| Literature DB >> 11958653 |
Jen-San Tsai1, Mark J Engler, James Liu.
Abstract
The number of linac monitor units (MU) from intensity modulated sequential tomotherapy (IMST) is substantially larger than the MU delivered in conventional radiation therapy, and the relation between MU and dose is obscure due to complicated variation of the beam intensities. The purpose of this work was to develop a practical method of verifying the MU and dose from IMST so that the MU of each arced beam could be double-checked for accuracy. MU calculations for 41 arced beams from 14 IMST patients were performed using the variables of vane open fraction time, field size, target depth, output factor, TMR, and derived intensity distribution. Discrepancy between planned and checked MU was quantified as 100 (MU(cal)-MU(plan))/MU(plan) percent. All 41 discrepancies were clustered between -5% to +4%, illustrated in a Gaussian-shaped histogram centered at -1.0+/-3.5% standard deviation indicating the present MU calculations are in agreement with the planned expectations. To confirm the correctness of the present calculated MUs of the IMST plans, eight of the calculated IMST plans are performed dose verifications using their hybrid plans, which are created by transporting patient's IMST plan beams onto a spherical polystyrene Phantom for dose distribution within the Phantom. The dose was measured with a 0.07 cc ionization chamber inserted in the spherical Phantom during the hybrid plan irradiation. Average discrepancy between planned and measured doses was found to be 0.6+/-3.4% with single standard deviation uncertainty. The spread of the discrepancies of present calculated MUs relative to their planned ones are attributed to uncertainties of effective field size, effective planned dose corresponding to each arc, and inaccuracy of quantification of scattered dose from adjacent arced beams. Overall, the present calculation of MUs is consistent with what derived from treatment plans. Since the MUs are verified by actual dose measurements, therefore the present MU calculation technique is considered adequate for double-checking planned IMST MUs.Entities:
Mesh:
Year: 2002 PMID: 11958653 PMCID: PMC5724610 DOI: 10.1120/jacmp.v3i2.2577
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1IMST beam ares, couch indices, and the coordinate system, T: target.
Figure 2The fraction of each vane's open time. The darker indicates longer open time, while the white vanes indicate all closed. The numerical on the left side shows the numerical fraction of time opened in percentage. The sketch shows the tumor depth variation as gantry rotates during IMST treatment.
Figure 3The tumor depth is varied as the gantry rotates. This figure illustrates how the average depth of the tumor is extracted in the present work of MU calculation. The radial lines are drawn diverging from the tumor center. All the radial lines are in equal angular (10°) intervals. Each radial quantifies a depth from the tumor center to the skin. The average depth is obtained by summing up all the radial line depths and divided by the total number of radial line depths.
Figure 4In the present modeled 〈D〉, the average dose around the tumor within the field size , we assumed .
Figure 5, 〈D〉, and isodose distributions.
Figure 6Measured output factors of the MIMiC for IMST beam dose calculation.
Figure 7A typical beam profile of 2 cm mode MIMiC. The small penumbra indicates that the radiation extending to the adjacent index positions is negligible.
Figure 8The setup of film and Rando Phantom for the study of the radiation dose contributing to the adjacent axial plane.
Figure 9(a) The setup of the spherical phantom and mini‐ionization chamber for absolute dose verification between the IMST plans and that is measured. (b) The dose distribution of the hybrid plan using a Spherical Phantom.
Figure 10The film dose distribution accumulated from its adjacent arcs.
Figure 11Discrepancies of all the IMST plans.
Figure 12The summed extrapatient MU discrepancies between the calculated MU and that from the IMST planned. The dashed curve is the Gaussian‐fitted curve and the vertical dashed line is the center of the Gaussian distribution.
Figure 13The average dose 〈D〉 quantification is sometimes complicated by 2) the high dose gradient of the image which is near the tumor edge, and 2) multiple targets contributed the dose to their high dose areas.
Absolute dose verified from the hybrid plans of the spherical polystyrene phantom.
| Trial # | Verified dose (Gy) | Hybrid planned dose (Gy) | Ratio (verified/hybrid) |
|---|---|---|---|
| 1 | 1.36 | 1.45 | 1.06 |
| 2 | 3.23 | 3.21 | 1.01 |
| 3 | 1.92 | 2.01 | 0.96 |
| 4 | 5.09 | 5.34 | 0.95 |
| 5 | 5.00 | 5.04 | 0.99 |
| 6 | 1.97 | 1.91 | 1.02 |
| 7 | 1.64 | 1.74 | 0.97 |
| 8 | 2.42 | 2.44 | 0.992 |
| Coordinates system | The coordinate system ( |
|
| Dose in Gray (Gy). |
|
| Depth in cm. |
| DVH | Dose volume histogram. |
|
| Gantry rotational axis. |
| IMRT | Intensity modulated radiation therapy. |
| IMST | Intensity modulated sequential tomotherapy, a special category of IMRT. |
| IVSC | Inverse square correction factor relative to the SAD (source to axial distance). |
| MIMiC | Multi‐vane intensity modulated collimator. 40 vanes, in two rows, operated in binary mode, are used for radiation intensity controlling. |
|
| Average field width spanned by the MIMiC at the Gx in the cross‐plane direction. |
| MU | The monitor unit when beam is on. Normally this MU is calibrated so that |
| O/P | Output factor, is the linear accelerator (linac) dose ratio at the maximum dose depth, |
| PCF | Practical Calibration factor. |
|
| A point in 3D space with implicit coordinates ( |
| TMR | Tissue maximum ratio, a dose ratio for a given field sizes at any depth to the maximum dose depth. |
|
| Fraction of time that remains opened in the rotation of gantry angle indices |
|
| The radiation field widths of the MIMiC at 1 cm, 2 cm mode respectively projected along the gantry rotational axis at the isocenter height. These are the widths that are measured at the couch height of isocenter, and the couch should move along the gantry rotational axis for the adjacent arcs irradiation. |
| θ | Gantry angle. |
| Θ | Couch angle. |
|
| The average field size Φ in cm2. |
| %DD | Percentage of depth dose. |