Literature DB >> 6809128

Sphygmomanometers in hospital and family practice: problems and recommendations.

M J Burke, H M Towers, K O'Malley, D J Fitzgerald, E T O'Brien.   

Abstract

The accuracy and working condition of 210 sphygmomanometers were tested: 100 (50 and mercury and 50 aneroid) models were used in family practices and 100 mercury models in hospitals. Faults in the inflation-deflation system were common and caused mainly by dirt or wear in the control valves. Leakage occurred in 48% of the hospital and 33% of the family practice sphygmomanometers. In the mercury models the mercury or air vents were often in an unsatisfactory condition or the calibrated glass tube dirty. The accuracy of the gauges was examined at 90 and 150 mm Hg: fewer than 2% of the mercury sphygmomanometers but 30% of the aneroid models had errors greater than +/- 4 mm Hg at either pressure. Over half of the cuffs examined had bladders widths less than the recommended size, and 94% had bladders shorter than the length recommended for use on normal adults. Mercury sphygmomanometers should be bought in preference to aneroid models as they are more accurate, less expensive in the long term, and can be maintained by the owner; they should be checked every six to 12 months depending on usage. Replacement parts should be kept readily available.

Entities:  

Mesh:

Substances:

Year:  1982        PMID: 6809128      PMCID: PMC1499274          DOI: 10.1136/bmj.285.6340.469

Source DB:  PubMed          Journal:  Br Med J (Clin Res Ed)        ISSN: 0267-0623


  10 in total

1.  SPHYGOMOMANOMETER CUFF SIZE AND THE ACCURACY OF INDIRECT MEASUREMENT OF BLOOD PRESSURE.

Authors:  M J KARVONEN; L J TELIVUO; E J JAERVINEN
Journal:  Am J Cardiol       Date:  1964-05       Impact factor: 2.778

2.  EFFECT OF SIZE OF CUFF BLADDER ON ACCURACY OF MEASUREMENT OF INDIRECT BLOOD PRESSURE.

Authors:  J A SIMPSON; G JAMIESON; D W DICKHAUS; R F GROVER
Journal:  Am Heart J       Date:  1965-08       Impact factor: 4.749

3.  Sphygmomanometers: errors due to blocked vents.

Authors:  A Shaw; C Deehan; J M Lenihan
Journal:  Br Med J       Date:  1979-03-24

4.  Accuracy of sphygmomanometers in hospital practice.

Authors:  L V Perlman; B N Chiang; J Keller; H Blackburn
Journal:  Arch Intern Med       Date:  1970-06

5.  Taking the blood pressure.

Authors:  G E King
Journal:  JAMA       Date:  1969-09-22       Impact factor: 56.272

6.  Blood pressure errors with aneroid sphygmomanometers.

Authors:  C E Bowman
Journal:  Lancet       Date:  1981-05-02       Impact factor: 79.321

7.  Management of high blood pressure in general practice.

Authors:  M O'Hanrahan; M Laher; C O'Boyle; K O'Malley; E T O'Brien
Journal:  Ir Med J       Date:  1982-04

8.  Isolation and activation of inactive renin from human kidney and plasma. Plasma and renal inactive renins have different molecular weights.

Authors:  J J Chang; M Kisaragi; H Okamoto; T Inagami
Journal:  Hypertension       Date:  1981 Sep-Oct       Impact factor: 10.190

9.  The sphygmomanometer.

Authors:  E T O'Brien; K O'Malley
Journal:  Br Med J       Date:  1979-10-06

10.  Defects in sphygmomanometers: an important source of error in blood pressure recording.

Authors:  S Conceiçao; M K Ward; D N Kerr
Journal:  Br Med J       Date:  1976-04-10
  10 in total
  27 in total

Review 1.  ABC of hypertension: Blood pressure measurement. Part II-conventional sphygmomanometry: technique of auscultatory blood pressure measurement.

Authors:  G Beevers; G Y Lip; E O'Brien
Journal:  BMJ       Date:  2001-04-28

2.  Blood pressure measuring devices: recommendations of the European Society of Hypertension.

Authors:  E O'Brien; B Waeber; G Parati; J Staessen; M G Myers
Journal:  BMJ       Date:  2001-03-03

3.  Unvalidated blood pressure devices with small cuffs are being used in hospitals.

Authors:  M Thomas; T Radford; I Dasgupta
Journal:  BMJ       Date:  2001-08-18

4.  The extent of inaccurate aneroid sphygmomanometers in a hospital setting.

Authors:  Dimitri A Cozanitis; Christopher J Jones
Journal:  Wien Med Wochenschr       Date:  2010-07

5.  Understanding and managing hypertension in an African sub-district: a multidisciplinary approach.

Authors:  Margaret Thorogood; Myles D Connor; Gillian Lewando Hundt; Stephen M Tollman
Journal:  Scand J Public Health Suppl       Date:  2007-08       Impact factor: 3.021

6.  Blood pressure measurement by family physicians.

Authors:  R C Dickson; N J Birkett
Journal:  Can Fam Physician       Date:  1988-02       Impact factor: 3.275

Review 7.  Accurate, reproducible measurement of blood pressure.

Authors:  N R Campbell; A Chockalingam; J G Fodor; D W McKay
Journal:  CMAJ       Date:  1990-07-01       Impact factor: 8.262

8.  Ankle blood pressure measured by automatic oscillotonometry: a comparison with Doppler pressure measurements.

Authors:  M Adiseshiah; F W Cross; P A Belsham
Journal:  Ann R Coll Surg Engl       Date:  1987-11       Impact factor: 1.891

Review 9.  Blood pressure measurement: current practice and future trends.

Authors:  E O'Brien; D Fitzgerald; K O'Malley
Journal:  Br Med J (Clin Res Ed)       Date:  1985-03-09

10.  Report of the Canadian Hypertension Society's consensus conference on the management of mild hypertension.

Authors:  A G Logan
Journal:  Can Med Assoc J       Date:  1984-11-01       Impact factor: 8.262

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.