| Literature DB >> 10835020 |
A M Gronowski1, S Copper, D Baorto, P R Murray.
Abstract
This study demonstrates that significant reproducibility problems can occur during routine use of the Abbott Laboratories LCx assay for Chlamydia trachomatis and Neisseria gonorrhoeae. These problems can go undetected by the quality control procedures outlined in the manufacturer's package insert. We outline here procedures for detecting and preventing contamination and reproducibility problems.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10835020 PMCID: PMC86828 DOI: 10.1128/JCM.38.6.2416-2418.2000
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948