Assessment of prosthetic aortic valve performance by magnetic resonance velocity imaging.

R Botnar1, E Nagel, M B Scheidegger, E M Pedersen, O Hess, P Boesiger.   

Abstract

OBJECTIVES: Magnetic resonance (MRI) velocity mapping was used to evaluate non-invasively the flow profiles of the ascending aorta in normal volunteers and in patients with an aortic (mechanical) valve prosthesis.
BACKGROUND: In patients with artificial aortic valves the flow profile in the ascending aorta is severely altered. These changes have been associated with an increased risk of thrombus formation and mechanical hemolysis.
METHODS: Velocity profiles were determined 30 mm distal to the aortic valve in six healthy volunteers and seven patients with aortic valve replacement (replacement within the last 2 years) using ECG triggered phase contrast MRI. Peak flow, mean flow and mean reverse flow were measured in intervals of 25 ms during the entire heart cycle. Systolic reverse flow, end-systolic closing and diastolic leakage volume were calculated for all subjects.
RESULTS: Peak flow velocity during mid-systole was significantly higher in patients with valvular prosthesis than in normals (mean + SD, 1.9 +/- 0.4 m/s vs. 1.2 +/- 0.03 m/s, P < 0.001) with a double peak and a zone of reversed flow close to the inner (left lateral) wall of the ascending aorta of the patients. Closing volume was significantly larger in patients than in controls (-3.3 +/- 1.2 ml/beat vs. -0.9 +/- 0.5 ml/beat; P < 0.001). There was reverse flow during systole in valvular patients amounting to 15.7 +/- 6.7% of total cardiac output compared to 2.3 +/- 1.2% in controls (P < 0.001). Diastolic mean flow was negative in patients after valve replacement but not in controls (-11.0 +/- 15.2 ml/beat vs. 6.8 +/- 3.2 ml/beat; P < 0.01).
CONCLUSIONS: The following three major quantitative observations have been made in the present study: (1) Mechanical valve prostheses have an increased peak flow velocity with a systolic reverse flow at the inner (left lateral) wall of the ascending aorta. (2) A double peak flow velocity pattern can be observed in patients with bileaflet (mechanical) prosthesis. (3) The blood volume required for leaflet closure and the diastolic leakage blood volume are significantly higher for the examined bileaflet valve than for native heart valves.

Entities:  

Mesh:

Year:  2000        PMID: 10697222     DOI: 10.1007/bf02613108

Source DB:  PubMed          Journal:  MAGMA        ISSN: 0968-5243            Impact factor:   2.310


  30 in total

1.  Appearance of poststenotic jets in MRI: dependence on flow velocity and on imaging parameters.

Authors:  R P Spielmann; O Schneider; F Thiele; M Heller; E Bücheler
Journal:  Magn Reson Imaging       Date:  1991       Impact factor: 2.546

2.  In vitro velocity and turbulence measurements in the vicinity of three new mechanical aortic heart valve prostheses: Björk-Shiley Monostrut, Omni-Carbon, and Duromedics.

Authors:  A P Yoganathan; H W Sung; Y R Woo; M Jones
Journal:  J Thorac Cardiovasc Surg       Date:  1988-05       Impact factor: 5.209

3.  Three-dimensional visualization of velocity fields downstream of six mechanical aortic valves in a pulsatile flow model.

Authors:  J M Hasenkam; M Giersiepen; H Reul
Journal:  J Biomech       Date:  1988       Impact factor: 2.712

4.  In vitro hemodynamic characteristics of tissue bioprostheses in the aortic position.

Authors:  A P Yoganathan; Y R Woo; H W Sung; F P Williams; R H Franch; M Jones
Journal:  J Thorac Cardiovasc Surg       Date:  1986-08       Impact factor: 5.209

5.  Magnetic resonance imaging of blood velocity distribution around St. Jude medical aortic valves in patients.

Authors:  K Houlind; O Eschen; E M Pedersen; T Jensen; J M Hasenkam; P K Paulsen
Journal:  J Heart Valve Dis       Date:  1996-09

6.  Laser anemometry measurements of pulsatile flow past aortic valve prostheses.

Authors:  K B Chandran; G N Cabell; B Khalighi; C J Chen
Journal:  J Biomech       Date:  1983       Impact factor: 2.712

7.  In vitro pulsatile flow velocity and turbulent shear stress measurements in the vicinity of mechanical aortic heart valve prostheses.

