Literature DB >> 24469384

Dead space reduction by Kolobow's endotracheal tube does not justify the waiving of volume monitoring in small, ventilated lungs.

Hans Proquitté1, Rena Wendel, Charles C Roehr, Roland R Wauer, Gerd Schmalisch.   

Abstract

In ventilated preterm infants the flow sensor contributes significantly to the total apparatus dead space, which may impair gas exchange. The aim of the study was to quantify to which extent a dead space reduced Kolobow tube (KB) without flow sensor improves the gas exchange compared with a conventional ventilator circuit with flow sensor [Babylog 8000 (BL)]. In a cross-over trial in 14 tracheotomized, surfactant-depleted (saline lavage) and mechanically ventilated newborn piglets (age <12 h; body weight 705-1200 g) BL and KB was applied alternately for 15 min and blood gases were recorded. The inner diameter of the endotracheal tube was 3.6 mm and the apparatus dead space of BL and KB including the endotracheal tube were 3.0 and 1.34 mL. Despite a 50 % apparatus dead space reduction with KB compared to BL statistically significant improvements were only observed for body weights <900 g. In this weight group median paCO2 was decreased by 5 mmHg (p < 0.01), whereas the improvement decreased with decreasing baseline paCO2. Furthermore, median paO2 was increased by 4 mmHg (p < 0.05) and O2 saturation was increased by 2.5 % (p < 0.05). No significant changes were seen in the circulatory parameters. In very small, ventilated lungs the use of KB improved the gas exchange; however, the improvement was moderate and does not justify the waiving of volume monitoring.

Entities:  

Mesh:

Year:  2014        PMID: 24469384     DOI: 10.1007/s10877-014-9559-5

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  23 in total

1.  Accuracy of deadspace free ventilatory measurements for lung function testing in ventilated newborns: a simulation study.

Authors:  B Foitzik; P Schaller; M Schmidt; G Schmalisch
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

2.  A comparison of intratracheal pulmonary ventilation to conventional ventilation in a surfactant deficient animal model.

Authors:  C A Velarde; B L Short; O Rivera; W Seale; R Howard; T Kolobow; K Rais-Bahrami
Journal:  Crit Care Med       Date:  2000-05       Impact factor: 7.598

3.  Aspiration of dead space allows isocapnic low tidal volume ventilation in acute lung injury. Relationships to gas exchange and mechanics.

Authors:  E De Robertis; G Servillo; R Tufano; B Jonson
Journal:  Intensive Care Med       Date:  2001-09       Impact factor: 17.440

4.  Augmented diffusion in the airways can support pulmonary gas exchange.

Authors:  J J Fredberg
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1980-08

5.  Impact of volume guarantee on synchronized ventilation in preterm infants: a randomized controlled trial.

Authors:  Nuray Duman; Funda Tuzun; Sumer Sutcuoglu; Cemile Didem Yesilirmak; Abdullah Kumral; Hasan Ozkan
Journal:  Intensive Care Med       Date:  2012-05-23       Impact factor: 17.440

Review 6.  The new bronchopulmonary dysplasia.

Authors:  Alan H Jobe
Journal:  Curr Opin Pediatr       Date:  2011-04       Impact factor: 2.856

7.  Accuracy of volume measurements in mechanically ventilated newborns: a comparative study of commercial devices.

Authors:  K Roske; B Foitzik; R R Wauer; G Schmalisch
Journal:  J Clin Monit Comput       Date:  1998-08       Impact factor: 2.502

8.  Lung inflammation in preterm infants with respiratory distress syndrome: effects of ventilation with different tidal volumes.

Authors:  Gianluca Lista; Francesca Castoldi; Paola Fontana; Roberta Reali; Alessandro Reggiani; Silvia Bianchi; Gilberto Compagnoni
Journal:  Pediatr Pulmonol       Date:  2006-04

9.  The impact of instrumental dead-space in volume-targeted ventilation of the extremely low birth weight (ELBW) infant.

Authors:  Sepideh Nassabeh-Montazami; Kabir M Abubakar; Martin Keszler
Journal:  Pediatr Pulmonol       Date:  2009-02

Review 10.  Tracheal gas insufflation for the prevention of morbidity and mortality in mechanically ventilated newborn infants.

Authors:  M W Davies; P G Woodgate
Journal:  Cochrane Database Syst Rev       Date:  2002
View more
  1 in total

Review 1.  Journal of clinical monitoring and computing 2014 end of year summary: respiration.

Authors:  D S Karbing; S E Rees; M B Jaffe
Journal:  J Clin Monit Comput       Date:  2015-03-04       Impact factor: 2.502

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.