| Literature DB >> 19591687 |
James A Tanyi1, Paige A Summers, Charles L McCracken, Yiyi Chen, Li-Chung Ku, Martin Fuss.
Abstract
PURPOSE: To assess the impact of two multileaf collimator (MLC) systems (2.5 and 5 mm leaf widths) on three-dimensional conformal radiotherapy, intensity-modulated radiotherapy, and dynamic conformal arc techniques for stereotactic body radiation therapy (SBRT) of liver and lung lesions.Entities:
Mesh:
Year: 2009 PMID: 19591687 PMCID: PMC2716348 DOI: 10.1186/1748-717X-4-22
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Isodose distributions and DVHs for a lung lesion generated from three different planning techniques and two MLC systems. A1 through A6 are axial isodose distribution corresponding to M120 MLC IMRT, M120 MLC 3DCRT, M120 MLC DCA, HD120 MLC IMRT, HD120 MLC 3DCRT, and HD120 MLC DCA plans, respectively.
Mean dose-spillage volume, normalized to PTV.p-values of the paired t-test included to assess difference between MLC systems.
| Technique | VHS | VIS | VLS | |||
| M120 | HD120 | M120 | HD120 | M120 | HD120 | |
| IMRT | 0.54 ± 0.30 | 0.50 ± 0.25 | 3.86 ± 1.38 | 3.66 ± 1.22 | 23.69 ± 9.21 | 23.14 ± 8.75 |
| 3DCRT | 0.47 ± 0.13 | 0.44 ± 0.10 | 4.08 ± 1.34 | 3.93 ± 1.12 | 23.64 ± 7.70 | 23.36 ± 7.70 |
| p = 0.04 | p = 0.24 | p = 0.01 | ||||
| DCA | 0.44 ± 0.13 | 0.43 ± 0.12 | 3.26 ± 0.61 | 3.19 ± 0.60 | 15.32 ± 4.36 | 14.76 ± 4.23 |
Cases where performance of HD120 MLC surpassed that of M120 MLC.
| Technique | CI | HI | VHS | VIS | VLS |
| IMRT | 15 (51.7%) | 22 (75.9%) | 19 (65.5%) | 24 (82.8%) | 22 (75.9%) |
| 3DCRT | 18 (62.1%) | 15 (51.7%) | 21 (72.4%) | 21 (72.4%) | 25 (86.2%) |
| DCA | 16 (55.2%) | 16 (55.2%) | 20 (69.0%) | 23 (79.3%) | 23 (79.3%) |
The values in the table are presented as the number of cases and their corresponding ratio (as a percentage) over the 29 patient cases assessed in the current study.
Mean (top) and max (bottom) percent improvement or worsening of HD120 MLC plans over M120 MLC plans.
| Technique | Improvement (%) | Worsening (%) | ||||||
| CI | VHS | VIS | VLS | CI | VHS | VIS | VLS | |
| IMRT | 3.9 | 4.6 | 5.5 | 3.5 | 2.1 | 3.1 | 8.5 | 5.1 |
| 3DCRT | 2.5 | 2.5 | 4.6 | 1.8 | 2.2 | 2.0 | 5.8 | 2.6 |
| DCA | 2.4 | 2.2 | 3.0 | 3.3 | 2.7 | 2.7 | 4.0 | 5.1 |
| IMRT | 18.5 | 20.4 | 26.5 | 22.7 | 10.4 | 17.7 | 27.6 | 14.0 |
| 3DCRT | 9.5 | 9.8 | 39.4 | 3.7 | 13.2 | 8.7 | 25.9 | 3.7 |
| DCA | 8.1 | 6.4 | 9.4 | 9.6 | 13.2 | 9.8 | 12.1 | 34.3 |
Mean number of monitor units (within one standard deviation) necessary to deliver one centigray of prescribed dose for different treatment plan categories.
| Technique | IMRT | 3DCRT | DCA | |||
| M120 | HD120 | M120 | HD120 | M120 | HD120 | |
| MU/cGy | 3.45 ± 1.06 | 3.63 ± 1.36 | 2.25 ± 0.54 | 2.26 ± 0.54 | 2.24 ± 0.54 | 2.28 ± 0.56 |