Literature DB >> 10768576

Photon energy recovery for crosstalk correction in simultaneous 99mTc/201Tl imaging.

P Hannequin1, J Mas, G Germano.   

Abstract

UNLABELLED: Simultaneous 99mTc sestamibi/201Tl imaging enables the acquisition of images of myocardial stress perfusion and myocardial viability in a single process. One of the major limits of this technique is the crosstalk of the 99mTc downscattered photons into the 201Tl window. We propose using the spectral deconvolution technique photon energy recovery (PER) for correcting this crosstalk.
METHODS: A planar line phantom made of 99Tc vertical lines and 201Tl horizontal lines and a cardiac SPECT phantom including an anterior (2 mL) and an inferior (1.5 mL) myocardial fixed defect were used. The phantoms were filled with an initial 99mTc/201Tl ratio of 5:1. Several successive acquisitions were made from time t = 0 to time t = 48 h (99mTc/201Tl ratio approximately 0) without moving the phantoms. Total number of counts, contrast, and normalized SD (NSD) were calculated on the Tl-raw and the Tl-PER planar images. SPECT datasets were analyzed. The Tl-raw images recorded at 48 h were considered the reference "virgin" 201Tl images.
RESULTS: Total number of counts, contrast, and NSD ranged from 336% to 201%, 15% to 29%, and 257% to 225% of the virgin 201Tl values, respectively, for Tl-raw planar images; whereas values for Tl-PER images ranged from 128% to 108%, 61% to 79%, and 154% to 108%, respectively. Anterior and inferior defect contrasts ranged from 1.18 to 1.22 and 1.12 to 1.16 for Tl-raw SPECT images, respectively; whereas for Tl-PER images, value ranges were 1.28-1.32 and 1.21-1.24, respectively. The corresponding reference virgin 201Tl values were 1.31 and 1.25 respectively. Summed score, average defect severity, and average defect extent ranges were 4-5, 0.4-0.52, and 4.7-5.9 for Tl-raw images, respectively, and 8-9, 0.59-0.79, and 7.4-8.8 for Tl-PER images. The reference virgin 201Tl values were 9, 0.73, and 8.7, respectively.
CONCLUSION: PER is quantitatively efficient to remove 99mTc crosstalk photons from 201Tl images for 99mTc/201Tl ratios ranging from 5:1 to 2:1.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10768576

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  5 in total

1.  Dual-isotope cardiac SPECT: the Twin Peaks Saga in nuclear cardiology.

Authors:  C Marcassa; O Zoccarato
Journal:  J Nucl Cardiol       Date:  2017-04-27       Impact factor: 5.952

2.  Preliminary clinical results of photon energy recovery in simultaneous rest Tl-201/stress Tc-99m sestamibi myocardial SPECT.

Authors:  P Hannequin; P Weinmann; J Mas; S Vinot
Journal:  J Nucl Cardiol       Date:  2001 Mar-Apr       Impact factor: 5.952

3.  Iterative deconvolution of simultaneous 99mTc and 201Tl projection data measured on a CdZnTe-based cardiac SPECT scanner.

Authors:  Krzysztof Kacperski; Kjell Erlandsson; Simona Ben-Haim; Brian F Hutton
Journal:  Phys Med Biol       Date:  2011-02-11       Impact factor: 3.609

4.  Evaluation of simultaneous 201Tl/99mTc dual-isotope cardiac SPECT imaging with model-based crosstalk compensation using canine studies.

Authors:  Y Du; J M Links; L Becker; A F DiPaula; T Frank; K H Schuleri; A C Lardo; E C Frey
Journal:  J Nucl Cardiol       Date:  2013-12-24       Impact factor: 5.952

5.  Impact of photon energy recovery on the assessment of left ventricular volume using myocardial perfusion SPECT.

Authors:  Alain Manrique; Anne Hitzel; Pierre Véra
Journal:  J Nucl Cardiol       Date:  2004 May-Jun       Impact factor: 5.952

  5 in total