J M Hasenkam1, E M Pedersen, J H Ostergaard, H Nygaard, P K Paulsen, G Johannsen, B A Schurizek.
Abstract
Since detailed knowledge about velocity fields downstream of heart valve prostheses obtained from in vitro studies has not been followed up by similar detailed studies in vivo a pig model for acute velocity field studies downstream of aortic valve prostheses was established. Two mechanical and two bioprosthetic valves were studied and a dynamic three dimensional visualisation of velocity fields one diameter downstream performed under different haemodynamic conditions in a total of 22 pigs. The Ionescu-Shiley pericardial valve had velocity fields very similar to the normal native porcine aortic valve. The Edwards-Carpentier porcine valve caused a jet type flow, and the valve design of the St Jude Medical and Björk-Shiley Monostrut valves was reflected in the velocity profile. Normalised (mean(SEM] systolic Reynolds normal stresses in the total cross sectional area were: native porcine 15(1.5) Nm-2; St Jude Medical 24(3.4) Nm-2; Björk-Shiley Monostrut 25(1.6) Nm-2; Edwards-Carpentier Supra-annular 51(6.6) Nm-2; Ionescu-Shiley Pericardial 19(2.0) Nm-2. Reynolds normal stresses were higher in areas of rapidly changing or constantly high velocity gradients.Entities:
Mesh:
Year: 1988 PMID: 3252971 DOI: 10.1093/cvr/22.7.472
Source DB: PubMed Journal: Cardiovasc Res ISSN: 0008-6363 Impact factor: 10.787