Intracranial stereotactic radiosurgery with an adapted linear accelerator vs. robotic radiosurgery: Comparison of dosimetric treatment plan quality.

Harald Treuer1, Moritz Hoevels, Klaus Luyken, Veerle Visser-Vandewalle, Jochen Wirths, Martin Kocher, Maximilian Ruge.   

Abstract

BACKGROUND AND
PURPOSE: Stereotactic radiosurgery with an adapted linear accelerator (linac-SRS) is an established therapy option for brain metastases, benign brain tumors, and arteriovenous malformations. We intended to investigate whether the dosimetric quality of treatment plans achieved with a CyberKnife (CK) is at least equivalent to that for linac-SRS with circular or micromultileaf collimators (microMLC). PATIENTS AND METHODS: A random sample of 16 patients with 23 target volumes, previously treated with linac-SRS, was replanned with CK. Planning constraints were identical dose prescription and clinical applicability. In all cases uniform optimization scripts and inverse planning objectives were used. Plans were compared with respect to coverage, minimal dose within target volume, conformity index, and volume of brain tissue irradiated with ≥ 10 Gy.
RESULTS: Generating the CK plan was unproblematic with simple optimization scripts in all cases. With the CK plans, coverage, minimal target volume dosage, and conformity index were significantly better, while no significant improvement could be shown regarding the 10 Gy volume. Multiobjective comparison for the irradiated target volumes was superior in the CK plan in 20 out of 23 cases and equivalent in 3 out of 23 cases. Multiobjective comparison for the treated patients was superior in the CK plan in all 16 cases.
CONCLUSION: The results clearly demonstrate the superiority of the irradiation plan for CK compared to classical linac-SRS with circular collimators and microMLC. In particular, the average minimal target volume dose per patient, increased by 1.9 Gy, and at the same time a 14% better conformation index seems to be an improvement with clinical relevance.

Entities:  

Mesh:

Year:  2014        PMID: 25416146     DOI: 10.1007/s00066-014-0786-y

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  35 in total

1.  Dosimetric comparison of CyberKnife with other radiosurgical modalities for an ellipsoidal target.

Authors:  Cheng Yu; Gabor Jozsef; Michael L J Apuzzo; Zbigniew Petrovich
Journal:  Neurosurgery       Date:  2003-11       Impact factor: 4.654

2.  Risk analysis of linear accelerator radiosurgery.

Authors:  J Voges; H Treuer; V Sturm; C Büchner; R Lehrke; M Kocher; S Staar; J Kuchta; R P Müller
Journal:  Int J Radiat Oncol Biol Phys       Date:  1996-12-01       Impact factor: 7.038

3.  A conformation number to quantify the degree of conformality in brachytherapy and external beam irradiation: application to the prostate.

Authors:  A van't Riet; A C Mak; M A Moerland; L H Elders; W van der Zee
Journal:  Int J Radiat Oncol Biol Phys       Date:  1997-02-01       Impact factor: 7.038

Review 4.  Stereotactic radiosurgery for treatment of brain metastases. A report of the DEGRO Working Group on Stereotactic Radiotherapy.

Authors:  Martin Kocher; Andrea Wittig; Marc Dieter Piroth; Harald Treuer; Heinrich Seegenschmiedt; Maximilian Ruge; Anca-Ligia Grosu; Matthias Guckenberger
Journal:  Strahlenther Onkol       Date:  2014-04-09       Impact factor: 3.621

5.  Impact of target point deviations on control and complication probabilities in stereotactic radiosurgery of AVMs and metastases.

Authors:  Harald Treuer; Martin Kocher; Moritz Hoevels; Stefan Hunsche; Klaus Luyken; Mohammad Maarouf; Jürgen Voges; Rolf-Peter Müller; Volker Sturm
Journal:  Radiother Oncol       Date:  2006-09-26       Impact factor: 6.280

6.  Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife ®) stereotactic system plans for acoustic schwannoma.

Authors:  Debnarayan Dutta; S Balaji Subramanian; V Murli; H Sudahar; P G Gopalakrishna Kurup; Mahadev Potharaju
Journal:  J Neurooncol       Date:  2011-09-04       Impact factor: 4.130

7.  Retrospective analysis of linac-based radiosurgery for arteriovenous malformations and testing of the Flickinger formula in predicting radiation injury.

Authors:  I A Cetin; R Ates; J Dhaens; G Storme
Journal:  Strahlenther Onkol       Date:  2012-11-07       Impact factor: 3.621

8.  Robotic radiosurgery versus micro-multileaf collimator: a dosimetric comparison for large or critically located arteriovenous malformations.

Authors:  Sławomir Blamek; Aleksandra Grządziel; Leszek Miszczyk
Journal:  Radiat Oncol       Date:  2013-08-23       Impact factor: 3.481

9.  Neurosymptomatic carvenous sinus meningioma: a 15-years experience with fractionated stereotactic radiotherapy and radiosurgery.

