What is the optimum leaf width of a multileaf collimator?

T Bortfeld1, U Oelfke, S Nill.   

Abstract

UNLABELLED: The following question is investigated: How narrow do the leaves of a multileaf collimator have to be such that further reduction of the leaf width does not lead to physical improvements of the dose distribution. Because of the physical principles of interaction between radiation and matter, dose distributions in radiotherapy are generally relatively smooth. According to the theory of sampling, the dose distribution can therefore be represented by a set of evenly spaced samples. The distance between the samples is identified with the distance between the leaf centers of a multileaf collimator. The optimum sampling distance is derived from the 20% to 80% field edge penumbra through the concept of the dose deposition kernel, which is approximated by a Gaussian. The leaf width of the multileaf collimator is considered to be independent from the sampling distance. Two cases are studied in detail: (i) the leaf width equals the sampling distance, which is the regular case, and (ii) the leaf width is twice the sampling distance. The practical delivery of the latter treatment geometry requires a couch movement or a collimator rotation. The optimum sampling distance equals the 20%-80% penumbra divided by 1.7 and is on the order of 1.5-2 mm for a typical 6 MV beam in soft tissue. The optimum leaf width equals this sampling distance in the regular case. A relatively small deterioration results if the leaf width is doubled, while the sampling distance remains the same. The deterioration can be corrected for by deconvolving the fluence profile with an inverse filter.
CONCLUSIONS: With the help of the sampling theory and, more generally, the theory of linear systems, one can find a general answer to the question about the optimum leaf width of a multileaf collimator from a physical point of view. It is important to distinguish between the sampling distance and the leaf width. The sampling distance is more critical than the leaf width. The leaf width can be up to twice as large as the sampling width. Furthermore, the derived sampling distance can be used to select the optimum resolution of both the fluence and the dose grid in dose calculation and inverse planning algorithms.

Mesh:

Year:  2000        PMID: 11128300     DOI: 10.1118/1.1319524

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Multicriteria VMAT optimization.

Authors:  David Craft; Dualta McQuaid; Jeremiah Wala; Wei Chen; Ehsan Salari; Thomas Bortfeld
Journal:  Med Phys       Date:  2012-02       Impact factor: 4.071

2.  Impact of the high-definition multileaf collimator on linear accelerator-based intracranial stereotactic radiosurgery.

Authors:  J A Tanyi; C M Kato; Y Chen; Z Chen; M Fuss
Journal:  Br J Radiol       Date:  2010-10-05       Impact factor: 3.039

3.  Improving IMRT-plan quality with MLC leaf position refinement post plan optimization.

Authors:  Ying Niu; Guowei Zhang; Barry L Berman; William C Parke; Byongyong Yi; Cedric X Yu
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

4.  The number of beams in IMRT--theoretical investigations and implications for single-arc IMRT.

Authors:  Thomas Bortfeld
Journal:  Phys Med Biol       Date:  2010-01-07       Impact factor: 3.609

5.  Dynamic conformal arc cranial stereotactic radiosurgery: implications of multileaf collimator margin on dose-volume metrics.

Authors:  J A Tanyi; E J Doss; C M Kato; D L Monaco; L ZMeng; Y Chen; C D Kubicky; C M Marquez; M Fuss
Journal:  Br J Radiol       Date:  2012-11       Impact factor: 3.039

Review 6.  Stereotactic radiosurgery for treatment of brain metastases. A report of the DEGRO Working Group on Stereotactic Radiotherapy.

Authors:  Martin Kocher; Andrea Wittig; Marc Dieter Piroth; Harald Treuer; Heinrich Seegenschmiedt; Maximilian Ruge; Anca-Ligia Grosu; Matthias Guckenberger
Journal:  Strahlenther Onkol       Date:  2014-04-09       Impact factor: 3.621

7.  New multileaf collimator with a leaf width of 5 mm improves plan quality compared to 10 mm in step-and-shoot IMRT of HNC using integrated boost procedure.

Authors:  Felix Zwicker; Henrik Hauswald; Simeon Nill; Bernhard Rhein; Christian Thieke; Falk Roeder; Carmen Timke; Angelika Zabel-du Bois; Jürgen Debus; Peter E Huber
Journal:  Strahlenther Onkol       Date:  2010-05-21       Impact factor: 3.621

8.  Intracranial stereotactic radiosurgery with an adapted linear accelerator vs. robotic radiosurgery: Comparison of dosimetric treatment plan quality.

Authors:  Harald Treuer; Moritz Hoevels; Klaus Luyken; Veerle Visser-Vandewalle; Jochen Wirths; Martin Kocher; Maximilian Ruge
Journal:  Strahlenther Onkol       Date:  2014-11-22       Impact factor: 3.621

9.  Implications of a high-definition multileaf collimator (HD-MLC) on treatment planning techniques for stereotactic body radiation therapy (SBRT): a planning study.

Authors:  James A Tanyi; Paige A Summers; Charles L McCracken; Yiyi Chen; Li-Chung Ku; Martin Fuss
Journal:  Radiat Oncol       Date:  2009-07-10       Impact factor: 3.481

10.  Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets.

Authors:  S A Yoganathan; Karthick Raj Mani; K J Maria Das; Arpita Agarwal; Shaleen Kumar
Journal:  J Med Phys       Date:  2011-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.