Beat-to-beat measurement of cardiac output by intravascular pulse contour analysis: a prospective criterion standard study in patients after cardiac surgery.

C Zöllner1, M Haller, M Weis, K Mörstedt, P Lamm, E Kilger, A E Goetz.   

Abstract

OBJECTIVE: To evaluate the accuracy of a new pulse contour method of measuring cardiac output in critically ill patients.
DESIGN: A prospective criterion standard study.
SETTING: Cardiac surgery intensive care unit in a university hospital. PARTICIPANTS: Nineteen cardiac surgery patients requiring intensive care treatment with pulmonary artery catheters after surgery.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: The pulse contour cardiac output monitor uses transpulmonary bolus thermodilution measurements to calibrate the system. In each patient, the pulse contour cardiac output values were compared with conventional thermodilution. The method described by Bland and Altman and linear regression analysis were used for comparison. The mean difference (bias) +/- standard deviation of differences (precision) was 0.31 +/- 1.25 L/min for pulmonary bolus thermodilution cardiac output versus pulse contour cardiac output and 0.21 +/- 0.73 L/min for pulmonary bolus thermodilution cardiac output versus transpulmonary bolus thermodilution cardiac output. Linear regression (correlation) analyses were pulse contour cardiac output = 0.97 thermodilution + 0.53 (r = 0.88), and transpulmonary cardiac output = 0.87 thermodilution + 1.09 (r = 0.96). There was a small increase 60 minutes after recalibration but not a statistically significant difference between pulse contour cardiac output and pulmonary bolus thermodilution cardiac output (p = 0.52).
CONCLUSIONS: Bias and precision are acceptable, and the system provides results that agree with conventional thermodilution. This study demonstrates the clinical applicability of the pulse contour cardiac output monitoring system.

Entities:  

Mesh:

Year:  2000        PMID: 10794327     DOI: 10.1016/s1053-0770(00)90003-x

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  22 in total

1.  Comparison between cardiac output measured by the pulmonary arterial thermodilution technique and that measured by the femoral arterial thermodilution technique in a pediatric animal model.

Authors:  M Rupérez; J López-Herce; C García; C Sánchez; E García; D Vigil
Journal:  Pediatr Cardiol       Date:  2003-12-23       Impact factor: 1.655

2.  Correlation between cardiac output measured by the femoral arterial thermodilution technique pulmonary arterial and that measured by contour pulse analysis in a paediatric animal model.

Authors:  Jesús López-Herce; Marta Rupérez; César Sánchez; Cristina García; Elena García
Journal:  J Clin Monit Comput       Date:  2006-02-28       Impact factor: 2.502

3.  Reliability of continuous pulse contour cardiac output measurement during hemodynamic instability.

Authors:  Anders Johansson; Michelle Chew
Journal:  J Clin Monit Comput       Date:  2007-06-01       Impact factor: 2.502

Review 4.  [Haemodynamic monitoring in the perioperative phase. Available systems, practical application and clinical data].

Authors:  U Wittkowski; C Spies; M Sander; J Erb; A Feldheiser; C von Heymann
Journal:  Anaesthesist       Date:  2009-08       Impact factor: 1.041

5.  Ability of pulse contour and esophageal Doppler to estimate rapid changes in stroke volume.

Authors:  Scott R Gunn; Hyung Kook Kim; Peter W J Harrigan; Michael R Pinsky
Journal:  Intensive Care Med       Date:  2006-07-28       Impact factor: 17.440

6.  Dynamic device properties of pulse contour cardiac output during transcatheter aortic valve implantation.

Authors:  Martin Petzoldt; Carsten Riedel; Jan Braeunig; Sebastian Haas; Matthias S Goepfert; Hendrik Treede; Stephan Baldus; Alwin E Goetz; Daniel A Reuter
Journal:  J Clin Monit Comput       Date:  2014-10-30       Impact factor: 2.502

7.  Cardiac output measurement by arterial pressure waveform analysis during optimization of biventricular pacing after cardiac surgery.

Authors:  Justin H Booth; T Alexander Quinn; Marc E Richmond; Santos E Cabreriza; Alan D Weinberg; Taylor Johnston; Henry M Spotnitz
Journal:  ASAIO J       Date:  2009 Nov-Dec       Impact factor: 2.872

8.  S3 guidelines for intensive care in cardiac surgery patients: hemodynamic monitoring and cardiocirculary system.

Authors:  M Carl; A Alms; J Braun; A Dongas; J Erb; A Goetz; M Goepfert; W Gogarten; J Grosse; A R Heller; M Heringlake; M Kastrup; A Kroener; S A Loer; G Marggraf; A Markewitz; D Reuter; D V Schmitt; U Schirmer; C Wiesenack; B Zwissler; C Spies
Journal:  Ger Med Sci       Date:  2010-06-15

9.  Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices.

Authors:  Xavier Monnet; Nadia Anguel; Brice Naudin; Julien Jabot; Christian Richard; Jean-Louis Teboul
Journal:  Crit Care       Date:  2010-06-10       Impact factor: 9.097

10.  [Noninvasive determination of cardiac output in ventilated patients . Clinical evaluation of a simplified quick method].

Authors:  J Reutershan; T Kapp; K Unertl; R Fretschner
Journal:  Anaesthesist       Date:  2003-09       Impact factor: 1.041

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.