Authors:  Y R Woo; A P Yoganathan
Journal:  Life Support Syst       Date:  1985 Oct-Dec

8.  Velocity fields and turbulent stresses downstream of biological and mechanical aortic valve prostheses implanted in pigs.

Authors:  J M Hasenkam; E M Pedersen; J H Ostergaard; H Nygaard; P K Paulsen; G Johannsen; B A Schurizek
Journal:  Cardiovasc Res       Date:  1988-07       Impact factor: 10.787

9.  Quantitation of the turbulent stress distribution downstream of normal, diseased and artificial aortic valves in humans.

Authors:  H Nygaard; P K Paulsen; J M Hasenkam; O Kromann-Hansen; E M Pedersen; P E Rovsing
Journal:  Eur J Cardiothorac Surg       Date:  1992       Impact factor: 4.191

10.  Three-dimensional visualization of velocity profiles in the ascending aorta in humans. A comparative study among normal aortic valves, St. Jude Medical and Starr-Edwards Silastic Ball valves.

Authors:  P K Paulsen; J M Hasenkam; H Stødkilde-Jørgensen; O Albrechtsen
Journal:  Int J Artif Organs       Date:  1988-07       Impact factor: 1.595

View more
  6 in total

1.  Quantification of periprosthetic valve leakage with multiple regurgitation jets by magnetic resonance imaging.

Authors:  A Pflaumer; M Schwaiger; J Hess; R Lange; H Stern
Journal:  Pediatr Cardiol       Date:  2005 Sep-Oct       Impact factor: 1.655

2.  Perioperative evaluation of regional aortic wall shear stress patterns in patients undergoing aortic valve and/or proximal thoracic aortic replacement.

Authors:  Emilie Bollache; Paul W M Fedak; Pim van Ooij; Ozair Rahman; S Chris Malaisrie; Patrick M McCarthy; James C Carr; Alex Powell; Jeremy D Collins; Michael Markl; Alex J Barker
Journal:  J Thorac Cardiovasc Surg       Date:  2017-11-16       Impact factor: 5.209

Review 3.  Heart valve disease: investigation by cardiovascular magnetic resonance.

Authors:  Saul G Myerson
Journal:  J Cardiovasc Magn Reson       Date:  2012-01-19       Impact factor: 5.364

Review 4.  2014 Korean guidelines for appropriate utilization of cardiovascular magnetic resonance imaging: a joint report of the Korean Society of Cardiology and the Korean Society of Radiology.

Authors:  Yeonyee E Yoon; Yoo Jin Hong; Hyung-Kwan Kim; Jeong A Kim; Jin Oh Na; Dong Hyun Yang; Young Jin Kim; Eui-Young Choi
Journal:  Korean J Radiol       Date:  2014-11-07       Impact factor: 3.500

Review 5.  2014 korean guidelines for appropriate utilization of cardiovascular magnetic resonance imaging: a joint report of the korean society of cardiology and the korean society of radiology.

Authors:  Yeonyee E Yoon; Yoo Jin Hong; Hyung-Kwan Kim; Jeong A Kim; Jin Oh Na; Dong Hyun Yang; Young Jin Kim; Eui-Young Choi
Journal:  Korean Circ J       Date:  2014-11-25       Impact factor: 3.243

6.  Ex Vivo Pilot Study of Cardiac Magnetic Resonance Velocity Mapping for Quantification of Aortic Regurgitation in a Porcine Model in the Presence of a Transcatheter Heart Valve.

Authors:  Nynke H M Kooistra; Freek Nijhoff; Masieh Abawi; Pierfrancesco Agostoni; Daniël M Araya Roos; Sjoerd van Tuijl; Niels Blanken; Michiel Voskuil; Pieter A F M Doevendans; Pieter R Stella; Tim Leiner
Journal:  J Cardiovasc Transl Res       Date:  2019-03-15       Impact factor: 4.132

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.