Accuracy of volume measurements in mechanically ventilated newborns: a comparative study of commercial devices.

K Roske1, B Foitzik, R R Wauer, G Schmalisch.   

Abstract

OBJECTIVE: Ventilatory measurements in ventilated newborns are increasingly used to monitor and to optimize mechanical ventilation. The aim of this study was to compare the accuracy of volume measurements by different instruments using standardized laboratory conditions.
METHODS: The accuracy of displayed volume values of different commercial devices (Bicore CP-100, Ventrak 1500, Ventrak 1550, Babylog 8000, PEDS IV and SensorMedics 2600) was investigated using adjustable calibration syringes (volume range 2-60 ml, breathing rates 30/min-60/min) and humidified (>95%), heated (35 degrees C) breathing gas with adjustable FIO2 (0.21-1.0). The pneumotach and also the tubes were placed within an incubator (37 degrees C).
RESULTS: The relative volume error of all devices was in conformity with clinically allowed tolerances (Bicore CP-100 6.4+/-0.5% (mean +/- SD), Ventrak 1500 3.6+/-4.2%, Ventrak 1550 6.5+/-2.7%, Babylog 8000 -5.5+/-1.5%, PEDS IV -4.0+/-1.4%, SensorMedics 2600 3.5+/-1.75%) for the measuring range studied (10 ml < V < 60 ml, rate 30-60/min, FIO2 = 0.21). Unacceptable errors were obtained for volumes lower than 10 ml with Bicore CP-100 (-28.5+/-26%) and PEDS IV (-10.3+/-3.4%). Changes in FIO2 had an important influence on volume measurements and only the SensorMedics 2600 and the PEDS IV corrected properly for FIO2 changes.
CONCLUSION: Most of the currently available neonatal spirometry devices allow sufficiently accurate volume measurements in the range of 10-60 ml and at frequencies between 30-60/min provided that an increased FIO2 is taken into account.

Entities:  

Mesh:

Year:  1998        PMID: 10023838     DOI: 10.1023/a:1009993916980

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  23 in total

1.  A plethysmographic method for assessment of lung function in mechanically ventilated very low birth weight infants.

Authors:  K E Edberg; K Sandberg; A Silberberg; B A Sjöqvist; B Ekström-Jodal; O Hjalmarson
Journal:  Pediatr Res       Date:  1991-11       Impact factor: 3.756

2.  Calibration of Fleisch and screen pneumotachographs for use with various oxygen concentrations.

Authors:  M J Turner; I M MacLeod; A D Rothberg
Journal:  Med Biol Eng Comput       Date:  1990-03       Impact factor: 2.602

3.  Patient-triggered ventilation.

Authors:  A Greenough
Journal:  Pediatr Pulmonol Suppl       Date:  1995

4.  Lung function in infants and young children: functional residual capacity, tidal volume, and respiratory rats.

Authors:  L M Taussig; T R Harris; M D Lebowitz
Journal:  Am Rev Respir Dis       Date:  1977-08

5.  Patient-triggered ventilation in neonates: comparison of a flow-and an impedance-triggered system.

Authors:  H D Hummler; T Gerhardt; A Gonzalez; J Bolivar; N Claure; R Everett; E Bancalari
Journal:  Am J Respir Crit Care Med       Date:  1996-10       Impact factor: 21.405

6.  The accuracy of pneumotachograph measurements during mechanical ventilation.

Authors:  J W Kreit; F C Sciurba
Journal:  Am J Respir Crit Care Med       Date:  1996-10       Impact factor: 21.405

7.  Patient-triggered ventilation: a comparison of tidal volume and chestwall and abdominal motion as trigger signals.

Authors:  W Nikischin; T Gerhardt; R Everett; A Gonzalez; H Hummler; E Bancalari
Journal:  Pediatr Pulmonol       Date:  1996-07

Review 8.  Principles of neonatal assisted ventilation.

Authors:  W A Carlo; R J Martin
Journal:  Pediatr Clin North Am       Date:  1986-02       Impact factor: 3.278

9.  Airway and body surface sensors for triggering in neonatal ventilation.

Authors:  J John; L J Björklund; N W Svenningsen; B Jonson
Journal:  Acta Paediatr       Date:  1994-09       Impact factor: 2.299

10.  Identifying lung overdistention during mechanical ventilation by using volume-pressure loops.

Authors:  J B Fisher; M C Mammel; J M Coleman; D R Bing; S J Boros
Journal:  Pediatr Pulmonol       Date:  1988
View more
  9 in total

1.  Accuracy of deadspace free ventilatory measurements for lung function testing in ventilated newborns: a simulation study.

Authors:  B Foitzik; P Schaller; M Schmidt; G Schmalisch
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

2.  Aerosolization of perfluorocarbons during mechanical ventilation: an in vitro study.

Authors:  Tobias Gregor; Gerd Schmalisch; Wolfram Burkhardt; Hans Proquitté; Roland R Wauer; Mario Rüdiger
Journal:  Intensive Care Med       Date:  2003-04-16       Impact factor: 17.440

3.  Reliability of measured tidal volume in mechanically ventilated young pigs with normal lungs.

Authors:  Mark J Heulitt; Shirley J Holt; Tracy L Thurman; Renée A Hall; Chan-Hee Jo; Pippa Simpson
Journal:  Intensive Care Med       Date:  2005-07-19       Impact factor: 17.440

Review 4.  Journal of clinical monitoring and computing 2014 end of year summary: respiration.

Authors:  D S Karbing; S E Rees; M B Jaffe
Journal:  J Clin Monit Comput       Date:  2015-03-04       Impact factor: 2.502

5.  Distribution of tidal ventilation during volume-targeted ventilation is variable and influenced by age in the preterm lung.

Authors:  Ruth K Armstrong; Hazel R Carlisle; Peter G Davis; Andreas Schibler; David G Tingay
Journal:  Intensive Care Med       Date:  2011-02-25       Impact factor: 17.440

6.  Dead space reduction by Kolobow's endotracheal tube does not justify the waiving of volume monitoring in small, ventilated lungs.

Authors:  Hans Proquitté; Rena Wendel; Charles C Roehr; Roland R Wauer; Gerd Schmalisch
Journal:  J Clin Monit Comput       Date:  2014-01-28       Impact factor: 2.502

7.  How safe is intermittent positive pressure ventilation in preterm babies ventilated from delivery to newborn intensive care unit?

Authors:  M Tracy; L Downe; J Holberton
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-01       Impact factor: 5.747

8.  The administration of 100% oxygen and respiratory drive in very preterm infants at birth.

Authors:  Jeroen J van Vonderen; Nadia E Narayen; Frans J Walther; Melissa L Siew; Peter G Davis; Stuart B Hooper; Arjan B te Pas
Journal:  PLoS One       Date:  2013-10-18       Impact factor: 3.240

9.  Accuracy of currently available neonatal respiratory function monitors for neonatal resuscitation.

Authors:  Charlotte Verbeek; Henriëtte A van Zanten; Jeroen J van Vonderen; Marcus J Kitchen; Stuart B Hooper; Arjan B Te Pas
Journal:  Eur J Pediatr       Date:  2016-06-08       Impact factor: 3.183

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.