Authors:  Sebastião Francisco Miranda Correa; Gustavo Nader Marta; Manoel Jacobsen Teixeira
Journal:  Radiat Oncol       Date:  2014-01-17       Impact factor: 3.481

10.  Risk-adapted single or fractionated stereotactic high-precision radiotherapy in a pooled series of nonfunctioning pituitary adenomas: high local control and low toxicity.

Authors:  Jan Patrick Boström; Almuth Meyer; Bogdan Pintea; Rüdiger Gerlach; Gunnar Surber; Guido Lammering; Klaus Hamm
Journal:  Strahlenther Onkol       Date:  2014-08-05       Impact factor: 3.621

View more
  20 in total

1.  Dosimetrics of intracranial stereotactic radiosurgery: Only "an exercise of style"?

Authors:  Alba Fiorentino; Niccolò Giaj-Levra; Rosario Mazzola; Sergio Fersino; Francesco Ricchetti; Filippo Alongi
Journal:  Strahlenther Onkol       Date:  2015-08-08       Impact factor: 3.621

2.  Authors' reply to "Dosimetric of intracranial stereotactic radiosurgery: only 'an exercise of style'".

Authors:  Harald Treuer; Moritz Hoevels; Klaus Luyken; Veerle Visser-Vandewalle; Jochen Wirths; Martin Kocher; Maximilian Ruge
Journal:  Strahlenther Onkol       Date:  2015-10       Impact factor: 3.621

3.  Hypofractionated stereotactic radiotherapy for brain metastases from lung cancer : Evaluation of indications and predictors of local control.

Authors:  Takeaki Ishihara; Kazunari Yamada; Aya Harada; Kenta Isogai; Yoshihiro Tonosaki; Yusuke Demizu; Daisuke Miyawaki; Kenji Yoshida; Yasuo Ejima; Ryohei Sasaki
Journal:  Strahlenther Onkol       Date:  2016-05-11       Impact factor: 3.621

4.  Predictors for occlusion of cerebral AVMs following radiation therapy : Radiation dose and prior embolization, but not Spetzler-Martin grade.

Authors:  Stefan Knippen; Florian Putz; Sabine Semrau; Ulrike Lambrecht; Arzu Knippen; Michael Buchfelder; Sven Schlaffer; Tobias Struffert; Rainer Fietkau
Journal:  Strahlenther Onkol       Date:  2016-10-18       Impact factor: 3.621

Review 5.  [Computer-controlled high-precision radiation].

Authors:  D Rueß; M Kocher; H Treuer; M I Ruge
Journal:  HNO       Date:  2017-01       Impact factor: 1.284

6.  Quality of life in patients with limited (1-3) brain metastases undergoing stereotactic or whole brain radiotherapy : A prospective study of the DEGRO QoL working group.

Authors:  Diana Steinmann; Dirk Vordermark; Wiebke Gerstenberg; Raimund Aschoff; Nadine Gharbi; Axel Müller; Christof Schäfer; Marilena Theodorou; Hans-Joachim Wypior; Hans Geinitz
Journal:  Strahlenther Onkol       Date:  2019-08-15       Impact factor: 3.621

7.  Iodine-125 brachytherapy as upfront and salvage treatment for brain metastases : A comparative analysis.

Authors:  Alexander Romagna; Christoph Schwartz; Rupert Egensperger; Juliana Watson; Jörg-Christian Tonn; Claus Belka; Friedrich-Wilhelm Kreth; Silke Birgit Nachbichler
Journal:  Strahlenther Onkol       Date:  2016-06-27       Impact factor: 3.621

8.  Stereotactic Radiosurgery of Cavernous Sinus Meningiomas.

Authors:  Daniel Rueß; Fenja Fritsche; Stefan Grau; Harald Treuer; Mauritius Hoevels; Martin Kocher; Christian Baues; Maximilian I Ruge
Journal:  J Neurol Surg B Skull Base       Date:  2019-03-12

9.  Radiosurgery with flattening-filter-free techniques in the treatment of brain metastases : Plan comparison and early clinical evaluation.

Authors:  J Rieber; E Tonndorf-Martini; O Schramm; B Rhein; S Stefanowicz; J Kappes; H Hoffmann; K Lindel; J Debus; S Rieken
Journal:  Strahlenther Onkol       Date:  2016-07-04       Impact factor: 3.621

10.  Tumor-dose-rate variations during robotic radiosurgery of oligo and multiple brain metastases.

Authors:  Maria-Lisa Wilhelm; Mark K H Chan; Benedikt Abel; Florian Cremers; Frank-Andre Siebert; Stefan Wurster; David Krug; Robert Wolff; Jürgen Dunst; Guido Hildebrandt; Achim Schweikard; Dirk Rades; Floris Ernst; Oliver Blanck
Journal:  Strahlenther Onkol       Date:  2020-06-25       Impact factor: 3.621

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.