Literature DB >> 11224784

The Nucleus Contour electrode array: a radiological and histological study.

B Richter1, A Aschendorff, P Lohnstein, H Husstedt, H Nagursky, R Laszig.   

Abstract

OBJECTIVES: To evaluate the handling and insertion trauma of the recently developed Nucleus perimodiolar Contour electrode array (Cochlear Ltd., Pty, Lane Cove, New South Wales, Australia) in human temporal bones compared with the Nucleus standard straight electrode array. STUDY
DESIGN: E-perimental control group.
METHODS: Twenty-nine fresh-frozen bones were implanted with different electrode arrays by an experienced cochlear implant surgeon, and evaluated both radiologically and histologically.
RESULTS: Intracochlear insertion of the standard Nucleus straight electrode array was found to be atraumatic, confirming previous findings in the literature. Insertion of the Nucleus Contour electrode array resulted in instances of localized basilar membrane penetration causing the electrode array to move from the scala tympani into the scala vestibuli. However, this trauma did not result in any observable damage to the osseous spiral lamina or the modiolus. Basilar membrane penetration was observed in six of eight cochlear bones when a standard cochleostomy size (approximately 0.8 mm) and site (anterior and superior to the round window) were used. However, when the surgical technique was modified to use a slightly larger cochleostomy ( approximately 1.8 mm) situated closer to the round window and employ a partial stylet withdrawal technique during electrode insertion, the frequency of penetrations was restricted to two of seven bones. This trauma rate is comparable to that observed with other cochlear implants designs.
CONCLUSIONS: Following our results, the design of the Nucleus Contour electrode appears to fulfill the safety requirements for an intracochlear electrode array, provided that the surgical insertion technique is modified in the manner outlined.

Entities:  

Mesh:

Year:  2001        PMID: 11224784     DOI: 10.1097/00005537-200103000-00023

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  14 in total

Review 1.  [Current developments in cochlear implantation].

Authors:  R Laszig; A Aschendorff; J Schipper; T Klenzner
Journal:  HNO       Date:  2004-04       Impact factor: 1.284

2.  Impact of the surgical experience on cochleostomy location: a comparative temporal bone study between endaural and posterior tympanotomy approaches for cochlear implantation.

Authors:  Clair Vandersteen; Thomas Demarcy; Coralie Roger; Eric Fontas; Charles Raffaelli; Nicholas Ayache; Hervé Delingette; Nicolas Guevara
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-10-16       Impact factor: 2.503

Review 3.  The development of the Nucleus Freedom Cochlear implant system.

Authors:  James F Patrick; Peter A Busby; Peter J Gibson
Journal:  Trends Amplif       Date:  2006-12

4.  Rotational tomography of the normal and reconstructed middle ear in temporal bones: an experimental study.

Authors:  Ch Offergeld; J Kromeier; A Aschendorff; W Maier; Th Klenzner; Th Beleites; Th Zahnert; J Schipper; R Laszig
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-10-18       Impact factor: 2.503

5.  Intracochlear assessment of electrode position after cochlear implant surgery by means of multislice computer tomography.

Authors:  Gijs K A van Wermeskerken; Mathias Prokop; Adriaan F van Olphen; Frans W J Albers
Journal:  Eur Arch Otorhinolaryngol       Date:  2007-07-14       Impact factor: 2.503

6.  Considerations for design of future cochlear implant electrode arrays: electrode array stiffness, size, and depth of insertion.

Authors:  Stephen J Rebscher; Alexander Hetherington; Ben Bonham; Peter Wardrop; David Whinney; Patricia A Leake
Journal:  J Rehabil Res Dev       Date:  2008

7.  Partial withdrawal of deeply inserted cochlear electrodes: observations of two patients.

Authors:  Maria-Izabel Kos; Colette Boex; Jean-Philippe Guyot; Marco Pelizzone
Journal:  Eur Arch Otorhinolaryngol       Date:  2007-06-12       Impact factor: 2.503

8.  [Insertion results for Contour™ and Contour Advance™ electrodes: are there individual learning curves?].

Authors:  A Aschendorff; T Klenzner; S Arndt; R Beck; C Schild; L Röddiger; W Maier; R Laszig
Journal:  HNO       Date:  2011-05       Impact factor: 1.284

Review 9.  Cochlear Implants and Children with Vestibular Impairments.

Authors:  Sharon L Cushing; Blake C Papsin
Journal:  Semin Hear       Date:  2018-07-20

Review 10.  Cochlear implants: system design, integration, and evaluation.

Authors:  Fan-Gang Zeng; Stephen Rebscher; William Harrison; Xiaoan Sun; Haihong Feng
Journal:  IEEE Rev Biomed Eng       Date:  2008-11-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.