350X            Impact factor:   3.275


  7 in total

1.  Aneroid sphygmomanometers. An assessment of accuracy at a university hospital and clinics.

Authors:  R H Bailey; V L Knaus; J H Bauer
Journal:  Arch Intern Med       Date:  1991-07

2.  How accurate are sphygmomanometers?

Authors:  D Mion; A M Pierin
Journal:  J Hum Hypertens       Date:  1998-04       Impact factor: 3.012

3.  Accuracy of prehospital sphygmomanometers.

Authors:  J S Jones; W Ramsey; T Hetrick
Journal:  J Emerg Med       Date:  1987       Impact factor: 1.484

4.  Clinical assessment of blood pressure.

Authors:  D W McKay; N R Campbell; L S Parab; A Chockalingam; J G Fodor
Journal:  J Hum Hypertens       Date:  1990-12       Impact factor: 3.012

5.  Guidelines for measurement of blood pressure, follow-up, and lifestyle counselling. Canadian Coalition for High Blood Pressure Prevention and Control.

Authors:  D Abbott; N Campbell; P Carruthers-Czyzewski; A Chockalingam; M David; G Dunkley; E Ellis; J G Fodor; D McKay; V R Ramsden
Journal:  Can J Public Health       Date:  1994 Sep-Oct

6.  Implementation of a province-wide computerized network in Quebec: the FAMUS Project.

Authors:  A Grant; E Delisle; S Dubois; T Niyonsenga; R Bernier
Journal:  MD Comput       Date:  1995 Jan-Feb

7.  Sphygmomanometers in hospital and family practice: problems and recommendations.

Authors:  M J Burke; H M Towers; K O'Malley; D J Fitzgerald; E T O'Brien
Journal:  Br Med J (Clin Res Ed)       Date:  1982-08-14
  7 in total
  1 in total

1.  Impact of terminal digit preference by family physicians and sphygmomanometer calibration errors on blood pressure value: implication for hypertension screening.

Authors:  Theophile Niyonsenga; Alain Vanasse; Josiane Courteau; Lyne Cloutier
Journal:  J Clin Hypertens (Greenwich)       Date:  2008-05       Impact factor: 3.738

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.