BACKGROUND: To analyze setup deviations using daily megavoltage computed tomography (MVCT) and to evaluate three MVCT frequency reducing protocols for gynecologic cancer patients treated with helical tomotherapy.Entities:
Mesh:
Year: 2013 PMID: 23638800 PMCID: PMC3671193 DOI: 10.1186/1748-717X-8-110
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Systematic and random deviations
| LO | 0.5 mm | 3.7 mm | 3.4 mm |
| V | −2.0 mm | 5.7 mm | 3.8 mm |
| LA | 0.5 mm | 4.0 mm | 6.1 mm |
Total systematic and random deviations prior to correction for gynecologic patients (n = 56) in the LO, V and LA directions. Shown are the populations’ mean of all systematic deviations (M), the standards deviations of all systematic deviations (Σ) and the mean of all random deviations (σ).
Figure 1Total systematic and random deviations for gynecologic patients (n = 56) in the LO, V and LA directions. Figures include the populations’ mean of all systematic deviations (M), the standards deviations of all systematic deviations (Σ) and the mean of all random deviations (σ).
Figure 2Residual deviations of MVCT protocols. Cumulative distribution of the residual deviations in V, LA and LO directions persisting when applying the FFF (blue rectangle), the FTF (pink triangle), the 505 protocol (green dot) for gynecologic tumor patients. p < 0.05 in V and LA for the 505 protocol.
CTV-PTV margins
| | ||||
|---|---|---|---|---|
| 14.9 | 10.2 | 11.6 | 1.9 | |
| 8.4 | 8.8 | 13.4 | 1.7 | |
| 7.0 | 6.6 | 11.5 | 1.7 | |
| 5.5 | 6.0 | 9.2 | 1.4 | |
The calculated minimal CTV-PTV set-up margins for the vertical (V), longitudinal (LO), lateral (LA) and angular (R) deviations for three different MVCT frequency protocols (FFF: first five fractions; FTF: first ten fractions; 505: on alternate weeks: 1st and 3rd week MVCTs).