On-line set-up corrections during radiotherapy of patients with gynecologic tumors.

J C Stroom1, M J Olofsen-van Acht, S Quint, M Seven, M de Hoog, C L Creutzberg, H C de Boer, A G Visser.   

Abstract

PURPOSE: Positioning of patients with gynecologic tumors for radiotherapy has proven to be relatively inaccurate. To improve the accuracy and reduce the margins from clinical target volume (CTV) to planning target volume (PTV), on-line set-up corrections were investigated. METHODS AND MATERIALS: Anterior-posterior portal images of 14 patients were acquired using the first six monitor units (MU) of each irradiation fraction. The set-up deviation was established by matching three user-defined landmarks in portal and simulator image. If the two-dimensional deviation exceeded 4 mm, the table position was corrected. A second portal image was acquired using 30 MU of the remaining dose. This image was analyzed off-line using a semiautomatic contour match to obtain the final set-up accuracy. To verify the landmark match accuracy, the contour match was retrospectively performed on the six MU images as well.
RESULTS: The standard deviation (SD) of the distribution of systematic set-up deviations after correction was < 1 mm in left-right and cranio-caudal directions. The average random deviation was < 2 mm in these directions (1 SD). Before correction, all standard deviations were 2 to 3 mm. The landmark match procedure was sufficiently accurate and added on average 3 min to the treatment time. The application of on-line corrections justifies a CTV-to-PTV margin reduction to about 5 mm.
CONCLUSIONS: On-line set-up corrections significantly improve the positioning accuracy. The procedure increases treatment time but might be used effectively in combination with off-line corrections.

Entities:  

Mesh:

Year:  2000        PMID: 10661359     DOI: 10.1016/s0360-3016(99)00386-7

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  7 in total

1.  Uncertainties and CTV to PTV margins quantitative assessment using cone-beam CT technique in clinical application for prostate, and head and neck irradiation tumours.

Authors:  X J Juan-Senabre; J López-Tarjuelo; A Conde-Moreno; A Santos-Serra; A L Sánchez-Iglesias; J D Quirós-Higueras; N de Marco Blancas; S Calzada-Feliu; C Ferrer-Albiach
Journal:  Clin Transl Oncol       Date:  2011-11       Impact factor: 3.405

2.  A hybrid strategy of offline adaptive planning and online image guidance for prostate cancer radiotherapy.

Authors:  Yu Lei; Qiuwen Wu
Journal:  Phys Med Biol       Date:  2010-03-30       Impact factor: 3.609

3.  A survey of image-guided radiation therapy use in the United States.

Authors:  Daniel R Simpson; Joshua D Lawson; Sameer K Nath; Brent S Rose; Arno J Mundt; Loren K Mell
Journal:  Cancer       Date:  2010-08-15       Impact factor: 6.860

4.  Positioning accuracy during VMAT of gynecologic malignancies and the resulting dosimetric impact by a 6-degree-of-freedom couch in combination with daily kilovoltage cone beam computed tomography.

Authors:  Lihong Yao; Lihong Zhu; Junjie Wang; Lu Liu; Shun Zhou; ShuKun Jiang; Qianqian Cao; Ang Qu; Suqing Tian
Journal:  Radiat Oncol       Date:  2015-04-26       Impact factor: 3.481

5.  Patient-specific daily pretreatment setup protocol using electronic portal imaging for radiation therapy.

Authors:  Michael H Wittmer; Thomas M Pisansky; Jon J Kruse; Michael G Herman
Journal:  J Appl Clin Med Phys       Date:  2005-11-21       Impact factor: 2.102

6.  Prospective evaluation of the setup errors and its impact on safety margin for cervical cancer pelvic conformal radiotherapy.

Authors:  Avinash Badajena; Vijay Parshuram Raturi; Kirti Sirvastava; Hidehiro Hojo; Hajime Ohyoshi; Yanping Bei; Toshiya Rachi; Chen-Ta Wu; Taku Tochinai; Masayuki Okumura; Haiqin Zhang; Hirotaki Kouta; Pragya Verma; Geeta Singh; Abhishek Anand; Anjali Sachan
Journal:  Rep Pract Oncol Radiother       Date:  2020-02-22

7.  Optimizing image guidance frequency and implications on margins for gynecologic malignancies.

Authors:  Carmen Stromberger; Arne Gruen; Waldemar Wlodarczyk; Volker Budach; Christhardt Koehler; Simone Marnitz
Journal:  Radiat Oncol       Date:  2013-05-02       Impact factor: 3.481

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.