How accurate are sphygmomanometers?

D Mion1, A M Pierin.   

Abstract

The objective of this study was to assess the accuracy and reliability of mercury and aneroid sphygmomanometers. Measurement of accuracy of calibration and evaluation of physical conditions were carried out in 524 sphygmomanometers, 351 from a hospital setting, and 173 from private medical offices. Mercury sphygmomanometers were considered inaccurate if the meniscus was not '0' at rest. Aneroid sphygmomanometers were tested against a properly calibrated mercury manometer, and were considered calibrated when the error was < or =3 mm Hg. Both types of sphygmomanometers were evaluated for conditions of cuff/bladder, bulb, pump and valve. Of the mercury sphygmomanometers tested 21 % were found to be inaccurate. Of this group, unreliability was noted due to: excessive bouncing (14%), illegibility of the gauge (7%), blockage of the filter (6%), and lack of mercury in the reservoir (3%). Bladder damage was noted in 10% of the hospital devices and in 6% of private medical practices. Rubber aging occurred in 34% and 25%, leaks/holes in 19% and 18%, and leaks in the pump bulb in 16% and 30% of hospital devices and private practice devices, respectively. Of the aneroid sphygmomanometers tested, 44% in the hospital setting and 61% in private medical practices were found to be inaccurate. Of these, the magnitude of inaccuracy was 4-6 mm Hg in 32%, 7-12 mm Hg in 19% and > 13 mm Hg in 7%. In summary, most of the mercury and aneroid sphygmomanometers showed inaccuracy (21% vs 58%) and unreliability (64% vs 70%).

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9607693     DOI: 10.1038/sj.jhh.1000589

Source DB:  PubMed          Journal:  J Hum Hypertens        ISSN: 0950-9240            Impact factor:   3.012


  14 in total

1.  The extent of inaccurate aneroid sphygmomanometers in a hospital setting.

Authors:  Dimitri A Cozanitis; Christopher J Jones
Journal:  Wien Med Wochenschr       Date:  2010-07

2.  [Evaluation of sphygmomanometers used by family physicians practicing outside the hospital environment in Bas-Saint-Laurent].

Authors:  A Vanasse; J Courteau
Journal:  Can Fam Physician       Date:  2001-02       Impact factor: 3.275

Review 3.  Principles and techniques of blood pressure measurement.

Authors:  Gbenga Ogedegbe; Thomas Pickering
Journal:  Cardiol Clin       Date:  2010-11       Impact factor: 2.213

4.  Use of mercury-based medical equipment and mercury content in effluents of tertiary care hospitals in India.

Authors:  Sharda Shah Peshin; Nabanita Halder; Chandrababu Jathikarta; Yogendra Kumar Gupta
Journal:  Environ Monit Assess       Date:  2015-02-26       Impact factor: 2.513

5.  Knowledge, attitude, behavior, and practice of interns toward errors in sphygmomanometer and blood pressure measurement.

Authors:  Ramachandran Meenakshisundaram; Shah Sweni; Govindaraj Dhanalakshmi; Raju Sridevi; Ponniah Thirumalaikolundusubramanian
Journal:  J Clin Hypertens (Greenwich)       Date:  2009-12       Impact factor: 3.738

6.  Italian society of hypertension guidelines for conventional and automated blood pressure measurement in the office, at home and over 24 hours.

Authors:  Gianfranco Parati; Stefano Omboni; Paolo Palatini; Damiano Rizzoni; Grzegorz Bilo; Mariaconsuelo Valentini; Enrico Agabiti Rosei; Giuseppe Mancia
Journal:  High Blood Press Cardiovasc Prev       Date:  2013-01-22

7.  Beyond the Evidence of the New Hypertension Guidelines. Blood pressure measurement - is it good enough for accurate diagnosis of hypertension? Time might be in, for a paradigm shift (I).

Authors:  Cornel Pater
Journal:  Curr Control Trials Cardiovasc Med       Date:  2005-04-06

8.  Impact of terminal digit preference by family physicians and sphygmomanometer calibration errors on blood pressure value: implication for hypertension screening.

Authors:  Theophile Niyonsenga; Alain Vanasse; Josiane Courteau; Lyne Cloutier
Journal:  J Clin Hypertens (Greenwich)       Date:  2008-05       Impact factor: 3.738

9.  Office blood pressure measurement practices among community health providers (medical and paramedical) in northern district of India.

Authors:  Bishav Mohan; Naved Aslam; Upma Ralhan; Sarit Sharma; Naveen Gupta; Vivudh Pratap Singh; Shibba Takkar; G S Wander
Journal:  Indian Heart J       Date:  2014-08-06

10.  Equipment errors: a prevalent cause for fallacy in blood pressure recording - a point prevalence estimate from an Indian health university.

Authors:  Badrinarayan Mishra; Nidhi Dinesh Sinha; Hitesh Gidwani; Sushil Kumar Shukla; Abhishek Kawatra; Sc Mehta
Journal:  Indian J Community Med       Date:  2013-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.